Калибровка терминала IND560 (Mettler-Toledo)

Калибровка терминала IND560 (Mettler-Toledo)

User's Guide
IND560x
Weighing Terminal

64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 1-1
This chapter covers
IND560x Overview
Terminal Versions
Warnings and Precautions
Operating Environment
Inspection and Contents Checklist
Physical Dimensions
Specifications
Main PCB
IND560x Options
Safe Area Communication Options
Additional Accessories
Display and Keypad
1 Introduction
1.1. IND560x Overview
The IND560x represents the latest in METTLER TOLEDO technology and is the
most versatile weighing terminal available today. The IND560x is suitable for
use in industrial weighing applications inside hazardous (classified) areas. The
IND560x terminal has global agency approvals that permit operation within
Division 1 or Zone 1/21 hazardous locations.
The IND560x terminal is a high-performance single- or multiple-range weighing
terminal designed for use with analog load cells or high-precision IDNet scale
interfaces. The IND560x delivers precision measurement data from milligrams to
tons in a single cost effective package that easily integrates into existing
systems.
Power options include an external power supply that converts high level AC
voltage to intrinsically safe DC voltages and is still suitable for installation inside
the hazardous area. An external battery for use in mobile applications is also
available for use with the harsh/analog model of the IND560x.
PLC and PC communication interfaces are available through easy to install safe area
communication modules. Internal and remote digital I/O options provide control for process
applications such as filling. The versatile IND560x can be upgraded with a variety of special
application software packages that add performance features where needed. These features make
the IND560x the perfect match for nearly any weighing application in many industries, including:
Pharmaceuticals Petrochemicals Refining
Milling Agriculture Cosmetics and Fragrance
Specialty Chemical Coatings and Inks

1-2 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Introduction
1.2. IND560x Terminal Versions
Figure 1-1: Harsh (left) and Panel Mount (right) Enclosures for the IND560x
The IND560x terminal is available in the following versions:
Harsh enclosure with analog load cell connection
Harsh enclosure with high-precision (IDNet) base connection
Panel-mount enclosure with analog load cell connection
Panel-mount enclosure with high-precision (IDNet) base connection
1.3. Standard IND560x Features
Basic weighing in hazardous (explosive) areas including zero, tare and printing
Panel-mount or harsh desk/wall-mount enclosures
Connect one analog load cell scale base (or up to four 350 ohm load cells) or an IDNet base
depending upon the version of the IND560x
128 × 64 dot-matrix graphic backlit liquid crystal display (LCD) with 21mm high weight
display
Real-time clock (battery backup)
Includes one intrinsically safe serial port (COM1) for asynchronous, bidirectional
communication and print output
Operates with the intrinsically safe APS768x or PSUx power supply. An external NiMH battery
pack is available for use with the harsh/analog model
Supports the following internal option boards:
Discrete I/O Interfaces
Dual channel intrinsically safe current loop interface
or
Dual channel fiber optic interface

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 1-3
Support for the following options utilizing the ACM500 communication module in the safe area:
Analog Output DeviceNet
EtherNet/IP – PROFIBUS® DP
Allen-Bradley® RIO
Ethernet TCP/IP with two serial
ports (Ethernet/COM2/COM3)
Modbus TCP
ID mode for prompted transaction sequencing
Selectable over/under classifying mode of operation
Selectable material transfer mode for simple filling or dosing
SmartTrac™ graphical display
Comparators for simple comparison of weight or rate
Two memory tables – store 25 Tare and 25 Target records
Unit switching between three different units, including custom units
Alibi memory storage for up to 60,000 records
Grand total and subtotal registers for accumulating weight
Five customizable print templates and report printing
TraxDSP™ digital filtering for analog load cells
TraxEMT™ performance monitoring and recording
CalFREE™ calibration without test weights
Support of the following Application Software Modules:
Fill-560 FillPlus COM-560
Drive-560 TaskExpert®

1-4 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Introduction
1.4. Warnings and Precautions
Please read these instructions carefully before putting the new terminal into operation.
Although the IND560x is ruggedly constructed, it is nevertheless a precision instrument. Use care in
handling and installing the terminal.
WARNING
DO NOT INSTALL OR PERFORM ANY SERVICE ON THIS EQUIPMENT BEFORE THE AREA
HAS BEEN SECURED AS NON-HAZARDOUS BY PERSONNEL AUTHORIZED TO DO SO BY
THE RESPONSIBLE PERSON AT THE CUSTOMER’S SITE.
WARNING
ONLY THE COMPONENTS SPECIFIED IN THIS MANUAL CAN BE USED IN THIS TERMINAL.
ALL EQUIPMENT MUST BE INSTALLED IN ACCORDANCE WITH THE INSTALLATION
INSTRUCTIONS DETAILED IN THIS MANUAL. INCORRECT OR SUBSTITUTE COMPONENTS
AND/OR DEVIATION FROM THESE INSTRUCTIONS CAN IMPAIR THE INTRINSIC SAFETY OF
THE TERMINAL AND COULD RESULT IN BODILY INJURY AND/OR PROPERTY DAMAGE.
1.5. Equipotential Bonding (EB)
Equipotential bonding must be installed by an electrician authorized by the owner. METTLER
TOLEDO Service performs only a monitoring and consulting function for this procedure.
Connect equipotential bonding of all devices (power supply unit, weighing terminal, interface
converter and weighing platform) in accordance with the terminal diagram and all country-specific
regulations and standards. In the process, it must ensured that:
All device housings are connected to the same potential via the EB terminals.
No circulating current flows via the cable shielding for intrinsically safe circuits.
The neutral point for equipotential bonding is as close to the weighing system as possible.
1.5.1. Warnings
1.5.1.1. 20BCENELEC
1.5.1.1.1. Connection of EB per Country-Specific Regulations
It must be ensured that the housings of all devices are connected to the same potential via the EB
terminals. No circulating current may flow via the shielding of the intrinsically safe cables.
1.5.1.2. 21BcFMus
1.5.1.2.1. Connection of EB per ANSI/NFPA 70, Article 504, and ANSI/IA RP 12.06.01 or Canadian Electric Code C22.2
It must be ensured that the housings of all devices are connected to the same potential via the EB
terminals. No circulating current may flow via the shielding of the intrinsically safe cables.
 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 1-5
1.6. Operating Environment
When selecting a location:
Use the weighing terminal only when electrostatic processes leading to propagation brush
discharge is impossible.
Keep the terminal away from processes that generate high charging potential such as
electrostatic coating, rapid transfer of non-conductive materials, rapid air jets, and high
pressure aerosols.
Choose a stable, vibration-free surface to mount the terminal
Ensure there are no excessive fluctuations in temperature and no direct exposure to sunlight
Avoid drafts on the weighing platform (for example, from open windows or air conditioning)
Calibrate the terminal after any major change of geographical location
1.6.1. Temperature and Humidity
The IND560x can be stored and operated at temperatures and relative humidity conditions as listed
in Table 1-1.
1.6.2. Environmental Protection
The IND560x harsh and panel-mount terminals have environmental protection as listed in Table
1-1.
1.6.3. Hazardous Areas
The intrinsically safe IND560x terminal is designed to operate within an area classified as Division
1, Zone 1 or Zone 21.
IMPORTANT: Not all models in the IND560 terminal family can be operated in areas classified
as Hazardous because of the combustible or explosive atmospheres in those areas. Before
installation, check the data plate of the terminal to confirm it is marked with the approval for the
environment in which it will be used.
1.7. Inspection and Contents Checklist
Verify the contents and inspect the package immediately upon delivery. If the shipping container is
damaged, check for internal damage and file a freight claim with the carrier if necessary. If the
container is not damaged, remove the IND560x terminal from its protective package, noting how it
was packed, and inspect each component for damage.
If shipping the terminal is required, it is best to use the original shipping container. The IND560x
terminal must be packed correctly to ensure its safe transportation.
The IND560x package should include:
IND560x Terminal Bag of miscellaneous parts

1-6 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Introduction
Printed Installation manual Gasket (panel-mount only)
Documentation CD
1.8. Physical Dimensions
The dimensions for the panel-mount and harsh enclosures of the IND560x terminal are shown
below.
1.8.1. Panel-mount Enclosure
The physical dimensions for the panel-mount enclosure of the IND560x terminal are shown in
Figure 1-2 in inches and [mm].
Figure 1-2: IND560x Panel-Mount Enclosure Dimensions
 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 1-7
The physical dimensions for the panel cutout of the IND560x terminal are shown in Figure 1-3 in
inches and [mm].
Figure 1-3: IND560x Panel-Mount Cutout Dimensions

1-8 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Introduction
1.8.2. Harsh Enclosure
The physical dimensions for the harsh desk and wall-mount enclosures of the IND560x terminal
are shown in Figure 1-4 and Figure 1-5, in inches and [mm].
Figure 1-4: IND560x Harsh, Desk Mount Dimensions

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 1-9
Figure 1-5: IND560x Harsh Wall Mount Dimensions - Shown With Optional Wall Mounting Bracket
1.9. Specifications
The IND560x terminal conforms to the specifications listed in Table 1-1.
Table 1-1: IND560x Specifications
IND560x Specifications
Enclosure Type Panel Mount: stainless steel front panel with an aluminum chassis
Harsh Environment desk/wall/column-mount: type 304L stainless steel
enclosure
Dimensions (h ×
w × d)
Panel Mount: 160 mm x 265 mm × 92 mm
(6.3 in. × x 10.4 in x 3.6 in.)
Harsh Environment: 170 mm × 289 mm × 183 mm
(6.7 in. × 11.4 in. × 7.2 in.)
Shipping Weight 3.5 kg (8 lb)

1-10 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Introduction
IND560x Specifications
Environmental
Protection
Panel Mount front panel sealing provides IP65 (comparable to TYPE 4 and
TYPE 12 protection)
-certified by DEKRA EXAM GmbH.
Harsh Environment enclosure meets IP65 requirements
certified by DEKRA EXAM GmbH, and IP68 and 69K certified by MTLS.
Storage
Environment
The terminal can be stored at temperatures ranging from –20°C to 60° C (–4°
to 140° F) at 10 to 95% relative humidity, non-condensing.
Operating
Environment
The terminal can be operated at temperatures ranging from -10° to 40° C (14°
to 104° F) at 10% to 95% relative humidity non-condensing.
Hazardous Areas The IND560x is approved for use in Division 1 and Zone 1/21 2 GD
hazardous areas.
Note that not all models in the IND560 terminal family can be operated in areas
classified as Hazardous because of the combustible or explosive atmospheres
in those areas. Before installation, check the data plate of the terminal to
confirm it is marked with the approval for the environment in which it will be
used.
Power All models of the IND560x can operate with the PSUx or APS768x intrinsically safe
power supplies. The PSUx and APS768x both provide several intrinsically safe
voltages to the IND560x for its correct operation. The PSUx and APS768x are available
in either a 120 VAC or a 230 VAC version.
The IND560x harsh/analog model with the IND560x-PAB option installed operates with
the METTLER TOLEDO intrinsically safe, external NiMH battery pack
(P/Ns 64060627 & 64060625).
Display 128 × 64 dot-matrix graphic LCD display with LED backlight. 21 mm weight
digits.
Display Update Rate: 10/second
Weight Display Displayed resolution of 100,000 counts for analog load cell scales.
Display resolution for high-precision IDNet bases is determined by the specific
base used.
Scale Types Analog load cells or High-Precision IDNet, Kx Line
(T-Brick Ex T4 type) or Point Ex module.
NOT COMPATIBLE WITH THE OLDER, PIK EX MODULE.
Number of
Analog Cells
Four 350-ohm load cells (87 ohm minimum resistance), 2 or 3 mv/V
Number of
Scales
Interface for one analog or one IDNet scale
Analog/Digital
Update Rates
Internal: Analog: >366 Hz; IDNet: determined by base;
Target Comparison: up to 50 Hz;
PLC Interface: up to 20 Hz

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 1-11
IND560x Specifications
Load Cell
Excitation
Voltage
5 VDC
Minimum
Sensitivity
0.3 µV/e approved
Keypad 26 keys; 1.5-mm thick polyester overlay (PET) with hardcoating.
Polycarbonate display lens (PC).
Communications Standard Interfaces:
One intrinsically safe RS-232 serial port (COM1), 300 to 115,200 baud
Optional Interfaces:
Ethernet/2 Serial Ports: Ethernet 10 Base-T and two additional serial ports
(COM2 and COM3) available with ACM500 in safe area.
Optional fiber optic interface communicates to ACM500 or legacy dual channel
fiber optic converter.
-- or --
Optional dual-channel intrinsically safe current loop interface to ACM500 for
PLC interface and/or Ethernet/COM2/COM3.
-- or --
Optional dual-channel intrinsically safe current loop interface to one or two
ACM200 modules for RS-232 serial port (COM4, COM5)
-- or --
IND560x with intrinsically safe interface directly to IND226x with InterfaceRemote installed for hazardous area remote display.
Protocols
Serial Inputs: Keypad input, ASCII commands for CTPZ (Clear, Tare, Print,
Zero), SICS (most level 0 and level 1 commands)
Serial Outputs: Continuous or Demand with up to five configurable print
templates or SICS host protocol, report printing, interfaces with external
ACM500 in safe area for Ethernet, COM2, COM3 and PLC interface.
PLC Interfaces Optional Allen-Bradley RIO, PROFIBUS DP, DeviceNet, EtherNet/IP, Modbus TCP
or Analog output are available using an expansion slot in the ACM500
(Accessory Communication Module) in the safe area.

1-12 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Introduction
IND560x Specifications
Approvals Weights and Measures
USA: Class II, 100,000d; Class III/IIIL, 10,000d; CoC #05-057A2
Canada: Class II, 100,000d; Class III, 10,000d; Class IIIHD, 20,000d;
AM-5593 Rev 1
Europe: Class II, resolution determined by platform approval; Class III
10,000e; TC6812 Rev 5
Hazardous Area
U.S.: IS, Class I, II, III; Div 1; Groups A-G T4
Ta=-10ºC to +40ºC, IP65, Type 4
I/1/AEx ib IIC T4
ATEX: Certificate BVS 09 ATEX E 010 and 1st
supplement
II 2 G Ex ib [op is Ga] IIC T4 Gb
II 2 D Ex ib [op is Da] IIIC T60°C Db
IP65
Canada: Class I, II, III; Div 1; Groups A-G T4
Ta=-10ºC to +40ºC, IP65
IECEx: Certificate IECEx BVS 10.0096X
Ex ib [op is Ga] IIC T4 Gb
Ex ib [op is Da] IIIC T60°C Db
IP65
1.10. Manufacture Date
The product manufacturing date or the date code for the terminal can be found on the serial data
plate (on the top of the enclosure).
1.11. Main PCB
The IND560x terminal’s main printed circuit board (PCB) provides the scale interface for analog
load cell or IDNet. The terminal must be ordered with the required scale interface - it cannot be
modified later.
Note: Connections for both the Analog load cells and an IDNet base are present on the main
board of the IND560x simultaneously. However, only one connection has been enabled in the
firmware of the board.
The main board also contains the DC power input connections, keyboard interface, bus connector
for the display, bus connector for the discrete I/O option board and a connector for the intrinsically
safe current loop or the fiber optic interface board (COM4 and COM5) connections.
An additional option board known as the IND560x-PAB (Power Adapter Board) must be installed
onto the main board of the IND560x to make the terminal compatible with the external NiMH battery

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 1-13
pack. The IND560x-PAB can only be used with revision D or higher of the analog version of the
main board. Refer to the Installation Manual for more information regarding the IND560x-PAB.
Note: The revision level is not indicated on the IND560x main board. Revision D (and higher)
main boards can be identified by the presence of the W3 jumper. If a main board has a W3
main jumper, it is at least revision D or higher. If the W3 jumper is not present, the main board
is lower than revision D and cannot be used with the IND560x-PAB.
1.11.1. Analog Load Cell Scale Base
The IND560x can provide power for up to an 87 ohm load (four 350-ohm analog load cells). A
jumper permits configuration for 2mV or 3mV load cells; the 3mV position is selected at the factory.
A removable terminal strip is provided for the load cell connection.
Note: When installed, the IND560x-PAB will cover the mV jumper. Please take this into
consideration during installation of the IND560x analog model.
1.11.2. IDNet Scale Base
The IND560x supports the newer T-Brick Ex T4 style of intrinsically safe high-precision base
through the main board IDNet port. This port provides the power and communication required to
operate this newer style base.
Note: The older PIK Ex module is not supported by the IND560x.
Note: IND560x IDNet terminals cannot be used with the IND560x-PAB and external NiMH
battery pack.
1.11.3. COM1 Serial Port
The main board also contains the COM1 serial port that provides intrinsically safe RS-232
communication. The port is bidirectional and can be configured and used for various functions
such as demand output, SICS host communications, continuous output, ASCII command input (C,
T, P, Z), ASCII character input, report printing, Shared Data Server access, or connection with the
InSite PC Tool.
Note: The approval entity parameters for the COM1 serial port must be checked for compatibility
with those of the remote device when making direct connections to this port inside the
hazardous area. When connecting COM1 to a peripheral in the safe area, a barrier will be
required to properly protect the port and ensure safety of the system. Compare entity values of
the IND560x terminal and the barrier to confirm safe compatibility before connecting.
1.12. IND560x Options
Some of the options and accessories available for use with the IND560x are described in the
following sections. Note that some options require the use of the ACM500 communication module
in the safe area.
1.12.1. Hardware Options
The following internal hardware options are available for the IND560x terminal:
Fiber optic interface to connect to the ACM500 accessory communication module

1-14 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Introduction
Intrinsically safe current loop interface to connect to the ACM500 module or ACM200 module
Intrinsically safe discrete I/O with 4 inputs and 6 outputs
IND560x-PAB (Power Adapter Board) to make the IND560x terminal compatible with the
external NiMH battery pack
1.12.1.1. Dual Channel Fiber Optic Interface
A dual channel fiber optic interface can be used to connect an IND560x terminal to the ACM500
located in the safe area. This interface also provides connectivity to the older, dual-channel fiber
optic converter (from an 8141, 8525, Puma, or ID3sTx installation).
The fiber optic interface provides one or two channels of communication designed to use a
hard clad silica fiber optic cable. High speed transmitters and receivers are used to increase the
throughput of data. When combined with the ACM500 communication module and its options,
this permits remote (safe area) operation of COM2 and COM3 serial ports, Ethernet and PLC
interfaces at distances up to 300 meters (1000 ft.) from the IND560x terminal.
When used with the standard dual-channel fiber optic converter, this interface provides one or
two additional bi-directional serial ports (COM4 and COM5) at the same distance.
The fiber optic interface can be used to communicate directly between two IND560x terminals
in a Master-Remote configuration
NOTE: Plastic core fiber optic cable can be used on short cable runs of 40 m (130 ft) or
less.
1.12.1.2. Dual Channel IS Current Loop Interface
Instead of the fiber optic cable interface, an intrinsically safe, dual channel current loop (CL)
interface can be used to connect an IND560x terminal to the ACM500 located in the safe area. This
interface can also provide connectivity to the ACM200 converter in the safe area. Alternatively, the
IND560x can be directly connected to an IND226x terminal acting as a remote display inside the
hazardous area. In this scenario, the IND226x must be equipped with the Interface-Remote option.
The current loop interface provides one or two channels of communication designed to use a
copper wire cable. High speed transmitters and receivers are used to increase the throughput of
data. When combined with the ACM500 communication module and its options, this permits
remote (safe area) operation of COM2 and COM3 serial ports, Ethernet and PLC interfaces at
distances up to 300 meters (1000 ft.) from the IND560x terminal.
When used with the ACM200 safe area communication module, this interface provides an
additional serial port (COM4 or COM5) at the same 300 meters (1000 ft.) distance from the
IND560x terminal.
This interface can also be used to operate an IND560x or IND226x terminal as a remote
display to the IND560x. The remote terminal can be located up to 300 meters (1000 ft.)
distance from the IND560x terminal.
1.12.1.3. Discrete I/O
Discrete I/O interface options include both internal and remote I/O. A maximum of 12 inputs and 18
outputs can be supported through a total of three options (internal and external, combined).

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 1-15
The internal discrete I/O option is available in three different combinations of active/passive
inputs and active/passive outputs. Both the inputs and outputs use solid state devices.
Active inputs and active outputs: The inputs are powered internally by the IND560x terminal
and are designed for use with simple switches within the hazardous area. The inputs are
not approved for direct connection to devices in the safe area - a remote ARM100 should
be used for this. The outputs are also powered internally by the IND560x terminal and
provide 12V switching at 50mA total. These outputs are intended for use with extremely low
power intrinsically safe piezo electric solenoids.
Note: The Active/Active I/O option is not available for use with the IND560x-PAB and
external NiMH battery pack.
Active inputs and passive outputs: The inputs are powered internally by the IND560x
terminal and are designed for use with simple switches within the hazardous area. The
inputs are not approved for direct connection to devices in the safe area - a remote
ARM100 should be used for this. Power to operate the outputs must be provided by an
external power supply and intrinsically safe barrier. Refer to the approval entity values for
restrictions and limitations.
Passive inputs and passive outputs: Power to operate both the inputs and outputs must be
provided by an external power supply and intrinsically safe barrier.
Safe area remote I/O can be configured to communicate to an ARM100 device (refer to page 1-
19) using the ACM200 with the CL/RS-422/RS-485 option or the ACM500 configured with
Ethernet/COM2/COM3 option (refer to page 1-17 for additional details on the ACM500 and
page 1-20 for ACM200). One other possibility is to use COM1 on the IND560x main board.
This signal can be passed through an intrinsically safe barrier (ISB) and then through an RS-
232 to RS-485 converter, both located in the safe area. The ARM100 provides six dry-contact
outputs and four passive low voltage DC inputs. An external 12 - 24 volt DC supply is required
to operate the ARM100.
NOTE: The ARM100 is NOT approved for use inside hazardous classified areas. An ARM100
must be installed in the safe area only.
1.12.1.4. IND560x-PAB
NOTE: Due to space limitations and battery life limitations, the IND560x-PAB (Power Adapter
Board) can only be used with the Harsh/Analog version of the IND560x. The IND560x-PAB will
not work with any panel mount or IDNet models of the IND560x. The IND560x-PAB and
external NiMH battery pack will also not support the Active-Active I/O option.
The IND560x-PAB option serves as a voltage splitter for the IND560x’s external NiMH battery. The
IND560x-PAB takes in the single voltage source from the external NiMH battery and splits it into the
various voltages that are required for specific functions in the IND560x. The IND560x with PAB can
only be used with the external, 8Ah NiMH battery pack, part #(s) 64060625 and 64060627.
1.12.2. Software Applications
The following add-on software applications are available for the IND560x terminal.
Fill-560 (filling and dosing application)
Fill-Plus (standard Fill-560 with a simple formula management add-on application)
Drive-560 (simple vehicle management application)

1-16 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Introduction
COM-560 (legacy host protocol application)
TaskExpert (custom programming capability)
1.12.2.1. Fill-560
The Fill-560 is a special application that can be added to the IND560x terminal to provide material
filling and dosing control for up to four materials. It provides control for the following combinations
of weigh-in and weigh-out sequences.
Fill only Blend only
Fill and dump Blend and dump
Dose out only Blend and dose out
Fill and dose out
Additional information can be found in the Fill-560 Technical Manual on the documentation CD
included with the Fill-560 option.
Only one formula can be structured at a time. Once a change is made to any formula, its
previous configuration is lost.
1.12.2.2. FillPlus
FillPlus is an add-on program created for the Fill-560 application. Like Fill-560, FillPlus is a special
application that can be added to the IND560x terminal to provide filling and dosing control for up to
4 materials.
FillPlus provides a formula management and storage program that adds the following features to
the Fill-560 application software:
Storage of up to 25 multiple material formulas
Scaling of formulas
A single auxiliary output assignable for each formula
The FillPlus application is written with TaskExpert™. The tasks associated with FillPlus run
independently of the Fill-560 base code.
Additional information is available in the Fill-560 & FillPlus Technical Manual, part # 64057366,
found on the documentation CD that accompanies the Fill-560 and FillPlus application software
kits, or a terminal that has been configured at the factory with the Fill-560 or FillPlus software.
1.12.2.3. Drive-560
The Drive-560 option is a specialized application solution focused on simple inbound-outbound
vehicle weighing requirements. An IND560drive has two modes of operation: Temporary Tare ID
Weighing and Permanent Tare ID Weighing. Some features of this software include:
Storage of up to 100 Permanent Tare IDs
Totalization of Permanent Tare IDs
One-Step processing of Temporary IDs
 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 1-17
Reprint of previous transaction ticket
Storage of up to 2000 transactions
Inbound weight/date/time available on both Inbound and Outbound tickets
Additional information can be found in the Drive-560 Manual, part # 64062160, found on the
documentation CD that accompanies the Drive-560 application.
1.12.2.4. COM-560
The COM-560 option is a specialized software module solution focused on the needs of users
utilizing legacy communication protocols or that have special command requirements. The
IND560com maintains all of the standard features and functions of the IND560 in addition to the
specific features and functions of the COM-560. It provides the following features and functions:
Custom ASCII command template
Mettler Toledo Continuous Short output
8142 Host Protocol
8530 Host Protocol
PT6S3 Protocol
SMA Protocol
Additional information can be found in the COM-560 Technical Manual on the documentation CD
included with the module.
1.12.2.5. TaskExpert™
TaskExpert functionality provides a way to modify the standard capabilities of the terminal so it can
more closely align with specific application requirements. TaskExpert is a combination of a
programming visualization tool, an execution engine and the basic functionality of the terminal.
Modifications may be made to the sequence of operation and additional functionality added to the
basic operation of the terminal.
1.13. Safe Area Communication Options
1.13.1. ACM500 Communication Module and Accessories
Figure 1-6: ACM500 Module

1-18 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Introduction
When Ethernet access, additional serial ports, or a PLC interface is required, an ACM500
communication module must be connected to the IND560x. The ACM500, which must be located
in a non-hazardous area, receives data from the IND560x through the fiber optic interface or the
intrinsically safe current loop interface. The ACM500 provides expansion slots for an
Ethernet/COM2/COM3 option board and a single PLC option. Refer to the ACM500 Technical
Manual or Installation Manual for details.
The following options for the IND560x require the use of the ACM500 communication module in the
safe area:
Ethernet/COM2/COM3
Programmable Logic Control (PLC) interfaces
Remote discrete I/O (in safe area) via ARM100 modules
1.13.2. Ethernet/COM2/COM3
The Ethernet port can be used for FTP transfer of tare and target tables and complete setup files. It
also provides a TCP/IP port to transmit demand templates, continuous data, for access to data via
the shared data server and to send email alerts when calibration expires or fails. This port can also
be used for backing up and restoring the terminal configuration, using the METTLER TOLEDO
InSite™ SL program.
COM2 provides RS-232 communication at rates from 300 to 115.2k baud. COM3 supports the
same baud rates and provides an RS-232, RS-422, or RS-485 connection. COM3 can be used to
update the firmware of the IND560x using the InSite™ CSL program.
When using COM2 or COM3 serial ports in the ACM500, it is recommended to use a network
transmission baud rate of at least 9,600 Baud. Using COM2 or COM3 serial ports with network
transmission baud rates of 4,800 Baud or less may result in continuous data lagging behind real
time conditions, or a delayed response to on-demand data requests (e.g., SICS requests).
1.13.3. PLC Interfaces
The IND560x terminal requires the use of the ACM500 module in the safe area to provide a PLC
interface. Interfaces supported include Analog Output, A-B RIO, DeviceNet, EtherNet/IP, Modbus TCP
and PROFIBUS DP.
Additional details on these interfaces can be found in the IND560 PLC Interface Manual on the
IND560x Documentation CD.
1.13.3.1. Analog Output
Analog Output refers to the representation of an internal system variable using a proportional
electrical signal. Analog Output can be used to transmit a measured value, such as the gross or net
weight or a rate. Another use for the Analog Output is as a control signal for some external device,
such as a control valve, where the amount of valve opening is proportional to the analog signal
commanding its operation. Such outputs are usually used to control the flow rate of material into or
out of a vessel.
Both 0-10 volt DC and 4-20 mA signals are provided. Only one signal can be used at a time.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 1-19
1.13.3.2. Allen-Bradley (A-B) RIO
The A-B RIO option enables data exchange by bi-directional communications using the Discrete
Data Transfer or Block Transfer mode. The terminal provides a communication exchange with the
PLC approximately 20 times per second utilizing the Allen-Bradley Discrete Data Transfer protocol.
This communication is a high-speed, real-time message interface between the terminal and the PLC
for process control. Division, integer, and floating point values are supported.
The A-B RIO interface also supports Block Transfer mode for transmission of larger amounts of
data.
1.13.3.3. DeviceNet
DeviceNet is an RS-485 based network utilizing CAN chip technology. The network can be
configured to run up to 500Kbits per second depending on cabling and distances. Messages are
limited to 8 un-fragmented bytes. The network is capable of supporting up to 64 nodes including
the master, commonly called the scanner.
1.13.3.4. EtherNet/IP
EtherNet/IP, short for "EtherNet Industrial Protocol," is an open industrial networking standard that
takes advantage of commercial, off-the-shelf EtherNet communication chips and physical media.
This networking standard supports both implicit messaging (real-time I/O messaging) and explicit
messaging (message exchange). The protocol is supported by ControlNet International (CI), the
Industrial Ethernet Association (IEA) and the Open DeviceNet Vendor Association (ODVA).
1.13.3.5. Modbus TCP
Modbus/TCP is used to establish master-slave/client-server communication between intelligent
devices. It is an open standard network protocol, widely used in the industrial manufacturing
environment. The ModbusTCP protocol takes the Modbus instruction set and wraps TCP/IP around
it.
1.13.3.6. PROFIBUS DP
The IND560x Terminal communicates to a PROFIBUS-DP master according to DIN 19 245. The
PROFIBUS option consists of software that resides in the IND560x Terminal and a printed circuit
board that installs in the ACM500 communication module to implement the data exchange.
1.13.4. ARM100 Module
Figure 1-7: ARM100 Module

1-20 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Introduction
The ARM100 Remote Discrete I/O Module is designed to extend the control network of METTLER
TOLEDO industrial terminals. This high performance module provides four inputs and six drycontact relay outputs at a remote location from a compatible terminal or communication module.
Digital control information is communicated between the ARM100 and the terminal via an RS-485
communication link.
An external 10 – 32 VDC power supply is required to operate the ARM100 module. The ARM100 is
only approved for installation in areas classified at SAFE.
1.13.5. ACM200 Safe Area Communication Module
Figure 1-8: ACM200 Safe Area Communication Module
The ACM200 converter can be used to convert the intrinsically safe current loop information from
COM4 or COM5 of the IND560x terminal to an RS-232 signal inside the safe area. The ACM200
converter can be used to obtain demand or continuous RS-232 data from the IND560x or can
convert RS-232 signals in the safe area to intrinsically safe ASCII data and send to the IND560x.
For a CL20mA or RS-422 or RS-485 signal inside the safe area, the ACM200 configured with the
CL/RS-422/RS-485 option can be used.If the ACM500 is already connected to the IND560x
terminal, it is not possible to connect the ACM200 converter. This is because both devices require
dedicated use of the COM4 and COM5 ports of the IND560x terminal. The ACM200 is designed to
communicate to the IND560x terminal using a copper wire communication cable at distances up to
300 m (1000 ft).
1.14. Additional Accessories
Additional accessories include:
InSiteSL PC Configuration Tool
Legacy, dual channel fiber optic converter (safe area) for RS-232 or 20mA CL
Glass core or plastic fiber optic cable
Various brackets for wall and column mounting of the harsh enclosure
1.14.1. InSite SL Configuration Tool
InSite™ SL is available to end users of IND560x terminals. The IND560x terminal can connect to a
PC running InSite™ SL via Ethernet or Serial to provide the following:
Saving configuration information locally on the PC
Loading a saved configuration file into other devices
Restoring to a known state for service purposes.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 1-21
1.14.2. InSite CSL Configuration Tool
InSite™ CSL is available to certified Mettler Toledo service representatives.
The IND560x terminal can connect to a PC running InSite™ CSL to provide the following:
Viewing and/or changing configuration.
Enabling device-free configuration work before hardware installation.
Saving configuration information locally on the PC, loading a saved configuration file into other
devices, or restoring to a known state for service purposes.
WYSIWYG print template editing tool with expanded viewing area, cut/paste functions, stored
clipboard library (MyData items), and template space usage display.
Printing documentation of configuration for users’ records.
Performing firmware upgrade services for the IND560x.
1.14.3. Legacy Fiber Optic Converters
The 0964-0043, -0058, -0059 and -0060 dual channel fiber optic converters can be used with
the IND560x terminal to provide one or two additional channels of bidirectional communication at
distances up to 300m (1000 ft). The converters provide either RS-232 or 20mA current loop
connectivity in the safe area, derived from the COM4 and COM5 ports of the terminal.
If the ACM500 is already connected to the IND560x terminal, it is not possible to connect the dual
channel fiber optic converter. This is because both devices require dedicated use of the COM4 and
COM5 ports of the IND560x terminal.
The IND560x and dual channel fiber optic converters are designed to use a hard clad silica fiber
optic cable, but plastic core fiber optic cable can be used on short cable runs of 40 m (130 ft) or
less.
1.15. Display and Keypad
The IND560x terminal has a Liquid Crystal Display (LCD), 128 × 64 dot matrix graphic type with a
white LED backlight. Figure 1-9 shows the IND560x’s display and keyboard layout.
Figure 1-9: The IND560x Display and Keyboard Layout
Scale function
keys
Numeric,
clear and
decimal keys
Navigation
and enter
keys
Power key
Display:
System line
Display:
Weight area
Display:
Softkey icons
Softkeys

1-22 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Introduction
The display provides a system line at the top of the display for system messages and data;
application area with weight display, legends, data entry, and other information in the middle; and
softkey labels (icons) across the bottom.
The system line shows system messages and any asynchronous errors. The middle portion of the
display is reserved for the weight display and/or SmartTrac display. Random data entry (e.g. tare or
target ID) is shown in the bottom of this area. The bottom of the display is reserved for showing the
graphic labels (icons) for the softkeys. Display positions for up to five softkey icons are provided.
To the right of the softkey icon area is a reserved space for a MORE UP ( ) or a MORE DOWN ( )
indicator. If present, these indicate additional softkey selections are available by pressing either the
UP or DOWN navigation keys. A total of 15 softkeys may be programmed, depending upon the
weighing options and terminal functions enabled. These are presented in three sets of five softkeys.
The softkey setup and key mapping capabilities of the terminal determine the positioning of the
softkeys and locations where they display.
Three dedicated scale function keys are located to the right of the display. These are used to zero or
tare the scale and to initiate a print.
The terminal’s 12-key numeric keypad is used to enter data and commands. The numeric keys
include a Clear key and a decimal point, and are located on the upper-right side of the terminal
front panel.
Five navigation keys are located below the three scale function keys. These keys enable the
operator to navigate through setup options in the menu tree and within setup and application
screens.
An On/Off power key is located at the lower right of the front panel. The On/Off power key is enabled
and disabled by a jumper on the main board.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 2-1
This chapter covers
 Overview
 Display Operation
 Understanding the Navigational Interface
 Home Screen
 Basic Functionality
2 Operation
2.1. Overview
This chapter provides information about navigation, basic functionality,
and report generation using the IND560x terminal.
Operation of the terminal depends on enabled functions and parameters
that are configured in setup as described in Chapter 3, Configuration.
Navigation and basic functionality, as well as basic report generation are
covered in the following sections.
2.2. Security
The IND560x supports the use of username/password for setup security at four levels. Administrator
An Administrator has unlimited access to all areas of the operating and setup system. There can be
multiple Administrator accounts. There is a Primary Administrator account, which can be changed
but never deleted. The terminal is pre-configured at the factory with the Primary Administrator
account with a username of “admin”. The factory default password is null (no password). The preconfigured username (admin) cannot be changed; the password can only be added or modified.
The unit as configured at the factory requires no login or password entry to enter the setup mode. All
functions of the terminal are available to all users until a password for the Primary Administrator
account is set up.
When the Metrology switch (SW1-1) is turned “on” (refer to the next section of this chapter) and a
region approval is selected in setup, all users with Administrator rights are reduced to the
Maintenance level. This is done to protect metrologically significant parameters that cannot be
changed when the terminal is approved.
Similarly, the metrology switch can also be used to prevent Administrator level users from
accessing metrology features even if the region approval is “None”.
Once a password is set up, be sure to remember the password. If the password is changed or
forgotten, access to the setup menu will not be available. Be sure to protect the password from
access by unauthorized personnel. The password provides access to the entire setup menu,
unless the metrology switch is placed in the approved position.
2.2.1. Maintenance
Access can be limited by configuring this security level as desired.

2-2 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Operation
2.2.2. Supervisor
Access can be limited by configuring this security level as desired.
2.2.3. Operator
One default operator account with a username of anonymous and password null (no password) is
provided. Sites with validation requirements might create many operator accounts, each with a
username and password entry requirement.
2.2.4. Password
If a password has been programmed for the Primary Administrator username in Setup, a security
challenge is presented when entering Setup and a username and password must be entered. The
user can enter a password at any security level supported by the User configuration in the User
setup.
The terminal is pre-configured at the factory to include an “anonymous” user account with a null
(blank) password. The default anonymous user record can be edited or deleted. The user can
increase to a higher level of security clearance by requesting a setup function and then meeting the
associated security challenge.
If a login fails, the display exits the login page and returns to the home position.
2.2.5. Metrology Switch
If the metrology switch (SW1-1) is placed in the approved position (On) access to the Scale branch
of setup and other metrologically significant areas is not permitted. The metrology switch can be
used to prevent Administrator level users from accessing metrology features even if the region
approval is “None”.
Selecting a specific region approval and turning “on” the Metrology switch specifically alters some
scale functions:
 If the approval is Canada, then the keyboard tare command carries out a rounded tare.
 If the approval is Canada, then the center of zero division is 0.2d. The default value for nonapproved mode is 0.25d.
 If the approval is OIML, then the power up zero range is +/- 10% and pushbutton zero range is
+/-2%. For non-approved mode, this range can be selected by the user.
 If the approval is Argentina, terminal functions are changed as follows:
 Access to the setup screens is prohibited while the metrology switch is engaged. If an
attempt is made to access the setup menu, a message is displayed that says “Access
denied. Scale is Approved.”
 All shared data server write ability via COM1 (serial) and Ethernet is disabled. Information
can still be read from shared data, but writing to shared data is not possible.
 The FTP server (both serial and Ethernet) can only read files from the terminal. It cannot
write anything to the terminal. Shared data field nf0101 will automatically be set to 2.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 2-3
 If SICS is selected as an assignment for a port, all SICS commands must be disabled.
When a SICS command is received, an “EL” response is returned.
 A preset tare from the front panel, from the tare table, received serially (for example 1.00T
via the CTPZ input assignment) or received from a PLC, can only be accepted if there is no
motion on the scale. If motion is detected, the terminal should wait for the programmed
stability timeout period and, if the motion continues, discard the request. The preset tare
feature will therefore operate just like the semi-automatic tare feature.
Access to the metrology switch may be sealed in conformity with local regulations in “legal-fortrade” applications. Figure 2-1 shows the location of the metrology switch.
Figure 2-1: Metrology Switch Location, Panel (left) and Harsh (right)
Refer to the Installation Manual, PCB Switch Settings for further information about SW1-1
and SW1-2 switch settings.
2.3. Display Operation
Key names and commands are identified in this manual by upper- and lower-case letters. Key
names, such as ENTER, are in all upper-case letters, and commands, such as “select,” are in
lower-case (unless they begin a sentence, in which case the first initial is upper-case). For
example:
“Press START...” means to press the START softkey .
“Select an option...” means to use the UP or DOWN navigation keys to select a setting, then
press ENTER.
2.3.1. Softkeys and Icons
Softkeys and application displays use graphic icons for identification. Table 2-1 shows graphic
icons and their functions, categorized by where they are used. Items with an asterisk (*) require a
supervisor level log-in or above to be in effect.
SW1

2-4 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Operation
Table 2-1: Icons and Functions
Icon Function Icon Function
Information Recall Options
Clear All* Recall information
Clear Subtotal* System Information Recall
Metrology Recall Total Recall
Print* Weight Recall
Softkey Menu and Setup
Enter Setup Menu More Softkey Selections (scroll
rows up and down)
Calibration Icons
CalFREE Calibration Capture Zero
Capture Span Step Calibration
Service Mode (access
the IDNet service mode) Start
Run Calibration Test Test Weight Information
Skip (skips failed
Calibration Test step
and continues with the
test)
Table, Memory and Function Access
Tare Memory (access
the tare table)
Target Memory (access the
target table)
Alibi Memory
Target (view or edit active
record values for the target
comparison)
View Table (access a
table to view or
retrieve a record)
Search (begin the search of
a table for a record or
group of records)
Tare (captures live
scale weight into the
tare record – Tare
Table edit only)
Comparators (view and edit
Comparator Limit values)

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 2-5
Icon Function Icon Function
Reports (recall and
print reports for Alibi
Memory, Tare Table,
or Target Table)*
ID (initiates a programmed
transactional sequence)
Transaction Counter Reset Counter
a
Time & Date Print
Repeat Print Customized Print Triggers
1, 2 and 3
Editing
Clear All* Exit (return to previous
screen)
Delete Insert
Edit Validate Entry/Transaction
(OK)
Escape (exit without
storing) Copy
Display Icons
Times 10 Display
(expands the displayed
weight by 10)
SmartTrac (turns SmartTrac
display off and on)
MinWeigh (enter
MinWeigh mode) Unit Switching
Adjust Contrast
Special Control
Target Control Start
Pause Stop
No Clear Total*
Reset

2-6 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Operation
Task Selection Icons (only when TaskExpert™ is installed)
Icon Function
Task List
Displays list of assigned TaskExpert applications
Task 1
Starts TaskExpert application designated as Task 1
Task 2
Starts TaskExpert application designated as Task 2
Task 3
Starts TaskExpert application designated as Task 3
2.4. Understanding the Navigational Interface
Navigate in the applications and configure the IND560x using
 Navigation keys  Alpha keys
 Softkeys  Numeric keys
 Scale function keys
The locations of these keys and the display screen area are shown in Figure 2-2.
Figure 2-2: Key Locations and Display Screen
2.4.1. Navigation Keys
Navigation keys (see Figure 2-2) enable navigation within the setup menu tree, setup screens, and
application screens. Navigation keys include:
 UP and DOWN navigation keys
Moves the focus up or down to different setup options within the menu tree (focus is indicated
by highlighted text), to different fields within a setup page, and to switch to another page of
available softkeys. The UP and DOWN navigation keys are also used to move to a next or
Softkey labels
System line
Softkeys
Weight and
Application area
Numeric keys
Navigation keys
Zero, Tare, and
Print keys
Power button

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 2-7
previous page when a scroll bar indicates that two or three pages are used for configuration
fields associated with a specific parameter. The example in Figure 2-3 shows a configuration
parameter that requires two pages to display all of the associated configuration fields. The
scroll bars indicate which page of the two is displayed.
Figure 2-3: Example of Scroll Bars, Indicating Two Pages of Configuration Fields
Press the DOWN navigation key to move from page 1 to 2, Press the UP navigation key to
move from page 2 to 1.
 LEFT and RIGHT navigation keys
Expand (RIGHT navigation key) the setup options in the menu tree
Collapse (LEFT navigation key) the setup options in the menu tree
Move the cursor position to a specific character in text areas
Enable left and right scrolling to view all information available on a screen
 ENTER key
Opens the setup page for viewing and editing setup parameters
Moves the focus from a field label to a setup value for that field
Accepts new values entered in a field and moves the focus to the next field label
2.4.2. Softkeys
There is a softkey setup page that is used to
 Change softkey positions
 Enable softkey functions
 Disable softkey functions
For example, a UNIT SWITCHING softkey can be enabled for selecting alternate units of
measure.
Five softkeys are located along the bottom of the display screen (see Figure 2-2). Some
applications might have up to three pages of softkeys for a total of 15 possible functions. A MORE
DOWN symbol displayed on the lower-right corner of the screen (to the far right of the softkey
icons) indicates that more softkey selections are available. Press the DOWN navigation key to
display additional softkey screens, and UP to display the previous softkey screen.
2.4.2.1. Customizing Softkeys
All the softkeys available for assignment to the home screen can be replaced with customized
bitmap images.
Scroll bar
indicating
page 1 of 2
Scroll bar
indicating
page 2 of 2

2-8 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Operation
The bitmaps must measure no more than 23 pixels wide by 15 pixels high. Each graphic must
have a white border at least one pixel wide at left, right and top – see Figure 2-4.
Figure 2-4: Softkey Graphic Dimensions
Graphics may be uploaded to the IND560x terminal using an ftp connection via Ethernet, or
through a serial port.
New bitmap graphics must have the same filename as the graphic they replace. Table 2-2 lists the
softkey graphics that can be replaced with customized images.
Table 2-2: Bitmap Graphic Filenames
Softkey Default Image Graphic File Name
Alibi alibi.bmp
Calibration Test cal_test.bmp
Comparators comprtr.bmp
ID Id.bmp
MinWeigh minweigh.bmp
Recall Info recall.bmp
Reports reports.bmp
Setup setup.bmp
SmartTrac sm_trac.bmp
Tare Table tare_mem.bmp
Target target.bmp
Target Control control.bmp
Target Start start.bmp
Target Table targ_mem.bmp

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 2-9
Softkey Default Image Graphic File Name
Trigger 1 trigger1.bmp
Trigger 2 trigger2.bmp
Trigger 3 trigger3.bmp
Time & Date timedate.bmp
Unit Switching select.bmp
X10 Display x10.bmp
2.4.3. Scale Function Keys
Scale function keys (see Figure 2-2) are:
ZERO When the scale platform or weighbridge is empty, the terminal should
indicate zero. The gross zero reference is recorded during calibration. Press
the ZERO scale function key to capture a new gross zero reference point if
pushbutton zero is enabled in configuration and the weight is within the
zero range.
TARE Tare is the weight of an empty container. Tare is normally used to determine
the net weight of the contents of a container. Press the TARE scale function
key when an empty container is on the scale. The terminal then displays a
zero net weight. As the container is loaded, the terminal then displays the
net weight of the contents. Pushbutton tare must be enabled to use this key
in this manner.
When the empty weight of the container is a known value, enter the tare
weight using the numeric keys and then press the TARE scale function key.
The terminal will display the net weight of the contents of the container.
Keyboard tare must be enabled in order to use this key in this manner.
PRINT Press the PRINT scale function key to initiate a demand print of an assigned
print template. A serial or Ethernet print connection must be configured to
link a template or report to the selected serial or Ethernet port. When a print
command is issued, Printing appears in the system line for 5 seconds –
see Figure 2-9.

2-10 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Operation
2.4.4. Alpha Keys
On some setup pages, softkeys function as alpha keys (see Figure 2-5) that are used to enter
alphabetic characters for setup parameters such as passwords.
Figure 2-5: Alpha Keys
The alpha keys are automatically displayed when the cursor is moved into a data entry box that
supports alpha entry. The first set of softkeys shown is upper-case letters (Figure 2-5). Three
additional sets of softkeys include lower-case letters and symbols. In each of these sets, a group of
four to six characters is shown per softkey.
Figure 2-6: Additional Character Sets
To enter a character, use the UP and DOWN arrow keys to move to the row of keys that shows the
group including the desired character. If the set of keys shown does not include the required
character, press the DOWN navigation key to view the next set. Press the DOWN navigation key
and UP navigation key as required until the set of keys shows the required character. The presence
of the MORE UP / MORE DOWN symbols to the right of the softkeys indicates that additional sets of
characters are available. Once the correct row is showing, press the softkey under the group of
characters that includes the required character.
Character sets
above associated
softkeys

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 2-11
At this point, the characters above the softkey will change to show one character per softkey (Figure
2-7).
Figure 2-7: One Letter Per Softkey
Now, press the softkey that corresponds to the desired character. Also note that the MORE DOWN
symbol is shown at the right of the softkeys. Press the DOWN navigation key to view the second
row of characters. This group will include an ESCAPE softkey (Figure 2-8). If the wrong group
of characters was selected, use this softkey to return to the character set selection screen.
Figure 2-8: ESC Character
Repeat this process until all alpha characters have been entered. The CLEAR key can be used to
delete unwanted characters in the entry box. The ESCAPE softkey on the symbol group of
softkeys (see Figure 2-6) exits the data entry process without accepting any of the new data typed.
The data in the entry field will revert to the original data (that was in the entry box before the new
entry was initiated).
Once desired alpha characters have been entered in appropriate fields on the screen, press the
ENTER key to accept the entries.
2.4.5. Numeric Keys
Use the terminal’s 12-key numeric keypad (see Figure 2-2) to enter data and commands.
To use numeric keys, position the cursor in the field (see Navigation Keys) and press the numeric
keys to enter the appropriate data. Press the DECIMAL key to enter decimal points where necessary.
The CLEAR key functions like a backspace key. Position the cursor at the end of data to be deleted
and press the CLEAR key. Press the CLEAR key once for each character to be deleted. When first
entering a data entry box, the previous entry will be in focus. Pressing CLEAR at this point will clear
the entire previous entry.

2-12 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Operation
2.5. Home Screen
The Home screen displays when the terminal is idle. The Home screen is the only screen that
provides operator access to the programmable softkeys. Figure 2-9 shows a sample screen with a
message showing in the system line.
Figure 2-9: Default Home Screen, System Line with Message Showing
The Home Screen includes:
 System Line Displays system messages and application data
 Application Area Displays weight, units, and other application-specific weighing data
 Softkey Icons Displays the icons for the active softkey functions. The appearance of a
MORE DOWN symbol or a MORE UP symbol indicates that more
softkey selections are available
2.6. Basic Functionality
This section provides information about the IND560x’s basic functionality. Additional areas of
functionality specific to application software available for the IND560x are addressed in the specific
application manuals. Basic functions addressed in this section include:
 External Battery
Compatibility
 MinWeigh  Reports
 Information Recall  Calibration Test
 Zero  Target Comparison  CalFREE™
 Tare  Comparators  Alibi Memory
 Unit Switching  ID Mode  Table Searches
 Expand By 10  SmartTrac  Remote Display
 Print  Time and Date  ACM500 Reconnect
2.6.1. Compatibility with External Battery Pack
An external NiMH battery pack is available as a power option for the harsh/analog model of the
IND560x terminal. Only the METTLER TOLEDO NiMH battery pack, part # 64060625, is approved
for use with the IND560x. The IND560x terminal requires a specific software configuration to enable
Application
area
MORE DOWN
symbol
indicating
more Softkey
selections
Softkey
icons
System
line

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 2-13
system compatibility with NiMH battery part # 64060625. Additional software parameters exist to
help extend the single charge life of the battery.
These battery-specific settings are found in the following locations within the terminal setup menu.
The settings can be addressed in the following order to ensure proper setup for external battery
compatibility.
2.6.2. Scale > Type
Load Cell parameters are shown only for IND560x analog main boards and only if scale type is
selected as Analog. The Load Cell parameter default values are 1 x 350 Ω. If scale type is set as
Remote, the number of cells is automatically set to 0.
These settings are shown in the Setup menu for all analog terminals whether support for the
battery has been enabled or not. However, only the harsh/analog model is physically capable
of supporting the external NiMH battery.
It is important to enter the correct values for the Load Cell parameters, as the current demand from
the load cells determines the switch off voltage set point of the external battery. Setting the Load Cell
parameters correctly helps to:
 Ensure that all the available capacity of the battery is being utilized
 Protect the battery pack from a deep discharge that will harm the NiMH cells and shorten the
lifespan of the pack
2.6.2.1. Terminal > Device > Battery Operation
This parameter is only shown in IND560x models with analog main boards. The default value is
Disabled. Setting this parameter to Enabled will trigger the IND560x to use the Load Cell
parameters established at Scale > Type. Enabling Battery Operation also unlocks an Auto Power Off
parameter elsewhere in the Terminal branch of the Setup menu.
2.6.2.2. Terminal > Display
Once the Battery Operation parameter has been set to Enabled, users can choose to activate an
auto-power off feature that will help to optimize single charge battery life. The Auto Off Timer
parameter monitors scale motion and keypad operation. If no keypad activity occurs within the
selected time value, the terminal automatically turns itself off. This auto-shutdown will occur with no
warning to the operator.
2.6.3. Recommendations for Battery Powered Systems
Battery powered systems of all types should always avoid unnecessary power consumption. The
following recommendations should be used whenever possible to prolong the single charge life and
long term performance of the METTLER TOLEDO NiMH battery pack (P/N 64060625).
2.6.3.1. IND560x System Hardware Considerations
 Use high impedance scale bases and load cells when possible. When considering only NiMH
battery single charge life, 1,000 Ohm load cells are preferred over 350 Ohm load cells. Single
charge battery life will increase as load cell impedance increases.

2-14 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Operation
 Only install options and connections that are required. If an option board has multiple
connections, only physically connect the ports that are required. (e.g. if only COM4 is required,
do not make a physical connection to COM5 also).
2.6.3.2. IND560x Software Considerations
 Enable Battery Operation at Terminal > Device > Battery Operation
 If feasible, enable the Auto Power Off setting at Terminal > Device > Auto Power Off
 If backlight is enabled, enable the Backlight Timeout setting found at Terminal > Device >
Display.
 Disable all unnecessary communications that might be configured at Communication >
Connections. This includes communications on COM4, COM5 or COM1.
2.6.3.3. General NiMH Battery Handling
 Recharge the NiMH battery as soon as possible after a “low battery” warning is seen on the
screen of the IND560x terminal.
 Rechargeable batteries self-discharge. To prevent the NiMH battery packs from experiencing
a damaging deep discharge, establish a quality control process to recharge them every
three months, even if not in use.
 Do not store or charge the NiMH battery packs in hot places. High temperatures increase
the rate of self-discharge.
2.6.3.4. Estimated Single Charge Battery Life
When users implement the recommendations listed above regarding IND560x hardware and
software, as well as the general NiMH battery handling recommendations, the estimated single
charge life of Mettler Toledo’s NiMH battery pack (P/N 64060625) in various IND560x applications
is shown in Table 2-3.
All estimates assume use of a new, freshly charged NiMH battery pack (NiMH batteries
experience up to 5-10% self-discharge in first 24 hours following charge.)
All estimates assume use at 20ºC
Approximate recharge time 6 is hours
500 recharge cycles is lifetime expectancy of NiMH battery pack P/N 64060625. Reduced
runtime will occur at the end of battery lifetime.
Table 2-3: Estimated NiMH Battery Single Charge Life
Total Load Cell
Resistance (TSR)
Installed IND560x
Hardware Options
Approximate
Current
Consumption
Estimated
Operating
Time
87.5
No options
OR
A/P Discrete I/O Only*
265 mA 25 hours
350 225 mA 32 hours
1000 215mA 35 hours
87.5 CL or FO with or without 275mA 24 hours

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 2-15
Total Load Cell
Resistance (TSR)
Installed IND560x
Hardware Options
Approximate
Current
Consumption
Estimated
Operating
Time
350 A/P Discrete I/O* 240mA 30 hours
1000 230mA 32 hours
* The influence of the A/P Discrete I/O on overall operation time is negligible.
2.6.3.5. Low Battery Warnings and Battery Protection
When the battery voltage reaches a pre-defined value, a “Low Battery!” message will appear in
IND560x system information line, blinking at three second intervals. The terminal will continue to
operate for one hour or less (depending on system configuration and battery condition) after the
“Low Battery!” message appears. During this period, turn the terminal off and either remove and
charge its battery, or replace the battery with a fully-charged one.
Note: Performance of the NiMH battery will be reduced if it is allowed to discharge too deeply.
The low battery warning and shut-off functions are designed to protect the battery’s efficiency
and life-span.
If the shut-off voltage is reached before the battery is recharged or replaced, a message will appear
on screen – “ALARM – Low Battery! Power off is in process!”. This message will display for ten
seconds, after which the terminal will power off automatically.
2.6.4. Zero
The Zero function is used to set or reset the initial zero reference point of the IND560x. There are
three types of zero setting modes:
 Automatic Zero Maintenance
 Power-Up
 Pushbutton
2.6.4.1. Automatic Zero Maintenance
Automatic Zero Maintenance (AZM) enables the IND560x to compensate for the buildup of small
amounts of weight and track itself back to the center of zero. Within the AZM operating range
(programmable from 0.00 to 10.00 divisions), when the terminal is in a no motion condition, it
makes small adjustments to the current zero reading to drive the weight reading toward the true
center-of-zero. When the weight is outside of the programmed AZM range, this feature is not
functional.
2.6.4.2. Power-Up Zero
Power-Up Zero enables the IND560x terminal to capture a new zero reference point after power is
applied. If there is motion during a power-up zero capture function, the terminal will continue to
check for a no-motion condition until zero is captured.
Power-up zero can be disabled or enabled, and a range above and below calibrated zero can be
configured. The range is programmable from 0% to 100% of capacity and can include a positive
range and also a range below calibrated zero.

2-16 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Operation
2.6.4.3. Pushbutton Zero
The pushbutton (semi-automatic) zero function can be accomplished by pressing the ZERO
scale function key, programming a discrete input, a PLC command or serial command, or by an
application.
The range for all types of semi-automatic zero is selectable (0% to 100%) plus or minus from
either the calibrated zero point (if power-up zero is disabled) or from the initial zero setting point (if
power-up zero is enabled).
Remote initiation of the semi-automatic Zero command is possible via a discrete input, an ASCII ‘Z’
command sent serially (CPTZ and SICS), a command initiated by the PLC interface, or from an
application.
2.6.5. Tare
Tare is the weight of an empty container. A tare value subtracts from the gross weight
measurement, providing the computation of the net weight (material without the container). The tare
function can also be used to track the net amount of material being added to or removed from a
vessel or container. In this second case, the weight of the material in the container is included with
the tare weight of the container as tare. The display then reflects the net amount being added to or
removed from the vessel.
Tare types and associated operations available on the IND560x include:
 Pushbutton Tare  Tare Clear
 Keyboard Tare (Preset Tare) — Manual Clear
 Net Sign Correction — Auto Clear
 Auto Tare
2.6.5.1. Pushbutton Tare
Pushbutton tare can be configured in setup as enabled or disabled. When disabled, the TARE scale
function key cannot be used to obtain a tare.
If enabled, pressing the pushbutton TARE scale function key initiates a semi-automatic tare.
The IND560x will attempt to perform a tare process. If successful, the display changes to a zero net
weight indication and the previous weight on the scale is stored as the tare value. The net mode will
be indicated on the display.
Several conditions could inhibit the pushbutton tare function:
 Motion – Pushbutton tare cannot be taken when the scale is in motion. If motion is detected
when a pushbutton tare command is received, the IND560x will wait up to three seconds
for a no-motion condition. If a stable (no motion) weight condition occurs before the three
seconds expire, the pushbutton tare command is executed.
If there is still motion at the end of three seconds, the command is aborted and a “Tare
Failure–In Motion” error displays.
 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 2-17
 Pushbutton Tare Disabled – If pushbutton tare is configured as disabled, the TARE scale
function key will not initiate a semi-automatic tare.
 Negative Gross Weight – Any pushbutton tare attempted when the gross weight is at or
below zero is ignored and a “Tare Failed–Too Small” error displays. Ensure that the gross
weight is above zero.
2.6.5.2. Keyboard Tare
A keyboard (preset) tare is a numeric tare that is entered manually through the numeric keypad,
received serially from a peripheral, or retrieved from the Tare Table memory. The preset tare value
cannot exceed the capacity of the scale. Data entered is interpreted to have the same units as the
current displayed value. Motion does not impact the entry of preset tare values.
Keyboard tare can be configured in setup as enabled or disabled. When disabled, the numeric
keypad and the TARE scale function key cannot be used to obtain a tare.
To enter a preset tare value manually, use the numeric keypad to enter the tare value (the data
entered will display above the softkey with a Data: label) and press the TARE scale function key
.
If configured in setup, remote equipment can enter a preset tare value using a serial command or
PLC command (Refer to Chapter 3, Configuration, the Communication section, Serial and PLC
setup, for further information).
If the preset tare is successful, the display changes to a net weight indication, and the entered
preset tare value is stored as the tare value in the Tare Table.
Several conditions could inhibit the preset tare function:
 Keyboard Tare Disabled – If keyboard tare is configured in setup as disabled, the numeric
keypad and the TARE scale function key cannot be used to obtain a tare.
 Over-Capacity or Under-Zero Conditions – Preset tare is not allowed when the weight
display indicates over capacity or under zero conditions. Any preset tare attempted when
the scale is over capacity is ignored and a “Tare Failed–Over Cap” error displays. Any
preset tare attempted when the weight display indicates an under zero condition is ignored
and a “Tare Failed–Too Small” error displays.
Preset tare can be entered in free format. If the entered value does not match the displayed weight
decimal point location or display interval, the entered tare value is rounded to the nearest display
interval and the decimal point adjusted to match the gross weight. The rounding method is that 0.5
or more of a display interval (d) is increased to the next display interval and 0.49 or less of a
display interval is decreased to the next lower display interval.
When entering a preset tare value less than 1.0, the operator can enter the data without the leading
zero (left of the decimal point), but all subsequent display, storage, or printing of this value will
include the leading zero. For example, a preset tare entry of .05 will display as 0.05.
If a preset tare has already been established and another preset tare is entered, the second preset
tare replaces the previous value (it does not add to the previous value). The replacement tare can
be larger or smaller than the original tare value.

2-18 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Operation
2.6.5.3. Tare Table
The IND560x terminal contains a tare table with 25 records for storing tare weights that can be
recalled by the operator, instead of manually entering them for each transaction. This is especially
useful when certain tare values are used repeatedly.
A 20 character description can be included for each record. This can be used to help distinguish
one tare record from another. Each tare record in the tare table also contains a totalization field.
When totalization is enabled for the tare table, each time a transaction is completed using a specific
tare ID, the selected weight value (gross or net weight) will be added to the total and the
corresponding counter will be incremented by one.
A tare memory can be recalled by either picking from a list of all available records or if the ID for the
tare value is know, it can be “quick” accessed (refer to the Quick Access section, below). A printed
report of the records in the Tare Table is also available.
2.6.5.4. Net Sign Correction
Net sign correction enables the terminal to be used for both shipping (inbound empty) and
receiving (inbound loaded) operations. Net sign correction can be either disabled or enabled on the
IND560x. Refer to Chapter 3, Configuration, the Scale section, for further information about
disabling and enabling net sign correction.
If net sign correction is disabled in setup, any stored weight value in the tare register is assumed to
be a tare regardless of the gross weight present on the scale at the time of the final transaction and
net values can be negative.
If net sign correction is enabled, the terminal will switch the gross weight and tare weight fields
when necessary, so that the larger weight is the gross weight, the smaller weight is the tare weight,
and the difference is always a positive net weight. Net sign correction affects the display, stored
data, weight recall, and printed data.
Net sign correction will operate with pushbutton tare, preset tare, or tare memories. An example of
weight values with and without net sign correction is shown in Table 2-4. In this example, the tare
register value is 53 kg and the live weight on the scale is 16 kg.
Table 2-4: Weight Values With and Without Net Sign Correction
Net Sign Correction
Printed and Displayed Disabled Enabled
Gross 16 kg 53 kg
Tare 53 kg 16 kg
Net 37 kg 37 kg
 When net sign correction is enabled, the tare weight field in the recall display will be
labeled with the letter “M” to indicate “Memory” instead of “T” or “PT”.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 2-19
2.6.5.5. Auto Tare
The IND560x can be configured so that tare is automatically taken (auto tare) after the weight on
the scale exceeds a programmed tare threshold weight. Auto tare can be configured in setup as
enabled or disabled. When auto tare is enabled, the display changes to a zero net weight indication
after the weight exceeds the threshold value. The previous weight on the scale is stored in the Alibi
Table as the tare value. Auto tare operations involve:
 Tare Threshold Weight – When weight on the scale platform exceeds the tare threshold
value, the terminal automatically tares.
 Reset Threshold Weight – The reset threshold weight must be less than the tare threshold
weight. When the weight on the scale platform falls below the reset threshold value, such
as when a load has been removed, the terminal automatically resets the auto tare trigger.
 Motion Check – A motion check is provided to control the re-arming of the auto tare
function. If disabled, the auto tare trigger will be reset as soon as the weight falls below the
reset value. If enabled, the weight must settle to no-motion below the reset threshold before
the next auto tare can be initiated.
Several conditions could inhibit the auto tare function:
 Motion – Auto tare cannot be taken when the scale is in motion. If motion is detected after
the weight on the scale exceeds a preset tare threshold weight, the IND560x will wait for a
no-motion condition. If a stable (no motion) weight condition occurs within three seconds,
the auto tare command is executed.
 Auto Tare Disabled – Auto tare can be configured in setup as enabled or disabled.
2.6.5.6. Clearing Tare
Tare values can be cleared manually or automatically.
2.6.5.6.1. Manual Clear
Manually clear tare values by pressing the CLEAR key on the numeric keypad when the
IND560x is in the net mode and has completed the weighing operation. Motion on the scale will
not impact a manual clear.
If configured in setup, pressing the ZERO scale function key will first clear the tare, then issue
a zero command (Refer to Chapter 3, Configuration, the Scale section, Auto Clear).
2.6.5.6.2. Auto Clear
The IND560x can be configured to clear tare automatically when the weight returns to a value
below a programmable threshold, or when a print command is issued. Once the tare is cleared, the
display returns to the gross weighing mode.
Auto clear is disabled or enabled in setup. If auto clear is enabled the following parameters,
configured in setup, affect the auto clear operation:
 Clear Threshold Weight – The clear threshold weight is the gross weight value below
which the IND560x will automatically clear a tare after settling to a value above this
threshold value.

2-20 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Operation
 Motion Check – A motion check is provided to control the automatic clearing of tare. If the
motion check is disabled, the tare value is cleared as soon as the weight drops below the
threshold weight (auto clear threshold), regardless of the motion status.
If the motion check is enabled, after meeting the requirements for weight value above and
then below the threshold weight (auto clear threshold), the IND560x waits for a no-motion
condition before automatically clearing the tare.
 Clear After Print – If enabled, tare is automatically cleared and the scale returned to the
gross mode after data has been transmitted by pressing the PRINT scale function key
or from a remote source.
Refer to Chapter 3, Configuration, the Scale section, Auto Clear for further information about
configuring auto clear.
2.6.6. Unit Switching
In order to support locations and applications that utilize multiple measurement units, the IND560x
supports unit switching. The UNIT SWITCHING softkey enables switching between primary units
(the main unit of measure) and alternate units (second or third units).
When the UNIT SWITCHING softkey is pressed, the display changes from the primary unit to the
second unit. When this softkey is pressed again, the display switches to the third unit (if one has
been programmed) or returns to the primary unit. Each subsequent time the UNIT SWITCHING
softkey is pressed the IND560x continues to switch the unit through the same sequence.
When units are switched, the units value changes to the correspondingly selected units, and the
conversion of the display value occurs. The display division changes to an equivalent weight value
in the switched unit (for example, from 0.02 lb to 0.01 kg) and the decimal location switches to
accommodate the conversion.
When switching units, the capacity of the converted units is dictated by the original number of
divisions established in the capacity and increments area of setup. In some situations, this may
reduce the capacity of the terminal when converting to second or third units.
2.6.7. Expand By 10
The EXPAND BY 10 softkey is used to increase the selected weight display resolution by one
additional digit. For example, a weight display of 40.96 could increase by one additional digit to
display as 40.958. The Expand By Ten mode is indicated on the display by x10, which displays
below the weight display. When the EXPAND BY TEN softkey is pressed again, the display
reverts to the normal display of weight.
If programmed as approved with the metrology switch (SW1-1) ON, the Expand By Ten mode is
displayed for five seconds then automatically returns to normal resolution. Printing is disabled
when the weight is expanded and the terminal is programmed as approved.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 2-21
2.6.8. Print
The print function (demand print) can be initiated by pressing the PRINT function key on the
main keyboard, by using custom print trigger softkeys or by automatic print settings. Demand
printing of data may also be initiated as part of a particular sequence of operation or special
application software. A ‘Printing’ system message (visible in Figure 2-9) appears for 5 seconds
when the terminal is carrying out a print command.
2.6.8.1. Print Interlock
Print Interlock is designed to enforce a single print output per transaction. Print Interlock can be
disabled or enabled. If enabled, the print command is ignored until the measured gross weight
exceeds the print interlock threshold. After the first print command is executed, subsequent print
commands are ignored until the gross weight indication falls below the print interlock reset
threshold. If a print command is blocked by print Interlock, a synchronous “2nd Print Prohibited”
error is generated and a “Print Not Ready” message appears in the system line.
2.6.8.2. Automatic Print
Automatic initiation of a demand print occurs after the gross weight exceeds the minimum threshold
and there is no motion on the scale. After initiation, the gross weight must return below the reset
threshold before a subsequent automatic print can occur.
Automatic print may be disabled or enabled. Automatic print can be triggered and reset by weight
exceeding thresholds or by weight deviation from a previously stable reading.
2.6.8.3. Repeat Print
The Repeat Print softkey permits the output of the most recent Demand Print or a Custom Print
from an application, to be printed again with a DUPLICATE header or footer to distinguish it from the
original print.
To enable the Repeat Print function, simply add the Repeat Print softkey to the Home Page.
Pressing this softkey will initiate a repeat print of the last Demand Print connection listed in the
assignments found in Setup at Communication > Connections.
An output template can be flagged with a “DUPLICATE” header or footer to indicate that the data in
the output template was generated as a repeat of a previous communication.
2.6.8.4. Custom Print Trigger
The three custom print triggers available in the IND560x function like the standard
demand print, except that they will not trigger alibi logging, totalization or transaction counter
update.
NOTE: This is a change in functionality introduced in v4.03 firmware. It affects how the IND560
handles print templates containing metrologically sensitive data.
With firmware versions 3.04 and earlier, the intended use of custom print triggers was simply to
transmit non-metrological data from the terminal. Over the life of the IND560, it became apparent
that custom print triggers were being used to transmit metrologically sensitive information such as
weight data. However, in all firmware versions prior to v4.03, the custom print triggers were not

2-22 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Operation
subject to the same types of validity checks that the standard demand print was. Custom print
triggers have historically been able to transmit metrologically sensitive weight data even under such
circumstances as motion, over capacity, etc.
In terminals using v4.03 firmware, templates transmitted by a custom print trigger are evaluated for
the following shared data types that indicate if the template contains metrologically sensitive scale
information: ce, cs, tz, wk, wm, ws, wt. If the template includes any of these shared data fields, the
following statuses are checked:
 Power Up Zero Not Captured
 Over Capacity
 Under Zero
 Data not OK
 Times 10 (x10) is enabled and terminal is in legal-for-trade mode
 Minimum print threshold not reached between print requests
 Print Interlock condition not met
 Motion
In v.4.03 firmware, except for motion each of these causes an immediate error. If there is motion,
the terminal waits for the motion timeout period, and either prints (if stability is achieved) or
displays an error.
2.6.9. MinWeigh
Certain industries such as pharmaceutical and food require a guarantee that the weighing
equipment selected for a particular measurement is adequate for the task. One way to ensure that
appropriate weighing equipment is selected is by the creation and use of a minimum weighment
value (MinWeigh), below which a particular piece of weighing equipment cannot be used.
When the MinWeigh function is enabled, the MINWEIGH softkey can be pressed to initiate the
MinWeigh function, which compares the current net weight with the programmed MinWeigh value.
If the net weight is greater than or equal to MinWeigh, all equipment functions behave normally. If
the absolute value of the net weight is less than MinWeigh, the weight display includes a flashing
MinWeigh symbol to the left of the weight. If the user attempts to record the weight while in this
condition, the printout will include an asterisk ().
For specific step-by-step instructions on how to set up MinWeigh, Refer to Chapter 3, Configuration,
Scale, MinWeigh.
2.6.10. Information Recall
Press the INFORMATION RECALL softkey to access the recall screen where the softkeys shown
below appear:
Weight Recall Displays the Gross, Tare, and Net active weight readings for the scale.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 2-23
System Information Displays system information, including the model and serial number,
Terminal ID #1, Terminal ID #2, Terminal ID #3, and software and
hardware information. This ID information is entered in the Setup mode
(Refer to the Chapter 3, Configuration, Terminal, Device, for more
information about entering Terminal ID information).
Metrology Recall Displays the terminal metrology control number (MCN), whether the
terminal is programmed as approved or not, and the last calibration time
and date. On the high-precision IDNet version, the Ident Code (calibration
tracking) is also displayed.
Totals Recall Displays the sub-total and grand-total transaction count and total weight
for the terminal. Press the CLEAR SUBTOTAL softkey to clear the subtotal amounts. Press the CLEAR softkey to clear both the grand-total
and sub-total amounts. Press the PRINT softkey to print a report of
the totals.
Notes: If either Clear Total or Clear Subtotal After Print is enabled in setup,
these values will clear after printing.
Table clear and print functions are accessible only to users with
Supervisor level log-in and above.
2.6.11. Target Comparison
Target comparison is used to compare either the gross or net weight on the scale to a
predetermined target value. This feature can be useful in both automatic and manual processes. As
an example, an automatic filling system using target comparison could provide a Start signal to the
IND560x and the IND560x could control the feeder system to fill a container to a desired target.
An example of a manual process could be a checkweighing station where an operator is checking
net weight of packages. The IND560x can use its target comparison feature to provide a bright Over
- OK - Under indication to help the operator determine if the weight of each package is acceptable
or not compared to an ideal weight.
Target (setpoint) comparison is most often used in two types of applications:
Material transfer applications – Requires that a control device deactivate when a target value
is achieved
Over/Under applications – Classify a load placed on the scale platform as above or below the
target value or within the target tolerance range
The IND560x target comparison rate (for analog load cell scales) is 50 comparisons per second.
The comparison rate for high precision IDNet bases is dependent upon the module in the base.
Refer to Chapter 3, Configuration, the Application section, Memory, Target Table and Operation,
Target for further information about configuring parameters for target comparison operations.
2.6.11.1. Material Transfer Applications
If the target comparison of the IND560x will be used to control the flow of material, the application
can be classified as a Material Transfer application. These types of applications are usually
automated but can also be manual. A single or two-speed feeding system is used to either add

2-24 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Operation
weight to or remove weight from the scale. The terminal monitors the change in weight and
compares it to a previously entered target and other control parameters.
The following definitions explain some terms used in Material Transfer applications:
Abort After a target comparison process has been paused, it can be totally stopped by
pressing the Abort softkey under the Target Control softkey or by triggering
a Target Pause/Abort discrete input. If abort is selected, the target comparison
process is aborted.
Coincidence
Outputs
A coincidence output is always active and does not require a start or stop signal. If
the weight on the scale is below the target minus the spill value, the outputs are “on”.
If the weight is above the target minus the spill value, the outputs are “off”. This type
of output typically requires external logic to provide the required control for feeding
systems.
Concurrent
Outputs
This describes the operation of the feed output in a 2-speed feeding system. If the
feeder type is programmed as concurrent, the feed output will turn on when the fast
feed output turns on and continue to operate until the weight reaches the target minus
the spill value. Both the fast feed output and the feed output are active at the
beginning of a target comparison. It is the opposite of independent outputs.
Fast Feed Output This refers to the physical output connection that is used for the faster feed in a 2-
speed feeding system. This output is not used in a single speed feeding system.
Feed Output This refers to the physical output connection that is used for the slower feed in a 2-
speed feeding system or the only feed output in a single speed feeding system.
Fine Feed The value entered for the amount of material that will be fed in the slower rate of feed
in a 2-speed feeding system. This value and the spill value are subtracted from the
target to determine the point that the fast feed output turns off.
Independent
Outputs
This describes the operation of the feed output in a 2-speed feeding system. If the
feeder type is programmed as independent, the feed output does not turn on until the
fast feed output turns off. Only one feeder speed is active at a time. It is the opposite
of concurrent outputs.
Latched Outputs Latched outputs turn off at the target weight minus the spill value and remain off
(regardless of additional weight changes) until a “start” signal is received. These
types of outputs do not usually require external logic to perform standard weigh-in or
weigh-out sequences.
Material Transfer
Mode
The target comparison mode that provides control for delivering a measured amount
of material from one container or vessel to another. The transfer can apply to material
entering or leaving a container or vessel. It includes the traditional terms such as
weigh-in, weigh-out, filling, and dosing.
Pause A pause function is provided in target comparison control in case a target
comparison must be temporarily paused. This can be done by pressing the Pause
softkey under the Target Control softkey or by triggering a discrete input
programmed as Target Pause/Abort. When paused, power is removed from the Feed
and Fast Feed (if used) discrete outputs. After a target comparison has been paused,
the process can either be resumed or aborted.
Resume After a target comparison process has been paused, it may be resumed by pressing
the Resume softkey or by triggering a discrete input programmed as Target
Start/Resume. When a target comparison is resumed, it continues to use the original
target values.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 2-25
Spill The amount of material that will be added (on a weigh-in) or removed (on a weighout) from the scale after the final feed is turned off. In a weigh-in process, this is the
material in suspension that will still fall onto the scale when the feed is turned off.
This value is subtracted from the target value to determine when the feed output turns
off.
Start When using latched targets, the default condition is with the latch set or in the “off”
condition. To turn any outputs “on”, a start signal is required. This could be the Start
softkey , a discrete input programmed as Target Start/Resume, or via the Target
Control softkey .
Target The target is the weight value that is the end goal of the material transfer process. If a
container should be filled with 10 kg of material, the target value is 10 kg.
Tolerance The weight range above and below the target value that will be acceptable as an “in
tolerance” target comparison. The tolerance can either be entered as a weight
deviation from the target or a percentage deviation from the target depending upon
setup.
2.6.11.2. Target Control
Two different levels of target control are provided by the terminal for the Material Transfer mode.
Latching outputs must be enabled to utilize either type of automated control. For simple
applications, only the START softkey is needed to begin the target comparison process. No
other control is provided. The comparison runs to completion then the latch sets the outputs FALSE
until the START softkey is pressed the next time.
For those applications where more control is needed for an occasional pause or to abort a
comparison process, the TARGET CONTROL softkey is used. Pressing the TARGET CONTROL
softkey provides a dedicated display for the target comparison with the comparison status and
appropriate softkeys shown. Possible status modes are “Ready”, “Running” or “Paused”. The
appropriate control softkeys include START , PAUSE and STOP .An example of each
target control display is shown in Figure 2-10, Figure 2-11 and Figure 2-12.
Figure 2-10: Example of Ready State in Target Control
START
softkey

2-26 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Operation
Figure 2-11: Example of Running State in Target Control
Figure 2-12: Example of Paused State in Target Control
2.6.11.3. Over/Under Applications
If the target comparison of the IND560x will be used to checkweigh the weight of a product, the
application can be classified as an Over/Under application. These types of applications are usually
manual processes but can also be automated. The SmartTrac display indicates the classification of
the weight placed on the scale and discrete I/O can be used to trigger external lights or other
controls. The terminal compares the current weight to a previously entered target with tolerance
values or to zone limits and indicates the results. The following definitions will explain some terms
used in Over/Under applications. The following definitions apply to over/under applications:
Over Limit If a tolerance mode of “Weight value” is selected in setup, the IND560x does not
require the entry of a target value. In this mode, only the upper and lower limits for
the OK zone are entered. When programmed for this mode, the over limit value is
the maximum weight that will be classified as OK. Any weight above than this value
will be classified as Over.
Over/Under Mode The target comparison mode that provides three zone classifications for weight
placed on the scale. The classification can be Under if the weight is below the target
minus the -tolerance, OK if within the tolerance range or Over if the weight is above
the target plus the +tolerance.
Target The target is the ideal weight value for the product being tested. If a container should
weigh exactly 10 kg, the target value is 10 kg.
Tolerance The weight range above and below the target value that will be acceptable as “OK”
in the target comparison. The tolerance can either be entered as a weight deviation
from the target, a percentage deviation from the target or as absolute weight (zone
edges) depending upon setup.
PAUSE
softkey
RESUME
(START)
softkey
ABORT
(STOP)
softkey

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 2-27
Under Limit If a tolerance mode of “Weight value” is selected in setup, the IND560x does not
require the entry of a target value. In this mode, only the upper and lower limits for
the OK zone are entered. When programmed for this mode, the under limit value is
the minimum weight that will be classified as OK. Any weight below than this value
will be classified as Under.
Zones The IND560x provides 3 different classifications or “zones” in the Over/Under mode.
The zones are Under, OK and Over.
2.6.11.4. Target Comparison Parameter Entry
To enter target comparison parameters, either directly edit the active records or load them with data
from a record in the Target Table.
2.6.11.4.1. Direct Editing of Active Records
To directly edit the active record:
1. Press the TARGET softkey . (Refer to Chapter 3, Configuration, the Terminal section,
Softkeys, for information about how to setup softkey display.) The Active Values screen
displays.
2. Use the UP and DOWN navigation keys to scroll through the fields available for editing.
3. Press the ENTER key to select a field to edit.
4. Depending whether the mode is material transfer or over/under, and the type of tolerance
selected, the active target record will show different fields to edit. Use the navigation and
numeric keys to enter values for all of the fields as needed.
Material Transfer,
Either Mode
Over/Under
Weight Deviation or % of Target Weight Value
Target Target Over Limit
Spill + Tolerance Under Limit
Fine Feed - Tolerance
+ Tolerance
- Tolerance
5. Use the alpha keys to enter a description.
6. Press the OK softkey to accept the edits to the active record.
7. Press the ESCAPE softkey to return to the weighing operation screen without saving the
active record edits.
2.6.11.4.2. Loading Records from the Target Table
The TARGET MEMORY softkey can be used in two ways to load records from the Target Table:
Quick Access and List Selection.
Quick Access to Records
Use the Quick Recall mode when the ID of the Target Table record to be loaded is known. Use the
numeric keypad to enter the ID and then press the TARGET MEMORY softkey to load the record.

2-28 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Operation
If the record is available, the data is loaded. If the record is not found, an “ID not found” error
displays.
List Selection
Use the List Selection mode when the ID of the Target Table record is unknown. To use the List
Selection mode:
1. Press the TARGET MEMORY softkey without any preceding data entry. The Target Search
screen displays.
2. Enter any search restrictions required or leave selections as they are to retrieve all records.
3. Press the SEARCH softkey to view the selected records in the table.
4. Use the UP and DOWN navigation keys to scroll through the list until the desired record is
highlighted.
5. Press the OK softkey to load the selected record from the list.
6. Press the ESCAPE softkey to return to the weighing operation screen without loading the
record.
Record Selection Using Shared Data
Alternatively, the Shared Data server can be used to activate target records stored in the Target
Table. Refer to the Shared Data Reference found on the documentation CD for specific instructions.
Target Units
If second and/or third units are established in the IND560x, target records that utilize either second
or third units can be retrieved from the Target Table into Active Target status. The IND560x will
convert the retrieved record into active units when the record is recalled. Pressing the TARGET
softkey will show the original units as recalled from the Target Table.
If an attempt is made to retrieve a record from the Target Table that does not use primary, second or
third units, a “Units Mis-match” error will display indicating that record recall was unsuccessful.
2.6.12. Comparators
Comparators are simple targets, five of which can be configured in setup. They are controlled either
by coincidence or by comparison with a range. The source for comparison can be the Gross
Weight, Displayed Weight, Rate or assigned by a custom TaskExpert application. When assigned to
the home screen, the Comparator softkey permits direct access to the Comparators. This
screen displays each Comparator’s ID, Description, Limit, Unit, High Limit (if applicable) and Active
operator. Only the limit value/s can be edited. To edit parameters other than limits, the Comparator
configuration screen must be accessed in setup.
To access a comparator’s limit editing page directly from the home screen, press the ID number of
the Comparator, then the Comparator softkey. If the ID number is not known, press the Comparator
softkey , use the UP/DOWN keys to select the desired Comparator from the list, and press the
Edit softkey .
For range comparators, the High Limit value must always be higher than the Limit value.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 2-29
When the Active value is <, <=, =, >= or >, the selected source is compared to a target weight.
When the Active value is within (>_ _<) or outside (_< >_) a range, the selected source is
compared to a range defined by the Limit and High Limit values.
2.6.12.1. Using Comparators
To use a Comparator, it must be assigned to a discrete output.
2.6.12.1.1. Target Value Comparators
In this case, the output will be triggered depending on the Active setting. For example, if the Active
operator is <= (less than or equal to) and the Limit is 1000 kg, the output will turn OFF when the
measured value reaches 1000 kg.
2.6.12.1.2. Range Comparators
In the case of a Range mode Comparator, the Limit value sets the lower of the two rates that define
the range, within or outside (depending on the operator) which the output is active. For example, if
the Active operator is Within (>_ _<), the Limit is 3.0 kg/sec and the High Limit is 8 kg/sec, the
output will be active provided the rate remains between 3 and 8 kg/sec. If the rate falls below 3 or
exceeds 8 kg/sec, the output will turn OFF.
Note that the Rate value configured in setup at Scale > Rate permits weight and time units
to be defined, together with the period between measurements and the number of
measurements to average to generate the output value.
If the weight information in the terminal becomes invalid, the Rate display will show 0.000.
A Range Comparator can also use displayed or gross weight as its source. For example, if
displayed weight (possibly a net value, allowing for the weight of a hopper) is the source, the
Active operator is Outside, and the Limit and High Limit are 1000 kg and 1500 kg respectively, the
output will remain active as long as the source value is below 1000 kg. In this case a large
enough value for the High Limit should be set to ensure that, once the Limit is reached and the
output turned off, there is not sufficient spill to reach the High Limit and reactivate the output.
2.6.13. ID
The ID function provides a convenient and simple way to ensure that a specific sequence of
operation is carried out the same way each time. Each step can include an on-screen message that
instructs the operator to perform an action, or a prompt asking the operator to enter some
information. These steps may include data entry (using the numeric keypad, alphanumeric keys, or
some other input such as a barcode scanner) and actions (such as placing a package on the
scale, taking a tare, or printing a ticket for the transaction). When a step is completed, pressing the
ENTER key moves to the next one.
As many as 20 steps may be programmed, and the sequence can be triggered in one of two ways:
 Manually, using the ID softkey
 Automatically, by threshold and reset weight readings from the scale
If the sequence is triggered manually, it may be programmed to loop, repeating all steps until either
the EXIT softkey is pressed, or an ESC softkey is pressed while an alphanumeric entry
field is selected.

2-30 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Operation
When ID mode is set to function automatically, placing on the scale a weight that exceeds a preconfigured threshold value initiates the sequence. When all the steps have been performed and the
weight is removed, the reading from the scale falls below a reset value. At this point the terminal is
ready to begin the next ID sequence.
For details on ID mode configuration, refer to Chapter 3, Configuration.
2.6.14. SmartTrac™
SmartTrac is a graphical display visualization of either the gross or net weight as selected in setup.
SmartTrac on the IND560x uses a bar graph visualization type for Material Transfer mode and a
three-zone graphic for Over/Under mode.
2.6.14.1. Material Transfer Mode
The operator adds material until the measured value is within an acceptable tolerance limit.
Normally an operator will add material quickly when the container is not nearly full and more
slowly as the target value approaches.
The SmartTrac display may be configured in setup as Small, Medium or Large. Each size provides
different target and weight information on the display. Figure 2-13 shows the small size, Figure
2-14 shows the medium size, and Figure 2-15 shows the large size.
Figure 2-13: Material Transfer SmartTrac, Small
Figure 2-14: Material Transfer SmartTrac, Medium
Weight display
SmartTrac
indicating weight
on target
SmartTrac
Target weight
Weight display
Target description

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 2-31
Figure 2-15: Material Transfer SmartTrac, Large
While the bar graph display appears to be one continuous bar graph, it can actually be broken into
four separate sections or ranges. Depending upon the target and tolerance values used, there could
be one, two, or three speeds at which these graphical sections are filled in. Figure 2-16 illustrates
the ranges.
Figure 2-16: SmartTrac Bar Graph Display Ranges
2.6.14.1.1. Under Tolerance
The Under Tolerance range is labeled “A” in Figure 2-16. This range represents the amount of
material from 10% of the target value to the target minus the negative tolerance value. The bar
graph begins to populate the “A” range when more than 10% of the target has been added. Until
reaching the 10% point, no fill is shown.
2.6.14.1.2. Acceptable Under Tolerance
Region B in Figure 2-16 represents the zone of acceptable tolerance below the target value.
2.6.14.1.3. On Target
When the measured value exactly equals the target value an unfilled triangle indication displays
(see Figure 2-17).
Figure 2-17: On-Target Triangle
2.6.14.1.4. Acceptable Over Tolerance
Region C in Figure 2-16 represents the zone of acceptable tolerance above the target value.
2.6.14.1.5. Over Tolerance
Region D in Figure 2-16 represents the zone of unacceptable tolerance above the target value. In
Region D, the fill rate reverts to the same rate as used in Region A. If the measured value is large
SmartTrac
Target weight
No weight display
Target description
Upper tolerance
Lower value
tolerance value

2-32 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Operation
enough that the number of dot columns required for display exceeds the number available, the
display saturates and further increases in the measured value do not affect it.
2.6.14.2. Over/Under Mode
There are three sizes of Over/Under SmartTrac selectable in setup. Each size provides different target
and weight information on the display. Figure 2-18 shows the small size, Figure 2-19 shows the
medium size, and Figure 2-20 shows the large size. Note that Figure 2-18 shows a composite
display with under, OK and over graphics displayed simultaneously.
Figure 2-18: Over/Under Display, Small
Figure 2-19: Over/Under Display, Medium
Figure 2-20: Over/Under Display, Large
2.6.14.2.1. Less than 10% of Target
Until the weight on the scale reaches 10% of the target value, the Under symbol is shown as a
dashed line symbol or , and the discrete outputs are disabled. The Under zone
symbol becomes solid and the discrete outputs become active when more than 10% of the target
has been added.
Weight display
SmartTrac
Weight display
SmartTrac
Target description
Target weight
Target description
Upper tolerance value
Lower tolerance value
Target weight
SmartTrac

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 2-33
2.6.14.2.2. Under Zone
The Under zone is shown in Figure 2-18 as the symbol to the left and also in Figure 2-19. This
zone classification is shown when the weight on the scale is greater than 10% of the target but less
than the target value minus the -tolerance value. If a discrete output has been assigned to the Under
Zone, that output will be “on” when the weight is in this zone.
2.6.14.2.3. OK Zone
When the weight on the scale is between the target minus the -tolerance and the target plus the
+tolerance, the OK symbol is shown. This symbol is the same symbol for all sizes of SmartTrac
and shown in Figure 2-18 as the center symbol. If a discrete output has been assigned as
Tolerance - OK, that output will be “on” when the weight is in this zone.
2.6.14.2.4. Over Zone
The Over zone is shown in Figure 2-18 as the symbol to the right and also in Figure 2-20. This
zone classification is shown when the weight on the scale is greater than the target value plus the
+tolerance value. If a discrete output has been assigned to the Over Zone that output will be “on”
when the weight is in this zone.
2.6.15. Time and Date
Time and date are used for reporting, error and transaction log timestamps, and triggering service
events. The TIME & DATE softkey enables access to the Set Time & Date screen where the
user can set the time and date, including the hours, minutes, day, month, and year. When the time
is set, seconds are set to 0.
Although the format for the time and date can be selected according to local preferences, the use of
a timestamp in log files is not selectable. Timestamp formats are always fixed as:
Date: YYYY/MM/DD (for example, July 23, 2005 becomes the fixed format date 2005/07/23)
Time: HH:MM:SS stored in 24 hour format (for example, 10:01:22 PM becomes the fixed format
time 22:01:22)
2.6.16. Reports
Reports may be generated, viewed, and printed from database tables including:
 Alibi Memory
 Tare Table
 Target Table
The REPORTS softkey must display on the operational screens to generate table reports.
Otherwise, table reports can only be generated from within setup.
To generate a report:
1. Press the REPORTS softkey . The Reports screen (Figure 2-21) displays.

2-34 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Operation
Figure 2-21: Reports Screen
2. Use the UP and DOWN navigation keys to select the desired type of database table report in the
selection box.
3. The CLEAR TOTALS softkey displays when the Tare Table is selected that has totals
enabled in setup. If Clear after Print is enabled in setup for the selected table, the CLEAR TOTALS
softkey does not display. Press the CLEAR TOTALS softkey to clear the totals in the
selected table. A warning displays that requires verification that the user wants to clear the table
totals.
4. Press the PRINT softkey on this screen to print the entire selected report.
5. The CLEAR TOTALS and PRINT functions are only accessible to users with a log-in level of
Supervisor and above.
6. Press the EXIT softkey to return to the home screen.
7. Press the VIEW TABLE softkey to select a sub-set of records from the table. The Search
Screen for the selected report type displays (see Figure 2-22).
Figure 2-22: Search Screen (Tare Table)
8. Press the SEARCH softkey to view all data in the table. Only fields enabled in setup display
in tables.
9. To limit the data:
a. Use the Search Field selection box to select a desired search field (for example, search
fields for the Tare table include Description, ID, n, Tare, and Total).
b. Use the Comparison selection box to select how the data will be limited (less than (<), less
than or equal to (<=), equal to (=), greater than or equal to (>=), greater than (>), or not
equal to (<>) are available in the selection box).
TYPE selection box
PRINT softkey
CLEAR TOTALS
softkey
VIEW TABLE
softkey
EXIT softkey
Search Field
selection box
Comparison
selection box
SEARCH
softkey

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 2-35
c. Use the numeric keypad to enter the limiting factor in the text box (the * character is the
“wild-card” character and returns all results).
d. Press the SEARCH softkey to view the limited data in the table. Search result records
are always listed in order by ID (lowest to highest ID numbers). Use the arrow keys to
scroll left and right to view additional columns. Figure 2-23 is an example of search results
for a Tare Table, with a series of views showing all available columns of data.
Figure 2-23: Tare Table Search Results
2.6.17. Calibration Test
The IND560x terminal provides a programmable Calibration Test sequence that can contain up to
25 individual steps. The Calibration Test is designed to lead the tester through a set of
predetermined steps and compare the terminal calibration with known test weights.
The test is accessed by pressing the CALIBRATION TEST softkey on the home page. A screen
displays where the name of the tester can be recorded and the serial number and capacity of each
of the test weights used can be entered. This provides a traceable record for the calibration test.
The calibration test is started by pressing the START softkey . A display is shown that provides
the active scale weight, the target and tolerance weight values, and two lines of instructions (as to
where to place the weights). When the tester completes these instructions, he or she presses the OK
softkey and the terminal compares the actual weight on the scale to the programmed target
weight for this step. If the comparison passes, the test continues to the next step. If it fails, a
message displays that indicates a failure. After acknowledging the error, the tester can abort the
test, retest this step, or skip this step knowing that it failed.
At the end of the complete test, a print can be initiated that prints all the data of the calibration test.
A full description of the calibration test can be found in Chapter 3, Configuration, Maintenance,
Configure/View, Calibration Test.
2.6.18. CalFree™
The IND560x terminal provides a method to calibrate a scale without using test weights. This is
based on manual entry of capacity and performance data from the load cell or load cell platform.
This method of calibration can be used for initial check-out and testing of systems or when a large

2-36 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Operation
structure is used as the weighing vessel and it is not possible to apply test weights to the structure.
METTLER TOLEDO highly recommends that test weights be used whenever possible as this provides
the most accurate method of calibration.
2.7. Alibi Memory Direct Access
The Alibi Memory stores transaction data that can be retrieved in order to verify transaction
information.
 The information stored in the Alibi Memory includes:
 Transaction counter value
 Date and time of transaction
Gross, net, and tare weights including units of measure
2.7.1. To view and print the Alibi Memory
1. Press the Alibi softkey if programmed as one of the home position softkeys or press the
REPORTS softkey if programmed as one of the home position softkeys then select Alibi
Memory from the selection box.
2. Press the VIEW TABLE softkey . The Alibi Search View screen displays.
3. Use the selection boxes and data entry fields to enter specific search information to limit the
search, or do not enter any search limits to view all Alibi Memory Table information.
4. Press the START SEARCH softkey . The Alibi Search View screen displays the search
results. Records are ordered by date and time with the most recent record shown last.
5. Press the PRINT softkey on this screen to print the entire selected Alibi Memory Table.
2.8. Table Searches
A table search can be accessed using various methods, including the TARE MEMORY softkey ,
TARGET MEMORY softkey , ALIBI softkey , REPORTS softkey , and from the menu tree.
2.8.1. To search a table
1. Access the table to be viewed using the appropriate method.
2. Press the VIEW TABLE softkey . The Search View screen displays.
3. Use the selection and data entry boxes to enter specific search information to limit the search,
or enter * (the “wild-card” character) to view all table information.
4. Press the START SEARCH softkey . The Search View screen displays with the search results.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 2-37
2.9. Remote Display
When properly configured, the IND560x can function as a remote display for another Mettler Toledo
terminal capable of handling Mettler Toledo Continuous Output.
If the Remote IND560x will be used to issue simple commands to another terminal (referred to as
the Master terminal), the Master must also be able to accept Command Inputs (CPTZ).
IMPORTANT: Proper barriers or approved Mettler Toledo communication modules are required
for safe installation and transmission of signals. Please refer to the Installation Manual or
METTLER TOLEDO control drawing 72191600 for specific instrument properties and equipment
guidelines.
2.10. Automatic Reconnection to ACM500 Communication
Module
The communication protocol between the IND560x and the ACM500 requires that the ACM500 be
powered up first. If the ACM500 experiences a power cycle, the message shown in Figure 2-24 will
display. In the past when this occurred, users were required to acknowledge this pop-up message
by pressing the ENTER key. Users were then further required to manually power cycle the IND560x
to re-establish communication between the terminal and the ACM500.
Figure 2-24: ACM500 COM Error Message
A new setting in v3.04 firmware and higher provides the user with two options to re-establish
communication automatically in the event of an ACM500 power loss. In setup, Communication >
Connections includes a new setting called Reconnect Mode. Available settings include Manual and
Automatic.
2.10.1.1. Manual
When Manual is selected, the operator will be presented with two options when the ACM500
communication error (Figure 2-24) is shown on-screen. Pressing ESC will acknowledge and clear
the error message, but no further action will be taken by the IND560x terminal. If OK is pressed, the
error message is acknowledged and the IND560x will complete a power cycle in order to reestablish communication with the ACM500.

2-38 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Operation
2.10.1.2. Automatic
If Automatic is selected, the ACM500 communication error pop-up message will not display onscreen and the user will not be given a choice of when the IND560x will go through a power cycle.
Instead, once the IND560x detects that the ACM500 is back on line, the IND560x will automatically
power cycle itself.
If the IND560x is performing a material transfer cycle, after the power cycle the terminal will re-start
with the cycle paused. The operator will be able to choose between continuing or abandoning the
cycle.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-1
This chapter covers
 Entering Setup Mode
 Exiting Setup Mode
 Setup Menu Tree
 Overview of Configuration
 Scale
 Application
 Terminal
 Communications
 Maintenance
 Restoring Factory Default
Settings
3 Configuration
This chapter provides information about how to configure the IND560x terminal’s
operating system. It describes access to the setup mode, where functions can be
enabled, disabled, or defined by entering parameter values in specific setup
screens.
3.1. Entering Setup Mode
The configuration of the IND560x terminal is accessed through the SETUP softkey . If password
security has been enabled, a login screen (Figure 3-1) displays and the user must enter the correct
password in order to advance into setup. (See the Security section in Chapter 2, Operation, and the
Configuration Options, Terminal, Users section of this chapter for further information about
password setup and security.) To exit back to the home screen without entering any login
information, press the ESCAPE softkey .
Figure 3-1: Login Screen
3.1.1. To enter a username
1. Press the ENTER key to access the Username entry box. The softkeys change to alpha keys
(see Figure 3-2).
Username entry box
Password entry box

3-2 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
Figure 3-2: Login Screen (Alpha Keys)
2. Use the alpha keys to enter the username in the username entry box.
3. Press the ENTER key. The alpha keys no longer display.
3.1.2. To enter a password
1. With focus on the Password field label, press the ENTER key. The softkeys change to alpha keys
(see Figure 3-2) and focus moves to the Password entry box.
2. Use the alpha keys to enter the password in the password entry box.
3. Press the OK softkey . If the username and password are correct, the terminal goes into
setup mode. Once the terminal is in setup mode, the setup menu tree displays.
3.2. Exiting Setup Mode
To exit the setup mode, select Home from the setup menu tree and press ENTER. The default
weighing operation screen displays.
As an alternative, press the first softkey whenever the menu tree is displayed to exit the menu
tree.
3.3. Setup Menu Tree
Each line of the setup menu tree is referred to as a branch (see Figure 3-3). Some branches have
additional branches that become visible when the view of the main branch is expanded. If a branch
has additional branches under it, the collapsed symbol (⊞) initially displays in front of the branch
name. After the branch view has been expanded, the expanded symbol (⊟) displays in front of the
branch name and the additional branches become visible. Any branch that is not expandable is
called a leaf node.
Sets of
alpha keys

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-3
The first branch of the menu tree shown with focus is the Scale branch (Figure 3-3).
Figure 3-3: Setup Menu Tree Example
Use the UP and DOWN navigation keys to move focus through the branches of the setup menu tree.
Press the RIGHT navigation key to expand a branch and the LEFT navigation key to collapse a
branch. When the focus is on a sub-branch, focus can quickly be moved back to the main branch
by pressing the LEFT navigation key.
When a leaf node (singular, non-expandable) branch such as Device or Display is in focus, press
the ENTER key to display the setup screen for that function.
3.3.1.1. Setup Screens
Setup screens enable access to data fields where parameters can be viewed, entered, or modified to
configure the terminal to meet specific application function needs.
3.3.1.2. Navigation
Press the UP and DOWN navigation keys to move through the field labels displayed on each setup
screen, and to move to a second or third setup screen when more than one screen is used for
configuration fields associated with a specific parameter (indicated by the presence of a scroll bar).
The shaded portion of the scroll bar (see Figure 3-4) indicates which screen is displayed.

3-4 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
Figure 3-4: Examples of Setup Screens (Format Time & Date)
3.3.1.3. Data Entry
Press the ENTER key to move the focus from the field label to either the selection box or data entry
box where data is to be entered or edited (see Figure 3-5).
Figure 3-5: Setup Screen
If the field values are presented in a selection box, the current selection will have focus when the
ENTER key is pressed.
To change the field value in a selection box:
1. Press the UP and DOWN navigation keys to scroll through the list and place the focus on the
value to be selected.
2. Press the ENTER key to accept the selection as the value for the field. The selection displays as
the value for the field and the focus moves to the next field label.
If the field value is for a data entry box and alpha/numeric character entry is possible, the alpha
keys will display.
Selection box
with focus
Scroll bar indicating
first setup screen
Field label
with focus
Scroll bar indicating
second setup screen

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-5
To change the field value using alpha/numeric characters:
1. When a data entry box is first entered, the previous data (if present) is in focus. To replace
previous data, use the alpha keys and the numeric keypad to enter the desired value.
OR
2. Press the LEFT and RIGHT navigation keys to move the cursor into position if the value needs to
be edited from a specific point rather than replaced. Position the cursor at the end of the data to
be deleted and press the C key once for each character to be deleted.
3. Press the ENTER key to accept the entered alpha/numeric characters for the field. The entry
displays as the value for the field and the focus moves to the next field label.
4. To exit a setup screen, press the EXIT softkey , which is in the first softkey position. The
setup menu tree displays with the focus on the branch for the setup screen that was exited.
3.4. Overview of Configuration
The setup menu tree can be expanded to show every branch and leaf node in the terminal’s
configuration. Use the navigation keys to select the desired setup screen.
There are five major branches in the setup menu:
 Scale
 Application
 Terminal
 Communication
 Maintenance

3-6 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
Details for each branch are provided in the following Configuration Options section. Figure 3-6
shows the setup menu tree with all branches expanded.
Figure 3-6: The IND560x Menu Tree with All Branches Expanded

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-7
The TaskExpert leaf node is only available if the TaskExpert Application Software module has
been installed into the IND560x.
3.4.1. Configuration Options
Configure terminal options on the setup screens that are available under the five major branches of
the setup menu.
If the metrology switch is in the approved position (SW1-1 = ON), access to the Scale branch
is not permitted. The Scale branch will be expandable and viewable, but changes to setup will
not be permitted. Access will be denied.
3.5. Scale
The Scale branch provides the following access to the configuration of
the connected scales:
 Type*
 Capacity and increment values*
 Calibration*
 Zero parameters*
 Tare parameters*
 Units
 Rate parameters
 Filter parameters*
 Stability parameters*
 Log or print threshold parameters
 Minimum weight parameters
Leaf nodes that are marked with an asterisk (*) above will have different parameters
depending on the connected scale type (analog or IDNet).
A Reset is present at the end of the branch to enable a limited reset to the factory default settings for
the Scale branch parameters.
3.5.1. Type
The Scale Type screen permits a name to be assigned to the scale, displays the scale PCB Type in
the terminal, provides a selection list for the Approval mode, and allows the IND560x to be
configured as a Remote Display for another (Master) terminal. The EXIT softkey will return the
display to the menu tree.

3-8 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
3.5.1.1. Name
The Name field enables entry of the scale identification. Enter the scale name (an alpha-numeric
string of up to 20 characters) in the Name entry box.
3.5.1.2. Scale Type
The Scale Type field automatically shows the hardware configuration. This field will display either
Analog or IDNet, along with Remote. If Remote is selected, the approval option will not be available.
3.5.1.3. Load Cells
Load Cell parameters are shown only for IND560x analog main boards, and only if scale type
selection is Analog. Load Cell parameters must be set correctly for analog IND560x terminals that
will operate under battery power, because the current demand from the load cells determines the
switch off voltage set point of the external battery.
Setting the Load Cell parameters correctly helps to ensure that all available capacity of the battery is
used, and protects the battery pack from a deep discharge that will harm the NiMH cells and
shorten the lifespan of the battery pack
Default values of the Load Cell parameters are 1 x 350 Ω. Enter the total number of analog load
cells into the first field, and the impedance (resistance) of each individual load cell into the 
(Ohm) field. If only the Total Scale Resistance is known, enter 1 as the number of load cells and
the TSR into the  field.
If scale type is set as Remote, the number of cells is automatically set to 0.
3.5.1.4. Approval
Approval (analog load cells only) refers to the metrological (weights and measures) approval
configuration for the specific scale. The Approval selection is not present for IDNet scale bases, but
rather is part of the Service Mode of the base. The selection list can be set to:
 None – no approval is required
 USA
 OIML
 Canada
 Australia
 Argentina
If an approval (USA, OIML, Canada, or Australia) is configured and the metrology security switch,
SW1-1, is set to ON, access to the Scale setup parameters in the menu tree will be limited to view
only.
If the approval is configured as Argentina and the metrology security switch, SW1-1, is set to ON,
access to the entire setup menu is prohibited. If an attempt is made to access the setup menu, an
on-screen message will display stating “Access denied. Scale is Approved.”

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-9
If an approval is selected but SW1-1 is not ON, it will not be possible to leave setup, and a
message will appear: Switch UNSECURED.
3.5.2. Scale Type: Remote
When properly configured, the IND560x can function as a remote display for another IND560x
terminal or for another Mettler Toledo terminal capable of handling Mettler Toledo Continuous
Output. If the remote IND560x will be used to issue simple commands to another terminal (referred
to as the Master terminal), the Master must be able to accept Command Inputs (CPTZ).
3.5.3. Physical Connection of Master and Remote Terminals
Communication for the remote display function occurs over a single serial connection. The
configurable serial ports (COM1, COM2, COM3, COM4, COM5) in the IND560x can support one
output and one input simultaneously, so only a single port is necessary for remote operation.
The standard COM1 port on the main board of the IND560x, COM4 and COM5 (on the optional
fiber optic and current loop interface boards), and the optional COM2 and COM3 ports which would
be installed in the ACM500, each support all three of the assignments used in the remote display
function: Remote Display, Continuous Output and CTPZ Input. Communication can occur via RS-
232, RS-422 or RS-485, but serial port parameters (baud, data bits, parity, flow control) must be
the same for the Master and Remote terminals.
IMPORTANT: Direct connection of COM1 on the IND560x to another Mettler Toledo terminal will
require the use of an appropriate signal barrier! Refer to the Installation Manual for details on
barriers known to work with the IND560x.
3.5.4. Remote Terminal Communication via COM1
If the IND560x will function as the Remote terminal, and the communication will take place over
COM1, the Scale > Type setting must be Remote, as shown in Figure 3-7. Note that the Approval
option is not available in this configuration.
Figure 3-7: Scale Type Setting for Remote Operation

3-10 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
In addition, at Communication > Connections, the COM1 port must be assigned to Remote Display,
as shown in Figure 3-8.
Figure 3-8: COM1 Connection Configured for Remote Operation
IMPORTANT: An appropriate signal barrier is required when connecting a non-intrinsically
safe device up to COM1 on the intrinsically safe IND560x.
3.5.5. Remote Terminal Communication via COM2 or COM3
If the IND560x will function as the Remote terminal and the communication will take place over
COM2 or COM3 in the ACM500, the Scale > Type setting must be Remote, as shown in Figure 3-7.
In addition, COM4 must be set as ACM500, as shown in Figure 3-9. Finally, COM2 or COM3 can
be assigned as Remote Display as shown in Figure 3-10
Figure 3-9: COM4 Connection Configured for ACM500
Figure 3-10: COM2/COM3 Connection Configured for Remote Operation
3.5.6. Remote Terminal Communication via COM4 or COM5
If the IND560x will function as the Remote terminal and the communication will take place via a
legacy fiber optic converter, another IND560x with fiber optic interface installed or the ACM500
 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-11
(COM4 or COM5), the Scale > Type setting must be Remote, as shown in Figure 3-7. In addition,
COM4 or COM5 must be set as Standard on the Connections setup page, as shown in Figure 3-11.
Finally, COM4 or COM5 can then be assigned as Remote Display on the Connections View page as
shown in Figure 3-12.
NOTE: The IND560x cannot function as a remote display for another IND560x if the current
loop interface is used. The current loop connection used in a remote display configuration
requires one active and one passive end in the communication link, and only an active
current loop interface is approved for use in the IND560x.
Figure 3-11: COM4 or COM5 Connection Configured for Fiber Optic Converter or ACM200
Figure 3-12: COM4 or COM5 Connection Configured for Remote Operation
3.5.7. Master Terminal Configuration
The Master terminal must have a port configured to send display information to the Remote
terminal. At Communication > Connections the serial port assignment must be Continuous Output.
If the Remote terminal will be used to send commands to the Master terminal, then the same port
must have a second connection assignment for CTPZ Input.
Figure 3-13 shows a typical Remote Display setup, with the Master terminal able to receive
commands sent from the Remote IND560x.

3-12 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
Figure 3-13: Example of Terminal Configuration for Remote Display Function
3.5.8. Capacity and Increment
When IDNet scale bases are installed, the Capacity & Increment setup screen does not display. It is
part of the Service Mode of the base.
For analog scales use the Capacity and Increment setup screen to select primary units, set the
number of ranges or intervals, and the blanking over capacity.
3.5.8.1. Primary Units
Set the primary units from the selection box choices, which include:
Grams (g) Pounds (lb) Tons (ton)
Kilograms (kg) Tonnes (t)
3.5.8.2. #Ranges
The number of ranges settings are available for analog load cells only. Set the number of ranges
from one to three in the selection box. Specify the capacity and increment size for each range.
3.5.8.3. Blanking Over Capacity
The blanking over capacity settings are available for analog load cells only. Blanking of the display
is used to indicate an over-capacity condition. Set the blank over capacity for the divisions (d)
(display increments) that the terminal is permitted to go over capacity. For example, if capacity is
set at 500 kg by 0.1 kg increments and the blank over capacity setting is 5 d, the terminal can
display weights up to 500.5 kg. At weights over 500.5, dashed lines will display instead of a
weight.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-13
3.5.8.4. X10 Always
The X10 Always selection is available for IDNet load cells only. X10 Always locks the IDNet base
into a higher resolution weight display. If the X10 Always mode is enabled, the high resolution
weight data from the IDNet base displays at all times and the EXPAND BY 10 softkey (if
configured to display) is inactivated.
Ony the Primary Unit field displays and it is not selectable. The unit displayed is the unit in which
the IDNet scale base was calibrated.
3.5.9. Calibration
The Calibration screen enables entry of a geo code adjustment value, the base serial number,
calibration units, linearity adjustment, and analog gain jumper setting.
3.5.9.1. Geo Code
Enter the geo code for the appropriate geo adjustment value for the current geographical location.
Geo codes are numbered 0–31.
The Geo Code field does not display when the IDNet scale base is used.
3.5.9.2. Base Serial Number
Enter the scale base serial number in this field. An entry of up to 14 characters is possible.
3.5.9.3. Calibration Units
The Calibration Units setting is available for analog load cells only. Calibration units are listed in the
selection box and are the same as the primary units selections, which include:
Grams (g) Tonnes (t)
Kilograms (kg) Tons (ton)
Pounds (lb)
3.5.9.4. Linearity Adjustment
The Linearity Adjustment setting is available for analog load cells only. Select the linearity
adjustment from the selection box. Selections are as follows:
Disabled – Use only zero and one span point7
3 point – Use zero, midpoint, and highpoint
4 point – Use zero, lowpoint, midpoint, and highpoint
5 point – Use zero, lowpoint, midpoint, mid-highpoint, and highpoint
3.5.9.5. Analog Gain Jumper
The Analog Gain Jumper Setting is available for analog load cells only. The analog gain jumper
setting on the Main PCB can be either 2 mv/V or 3 mv/V. The terminal is shipped from the factory in
the 3 mv/V setting. In order for the CalFree feature to operate properly, this parameter must indicate
the position of the jumper on the Main board. The jumper settings are:

3-14 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
2 mv/V – Jumper installed on both pins
3 mv/V – Jumper removed
3.5.9.6. Calibration Softkeys
The Calibration setup screen displays softkeys that can be used for calibration procedures,
including:
Capture Zero
Capture Span
Step Calibration
CalFree
Service Mode (IDNet scales)
3.5.9.6.1. Capture Zero
The Capture Zero function in the Calibration setup is available for analog load cells only. The
CAPTURE ZERO softkey triggers an independent operation to reset the zero condition of the
scale. A status message displays when this softkey is pressed that directs the user to empty the
scale and press the START softkey . The status of the capture zero operation displays. When the
operation is complete, a final status message displays that verifies the completion of the capture
zero operation.
If motion is present during the zero capture process, the terminal will average the weight
readings then display an error message indicating unstable weight values were used. This
message must be acknowledged to save the these values.
3.5.9.6.2. Capture Span
The Capture Span function in the Calibration setup is available for analog load cells only. The
CAPTURE SPAN softkey initiates a sequence to capture span that can be performed
independently of capturing zero.
To capture span:
1. Press the CAPTURE SPAN softkey . The Capture Span setup screen displays.
2. Enter the weight for test load 1 and all other test loads if linearity has been enabled. Press the
ENTER key.
3. Place test load weight 1 on the scale.
4. Press the START softkey . The status of the weight capture operation displays. When the
operation is complete, a status message displays that verifies the completion of the weight
capture.
Press the ESCAPE softkey to abort the calibration process and return the display to the
previous screen.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-15
5. After the first calibration step has completed, the menu will either display a prompt for the next
calibration weight to be added (if 2, 3, or 4 test load steps are enabled by the linearity
adjustment parameter) or will show a successful or failed calibration sequence.
If motion is present during the span capture process, the terminal will average the weight
readings then display an error message indicating unstable weight values were used. This
message must be acknowledged to save the these values.
6. Repeat steps 3–4 for test loads 2, 3, and 4 if enabled by linearity adjustment.
7. If the capture span operation was successful, a verification message that reads “Capture Span
OK” displays If the capture span operation was not successful, an error message that reads
“Calibration Failure” displays. If the calibration fails, repeat the capture span procedures. If the
calibration continues to fail, contact a local METTLER TOLEDO representative for assistance.
8. Press the EXIT softkey to return to the Calibration screen.
3.5.9.6.3. Step Calibration
The Step Calibration function in the Calibration setup is available for analog load cells only. The
STEP CALIBRATION softkey initiates a procedure that enables a “build-up” calibration for tanks
and hoppers. For step calibration, the same amount of test weight is added for each step of the
calibration procedure.
To perform a step calibration:
1. Press the STEP CALIBRATION softkey . The Step Calibration setup screen displays.
2. Enter the target weight for the test load (the same amount of test load weight is used in each
step).
3. Press the START softkey . The Step Calibration screen displays. This screen shows an active
(live) weight display (using the current span factor), the intended target weight as entered on
the previous screen, and the next step for the operator “Add test weight.”
4. Add test weight to the tank/hopper as prompted by the display. Each time test weight is added
to the scale, the active display shows the weight.
5. When the full amount of test weight has been added, press the OK softkey . The test load is
captured and span factors are saved. During the time the weight is captured and the new span
factor is being calculated, the display “Add test weight” changes to “Capturing span.” The
display then changes to the next prompt “Remove test weight then fill to target.”
6. Remove the test weight. The active display returns to zero.
7. Fill the tank/hopper with a substitute material up to the target weight.
8. Press the OK softkey . The target weight value recalculates to show the substitute material
weight plus the initial intended target weight. The display changes to the next prompt “Add test
weight.”
9. Add test weight to the tank/hopper. The active display shows the weight.
10. Press the OK softkey . If the actual weight does not equal the target, a new span factor
calculates and the display changes to “Capturing span.” The active weight display changes to
match the target weight value. The display changes to the next prompt “Remove test weight
then fill to target.”

3-16 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
11. Remove the test weight. The active display returns to the previous weight that displayed for the
last test load.
12. Repeat steps 7 through 11 until an appropriate number of calibration steps have been
completed for the specific application.
13. Press the EXIT softkey at any step in the procedure to stop the step calibration process and
return to the Calibration screen.
3.5.9.6.4. CalFree
The CalFree procedure is applicable for analog load cells only. The CalFree softkey provides
access to the span calculation screen for precalibration of a scale without test weights. Highprecision scale bases are precalibrated when received from the factory so this feature is not used.
To perform precalibration of span using CalFree:
1. Press the CalFree softkey . The CalFree screen displays.
2. Enter the load cell capacity and rated load cell output values in the associated fields:
The total load cell capacity should be entered here. For example, for a tank with three 5000 kg
cells, cell capacity would be 3 x 5000 kg or 15000 kg.
If multiple load cells are used, the average output of all cells should be entered here.
3. Enter an estimated preload value in the associated field. Estimated preload is an optional entry.
During calculation, the terminal also tests for analog/digital (A/D) converter input saturation at
full scale capacity. Estimated preload is included in this calculation if entered in this field. If the
preload is unknown, leave this field blank.
This preload value is used only to determine overload conditions and is not used as a zero
reference point. The zero reference point must be captured using the normal zero calibration
procedure.
4. Press the OK softkey . The span is calculated using the parameters entered.
5. If the calibration operation was successful, a verification message that reads “Calibration OK”
displays If the calibration operation was not successful, an error message that reads
“Calibration Failure” displays. If the precalibration fails, repeat the CalFree procedures. If the
precalibration continues to fail, contact a local METTLER TOLEDO representative for assistance.
6. Press the ESCAPE softkey to return to the Calibration screen.
3.5.9.6.5. Service Mode
Service Mode is available for IDNet load cells only. This screen displays a single box that shows
messages from the IDNet base up to 16 characters. When either the (Yes) or (No) softkey
is pressed, the terminal sends the respective response to the scale base and the next message to
be displayed is transmitted back to the terminal from the base. This sequence remains constant
during the complete service mode communication sequence. Refer to the service mode flow chart
for the IDNet base used to view the exact sequence.
After the last communication from the scale base has been completed, the display returns to the
Scale Type screen.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-17
3.5.10. Analog Saturation Test
During the calibration process, the IND560x will determine if there is a potential for analog section
saturation. The IND560x will compare the available number of counts with the captured span to
confirm that the desired capacity is achievable. If the potential exists to saturate the analog section
before system capacity is reached, the IND560x will display a warning message and
recommendation that the system be recalibrated with a new, lower, capacity.
Figure 3-14: Analog Saturation Warning
This message functions only as a recommendation to recalibrate. After the warning message is
acknowledged by pressing the ENTER key, the user is presented with the “Capture Span OK” screen
and the BACK softkey. Service personnel should then carry out the following:
 Confirm that the analog gain jumper on the main board is set correctly for the actual load cell
output, either 2mV/V or 3mV/V (refer to the Installation Manual, for jumper positions).
 If the analog gain jumper is correctly set, the scale should be recalibrated with a lower
capacity. Potential analog section saturation does not force a failure in calibration.
IMPORTANT: It is critical that service personnel follow the recommendation to recalibrate or
confirm that the analog output jumper is properly positioned on the main board.
Analog section saturation will result in a ‘freeze’ of the display, where the display is
no longer showing increases in weight, creating the possibility of overfills and spills
during the weighing process.
3.5.11. Zero
This section provides access to Auto Zero Maintenance (AZM) settings, under zero blanking, powerup zero, and pushbutton zero parameters.
3.5.11.1. AZM & Display
AZM is a means of tracking zero when the scale is empty. AZM compensates for conditions such as
terminal or load cell drift or debris on a scale platform.
3.5.11.1.1. Auto Zero
Use the AZM & Display setup screen to enable auto zero for gross weighing or gross and net
weighing, or to turn auto zero off.

3-18 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
3.5.11.1.2. Auto Zero Range
Analog load cells: Set the auto zero range for the number of divisions (d) around the current zero
setting in which auto zero will operate.
IDNet load cells: Either enable or disable the auto zero function.
3.5.11.1.3. Under Zero Blanking
The Under Zero Blanking function in the Zero setup is available for analog load cells only. Blanking
of the display is used to indicate an under-zero condition when the weight on the scale falls below
the current zero reference. Set the under zero blanking for the number of divisions (d) that the
terminal is permitted to go under zero before blanking.
A value of 99 disables blanking under zero and the terminal will display a weight as far under
zero as possible.
3.5.11.1.4. Power Up
A Power Up setting of Restart enables the terminal to save and reuse the last zero reference weight
after a power cycle so it returns to the same gross weight value. If reset is selected, the last zero
calibration value will be used as the zero reference point. Select either Reset or Restart from the
Power Up selection box.
3.5.11.2. Ranges
Use the settings on the Ranges screen to enable or disable Power Up Zero and Pushbutton Zero and
to set the ranges around the original zero condition for the scale for applying these functions.
3.5.11.2.1. Power Up Zero
The Power Up Zero function in the Zero setup is available for analog load cells only. If Power Up
Zero is enabled, the terminal tries to capture zero upon power up.
3.5.11.2.2. Range
The Range setting for Power Up Zero is available for analog load cells only. If Power Up Zero is
enabled, –Range and +Range fields will display for setting the range around the original zero
condition for the scale within which Power Up Zero can be applied. The range units are percent.
For example, if the +Range setting for Power Up Zero is set at 2%, Power Up Zero will only occur
when the weight reading on the scale is at or below 2% of the original zero condition. If the –Range
setting for pushbutton zero is set at 2%, Power Up Zero will only occur when the weight reading on
the scale is at or above –2% of the original zero condition.
If Power Up Zero capture is enabled and the weight on the scale is outside of the zero capture
range, the display will indicate EEE until the weight is removed and zero is captured.
3.5.11.2.3. Pushbutton Zero
If Pushbutton Zero is enabled, the front panel ZERO pushbutton will operate to capture zero
reference points.
If Pushbutton Zero is Disabled, execution of a remote Zero is still possible via SICS, CPTZ, or
SMA commands, discrete inputs or from the PLC commands.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-19
3.5.11.2.4. Range
The Range setting for Pushbutton Zero is available for analog load cells only. If Power If Pushbutton
Zero is enabled, –Range and +Range fields will display for setting the range around the original
zero condition for the scale within which Pushbutton Zero can be applied. The range units are
percent.
For example, if the +Range setting for Pushbutton Zero is set at 2%, the Pushbutton Zero can only
be used when the weight reading on the scale is at or below 2% of the original zero condition. If
the –Range setting for Pushbutton Zero is set at 2%, the Pushbutton Zero can only be used when
the weight reading on the scale is at or above –2% of the original zero condition.
3.5.12. Tare
Tare is used to subtract the weight of an empty container from the gross weight on the scale to
determine the net weight of the contents. Tare is inhibited if the scale is in motion.
Three setup screens are available to configure tare:
Tare Types Auto Tare Auto Clear
3.5.12.1. Tare Types
Use the Tare Types setup screen to enable or disable pushbutton tare, keyboard tare, net sign
correction, terminal tare (IDNet load cells only).
3.5.12.1.1. Pushbutton Tare
When pushbutton tare is enabled, the TARE scale function key can be pressed when an
empty container is on the scale to determine tare. The terminal displays a zero weight and net
mode. When the container is loaded and placed back on the scale, the terminal displays the net
weight of the contents.
If Pushbutton Tare is Disabled, execution of a remote Tare is still possible via SICS, CPTZ, or
SMA commands, discrete inputs or from the PLC commands.
3.5.12.1.2. Keyboard Tare
When keyboard tare is enabled, the known value for the empty weight of a container (tare) can be
entered manually. The terminal will then display the net weight of the contents of the container.
Keyboard tares are automatically rounded to the closest display division.
3.5.12.1.3. Net Sign Correction
Net sign correction enables the IND560x terminal to be used for both shipping (inbound empty)
and receiving (inbound loaded) operations. If net sign correction is enabled, the terminal will switch
the gross and tare weight fields on the printed ticket, if necessary, so that the larger weight is the
gross weight, the smaller weight is the tare weight, and the difference is always a positive net
weight. Net sign correction affects the printed data output, the recalled weigh display and the
displayed weight. Continuous data output will continue to show a negative net weight value.
Net sign correction will operate with pushbutton tare, preset tare, or tare memories. An example of
weight values with and without net sign correction is shown in Table 3-1. In this example, the tare
register value is 53 kg and the live weight on the scale is 16 kg.

3-20 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
Table 3-1: Weight Values With and Without Net Sign Correction
Net Sign Correction
Printed and Displayed Disabled Enabled
Gross 16 kg 53 kg
Tare 53 kg 16 kg
Net 37 kg 37 kg
 When net sign correction is enabled, the tare weight field in the recall display will be
labeled with the letter “M” to indicate “Memory” instead of “T” or “PT”.
3.5.12.2. Terminal Tare
The Terminal Tare field only displays when an IDNet scale base is connected. Terminal Tare is
disabled by default. When Terminal Tare is disabled, all Tare commands received by the IND560x
via SICS, CTPZ, PLC or keypad will be passed through to the IDNet base for execution. The base will
calculate the tare and net weight values and pass this information back to the IND560x. When
Terminal Tare is enabled, tare and net weight values are calculated in the terminal and not in the
high precision IDNet base.
When Terminal Tare is enabled, the stability timeout value will be set to 3 seconds, but can be set
to any value between 0 and 99 at Scale > Stability. A value of “99” instructs the terminal to wait
indefinitely for a stable condition – a Tare command would never be aborted due to motion.
When Terminal Tare is disabled, the stability timeout value defaults to “99”, and cannot be
modified. The terminal will not display or transmit Net status, Net weight or Tare weight until it
receives this information from the scale base. Only when the IDNet base responds will the terminal
display and transmit the tare value and Net status to the PC/PLC.
Terminal Tare should be disabled for approved (legal-for-trade) systems.
3.5.12.3. Auto Tare
Use the Auto Tare screen to enable or disable automatic tare and to set the tare and reset threshold
weights, and enable or disable motion check.
3.5.12.3.1. Auto Tare
When auto tare is enabled, the tare weight is taken automatically when a container above the
threshold weight is on the scale and settles to no-motion.
3.5.12.3.2. Tare Threshold Wt.
When weight on the scale platform exceeds the tare threshold value and settles to no-motion, the
terminal automatically tares.
3.5.12.3.3. Reset Threshold Weight
The reset threshold weight must be less than the tare threshold weight. When the weight on the
scale platform falls below the reset threshold value, such as when a load has been removed, the
terminal automatically resets the auto tare trigger, depending upon the programming of motion
checking.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-21
3.5.12.3.4. Motion Check
Enable the motion check setting to prevent auto tare reset rigger from occurring when the scale is in
motion. When enabled, the scale must detect a non-motion condition below the reset value to reset
the trigger.
3.5.12.4. Auto Clear
Use the Auto Clear screen to enable or disable auto clear tare, clear after print, to set the clear
threshold weight, and enable or disable motion checking for auto clearing of tare.
3.5.12.4.1. Auto Clear Tare
To clear tare automatically when the scale returns to below the threshold weight, enable the auto
clear tare setting.
3.5.12.4.2. Clear Threshold Wt.
When the gross scale weight exceeds then falls below the clear threshold weight value, the terminal
automatically clears tare and returns to gross mode.
3.5.12.4.3. Motion Check
The Motion Check field displays only when Auto Clear Tare is enabled. Enable the motion check
setting to prevent auto clear from occurring when the scale is in motion.
3.5.12.4.4. Clear After Print
The Clear After Print field displays only when Auto Clear Tare is enabled. To clear tare automatically
after printing, enable the clear after print setting.
3.5.12.4.5. Clear With Zero
To clear tare automatically when capturing zero, enable the clear with zero setting.
3.5.12.4.6. Power Up
A Power Up setting of Restart enables the terminal to save and reuse the last tare weight after a
power cycle. If Reset is selected, the terminal returns to gross mode upon power up and the last
tare weight after the power cycle is cleared. Select either Restart or Reset from the Power Up
selection box.
3.5.13. Units
This setup screen enables the selection of a second and a third unit and power up units.
3.5.13.1. Second Unit and Third Unit
Use the Second Unit and Third Unit selection boxes to select second and third weighing units,
configure custom second and third units, or to select none. Only one custom unit is possible.
If printing of two units is required, establish Third Units. Both Primary and Third units can be printed
simultaneously in an output template.

3-22 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
Available weighing units include:
Custom Pounds-Ounces (lb-oz)
Pennyweight (dwt) Ounces (oz)
Grams (g) Troy Ounces (ozt)
Kilograms (kg) Tonnes (t)
Pounds (lb) Tons (ton)
3.5.13.2. Power Up
Power up units defines the units the terminal defaults to after power up. Use the selection box to
select either Primary Units or Restart, which makes the power up units those that were active when
power was removed.
3.5.13.3. Custom Units
If Custom is selected for second or third units, three additional fields display for configuring custom
units. The terminal displays the first three letters of the custom name in the units area of the display
when custom units are in use.
3.5.13.3.1. Custom Factor
Enter a multiplication factor for the custom unit in this field, such as 0.592 or 1.019. The primary
unit is multiplied by the custom factor to obtain the custom value.
3.5.13.3.2. Custom Name
The softkeys turn into alpha keys when this field is selected. Use the alpha keys to enter the name
for the custom unit. Up to 12 characters may be entered.
3.5.13.3.3. Custom Increment
Enter a custom increment in this field, such as 0.1 or 0.5. This step controls both the decimal
position and the increment size of the custom unit value.
3.5.14. Rate
A Rate, expressed as change in primary or secondary weight units per time unit, may be
programmed for display on the IND560x (Figure 3-15) and/or used in controlling discrete outputs
– for example as the source for a Comparator (described later in this chapter).
The Rate display is over-ridden by SmartTrac. If SmartTrac is enabled, the Rate display
option is not available in the Terminal > Display screen.
If the weight information in the terminal becomes invalid, the Rate display will show 0.000.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-23
Figure 3-15: Rate Display on Home Screen
The following Rate parameters can be configured:
3.5.14.1.1. Weight Units
Weight Units may be set to None, Primary or Secondary. The Primary and Secondary selections
correspond to the Units setting in the Scale > Capacity & Increment screen. Secondary units must
be enabled at Scale > Units in order for secondary to be a unit selection for the Rate function. When
the default value, None, is selected, the rate function is disabled. If the unit is a Custom Unit, a
Custom Unit Name of up to 3 characters can be displayed.
3.5.14.1.2. Time Units
Time Units may be set to Seconds, Minutes or Hours. The default value is Seconds.
3.5.14.1.3. Measurement Period
The Measurement Period value sets the frequency of weight measurements. 1 second is the default
value. Other options are 0.5 and 5 seconds.
3.5.14.1.4. Output Average
Output from the rate function is set by a computed rolling average of measurements. This possible
range of values is 1 to 99 seconds, with a default value of 1. The Output Average value should be
sufficiently larger than the Measurement Period to allow the terminal to take multiple measurements
for each average.
3.5.14.2. Filter
The IND560x terminal has a low-pass, multi-pole vibration filter that can be set for several conditions
when using analog load cells. The heavier the filtering, the slower the display settling time will be.
3.5.14.2.1. Low Pass Frequency
Low Pass Frequency configuration is available for analog load cells only. Low pass frequency is the
frequency above which all disturbances are filtered out. The lower the frequency, the better the
disturbance rejection, but the longer the settling time required for the scale.
3.5.14.2.2. Low Pass # of Poles
Low Pass # of Poles configuration is available for analog load cells only. The number of poles
determines the slope of the filtering cutoff. For most applications, a slope value of 8 is acceptable;
however, decreasing this number will improve settling time slightly.

3-24 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
3.5.14.2.3. Notch Filter Frequency
Notch Filter Frequency configuration is available for analog load cells only. The notch filter allows
selection of one specific frequency below the low pass filter value that can also be filtered out. This
enables setting the low pass filter higher to filter out all but one frequency (that the notch filter will
handle) and obtain a faster settling time. Values from 0 (disable) to 99 are possible.
3.5.14.2.4. Stability Filter
Stability Filter configuration is available for analog load cells only. The stability filter works in
conjunction with the standard low pass filter to provide a more stable final weight reading. The
stability filter should only be used in transaction weighing applications, since the nonlinear action of
the filter switching may cause inaccurate cutoffs in batching or filling applications. The stability filter
can be enabled or disabled on this setup screen.
3.5.14.3. IDNet Filtering
Configuration for IDNet filtering parameters is performed on the Filter screen. Use the Vibration
selection box to select a condition setting that reflects the specific location’s conditions. Selections
include:
Ideal Conditions The weighing platform operates very quickly. However, it is very sensitive.
This setting is suitable for a very calm and stable weighing location.
Average Conditions This is the factory default setting and is suitable for most normal
environments.
Extreme Conditions The base reacts to changes in weight more slowly but is much more stable
in unstable environments.
Use the Weighing Process selection box to select the specific weighing process in use. Selections
include:
Fine Filling Used when liquid or fine powders are being weighed
Universal Weighing For solid materials coarse filling or checkweighing
Static Weighing For solid materials and weighing under extreme conditions such as strong
vibrations
Dynamic Weighing For weighing products that will not be fully stable or still during the weighing
process. Not intended for animal weighing.
3.5.15. Stability
The IND560x terminal includes a stability detector (weight in motion). For analog load cells, the
Stability setup screen enables setting a motion range, no-motion interval and timeout period. For
IDNet bases, a stability measure and timeout period may be set.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-25
3.5.15.1. Analog Load Cell Stability
3.5.15.1.1. Motion Range
Sets the motion range to the weight value (in divisions) that the weight is permitted to fluctuate and
still have a no-motion condition. Values from 0.1 to 99.9 are possible with the default value being
1.0.
3.5.15.1.2. No-motion Interval
The no motion interval defines the amount of time (seconds) that the scale weight must be within the
motion range to have a no-motion condition. Values from 0.0 (motion detection disabled) to 2.0 are
possible, the default value being 0.3 seconds. A shorter interval means that a no-motion condition is
more likely, but may make weight measurement less precise.
3.5.15.1.3. Timeout
Defines the period (in seconds) after which the terminal stops attempting to perform a function that
requires a no-motion condition (such as a zero, tare or print command) and aborts the function. This
timeout is used regardless of the source of the command such as the keypad, discrete input, PLC or
SICS. Values from 0 to 99 are possible, the default value being 3. A smaller value means that less
time will be used to check for no-motion before aborting a command. When a value of 0 is entered,
there must be no motion when a command is given or it will fail immediately. A value of 99 is a
special condition which permits the terminal to wait indefinitely for a no-motion condition - a
command would never be aborted.
3.5.15.2. IDNet Stability
3.5.15.2.1. Stability
Stability settings for IDNet load cells are configured by selecting a number setting of 0 (disabled), 1
(fast display, good repeatability), 2 (slow display, better repeatability), 3 (slower display, better
repeatability), or 4 (very slow display, very good repeatability) in the Stability selection box.
3.5.15.2.2. Timeout
Defines the period (in seconds) after which the terminal stops attempting to perform a function that
requires a no-motion condition (such as zero or tare) and aborts the function. This timeout is used
regardless of the source of the command such as the keypad, discrete input, PLC or SICS. Values
from 0 to 99 are possible with the default value being 3. A smaller value means that less time will be
used to check for no-motion before aborting a command. When a value of 0 is entered, there must be
no-motion when a command is given or it will fail immediately. A value of 99 is a special condition
which permits the terminal to wait indefinitely for a no-motion condition - a command would never be
aborted.
3.5.15.3. Log or Print
The Log or Print setup branch is where the thresholds to control how and when data is saved or
output to a printer are defined. Normal demand mode printing occurs whenever a print request is
made, providing there is no motion on the scale and zero has been captured (a negative gross
weight will not be printed).

3-26 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
The weight values entered are the gross weight values in primary units. Gross weight in primary
units is used regardless of whether the IND560x is in Gross or Net mode and regardless of the units
that are displayed.
3.5.15.3.1. Minimum Weight
The minimum weight setting is the threshold below which log or print functions will not initiate.
Primary units are displayed for this field.
3.5.15.3.2. Interlock
Interlock prevents repeat logging and printing. If enabled, interlock requires that the weight reading
be reset per the Reset on parameter setting (see below) and then settle to a weight greater than the
minimum print value before responding to the next log or print request.
3.5.15.3.3. Automatic
Enable the automatic setting to log data and send a print request every time the weight on the scale
settles to a positive value that is larger than the entered threshold weight value.
If Automatic is set to Disable, the Threshold Weight field does not appear.
3.5.15.3.4. Reset on
The resetting of auto printing or logging can be based on weight threshold or weight deviation
values. Select Return (the weight must return to below this value to reset) or Deviation (the weight
must change more than this value to reset) from the selection box and enter the weight value in the
“Reset on” field.
If both Interlock and Automatic settings are disabled, the Reset on field does not display.
If Reset On is set to Deviation, The Threshold Weight and Motion Check fields do not display.
3.5.15.3.5. Threshold Weight
The Threshold Weight is the value above which an Automatic logging or printing of data can/will
occur. Threshold Weight does not appear if Automatic is set to Disabled or if Reset On is set to
Deviation.
3.5.15.3.6. Motion Check
Enable the motion check setting to prevent interlock and automatic log and print functions from
resetting when the scale is in motion below the “Reset On” return point. Motion Check does not
appear if Reset ON is set to Deviation.
3.5.15.4. MinWeigh
Enable the MinWeigh function to compare the current net weight with a MinWeigh value. If the net
weight is greater than or equal to the MinWeigh value, all terminal functions behave normally. If the
absolute value of the net weight is less than MinWeigh, the weight display includes a flashing
MinWeigh symbol to the left of the weight. If the user attempts to record the weight while in this
condition, the printout will include an asterisk ().
3.5.15.4.1. Entry Method
The MinWeigh value entry method can be either calculated by the IND560x or directly entered.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-27
Calculated
MinWeigh is composed of 4 factors, each determined independently:
MinWeigh =
U0  SF  100%
T – (c  SF  100%)
Where:
U0 = Uncertainty in measurement as the applied load approaches 0. U0 is computed
differently in each country and is entered in Primary units of measure.
T = Tolerance in percent, which reflects the tolerances required for the specific process and
facility. The range is 0.1%–99.9%.
SF = Safety factor, which is another adjustment means. Usually SF = 1. The integer value
range is 1–10.
c = Uncertainty factor related to the portion of uncertainty in measurement that is
proportional to the applied load. For reference only, c = (UMAX – U0) / Max, where Max
= maximum weighment and UMAX = measurement uncertainty at maximum weight.
When Calculated is selected, the user is prompted to enter the uncertainty U0, c factor, tolerance
percent, and safety factor in data entry boxes. The new MinWeigh value is then computed by the
weighing equipment.
Direct
For direct entry of a MinWeigh value, user enters the desired value for MinWeigh directly in the
MinWeigh data entry box. The MinWeigh value is entered in Primary units of measure.
3.5.15.5. Reset
The Reset screen enables the Scale branch setup values to be reset to factory default settings.
3.5.15.5.1. Scale Reset
To initiate a reset, press the OK softkey . If the reset was successful, a verification message that
reads “Reset Successful” displays. If the reset was not successful, an error message that reads
“Reset Failure” displays. If the reset fails, try to initiate the reset again. If the reset continues to fail,
contact a local METTLER TOLEDO representative for assistance.
Press the ESCAPE softkey to exit without resetting.
Scale reset does NOT include the reset of metrologically significant parameters – scale type,
approval, weight units, capacity, increment, or calibration data. This data is reset only by
performing a Master Reset with both switches 2-1 and 2-2 in their ON positions.

3-28 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
3.6. Application
Use application setup screens to configure:
 Memory and tables
 Operation of targets, totalization and the ID function
 Discrete I/O
3.6.1. Memory
Memory setup screens include:
 Alibi
 Tare Table
 Message Table
 Target Table
3.6.1.1. Alibi
The Alibi Memory can be enabled or disabled in the selection box. Alibi memory is configured as a
“ring” buffer that overwrites the oldest record when it reaches its memory limit. Alibi memory can
hold approximately 60,000 transactions before it reaches its limit and begins overwriting old
transactions. More detailed information about Alibi memory can be found in Chapter 2, Operation.
The Alibi Memory table stores basic transaction information that is not user-definable. This
information always includes:
 Date and time stamp
 Transaction counter value
 Gross, tare, and net weights and weight unit
If the IND560x terminal has been programmed as “approved”, Alibi Memory enabling or
disabling is only accessible if the security switch (SW1-1) is in the OFF position.
3.6.1.2. Tare Table
The Tare Table displays stored tare records, including:
 Tare record ID
 Tare weight value and unit
 Description
 Total number of transactions using each stored tare record
 Total (accumulation of tare weights for each stored tare record)
More detailed information about the tare table memory can be found in Chapter 2, Operation.
Use the Tare Table setup screen to enable or disable description and to configure totalization.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-29
3.6.1.2.1. Description
Description is an alphanumeric string that describes the Tare Table entry. Use the Description
selection box to enable or disable Description in the Tare Table structure.
3.6.1.2.2. Totalization
Totalization is a field that tracks the total weight for all transactions involving each tare in the table.
Use the Totalization selection box to select None, Displayed Weight, or Gross Weight for totals in the
Tare Table structure.
Press the CLEAR softkey to reset the table. Log-in must be at supervisor level or above to reset the
table.
3.6.1.2.3. To view Tare Table records
Press the VIEW TABLE softkey The Tare Search screen displays.
1. Use the selection boxes and associated fields to enter specific search information to limit the
search, or enter * (the “find all” character) to view all Tare Table information.
2. Press the START SEARCH softkey . The Tare Search View screen displays with the search
results. Only records with non-null tare values display. Records are ordered by ID, with the
lowest ID number shown first.
3. Press the UP, DOWN, LEFT, and RIGHT navigation keys to scroll up and down and across the
screen in order to view all data and all records listed.
3.6.1.2.4. To modify or add Tare Table records
1. Press the UP and DOWN navigation keys to select (highlight) a record in the table.
2. Press the EDIT softkey to open the setup screen for editing a record or press the INSERT
softkey to open the setup screen to create a new table record.
3. Press the UP and DOWN navigation keys to move the focus to the field name to be edited or
inserted.
4. Press the ENTER key to select a field value to edit or insert. The alpha keys display.
5. Use the alpha keys and the numeric keypad to edit or enter the desired value.
6. Press the TARE softkey to capture the current live scale weight and unit, which will
display in the Tare field.
7. Press the OK softkey to accept the modifications or additions to the Tare Table.
8. Press the ESCAPE softkey to return to the Tare Search View screen without saving
modifications or additions.
9. Press the DELETE softkey to delete a tare record in the list.
10. Press the PRINT softkey to print the list. Login must be at supervisor level or above to
print the table.
11. Press the EXIT softkey to return to the Tare Search Screen.

3-30 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
3.6.1.3. Message Table
The Message Table displays stored text messages and corresponding ID numbers that can be used to
print templates. Only message records with non-null values display.
Press the CLEAR softkey to clear the entire table. Login must be at supervisor level or above to clear
the table.
To view Message Table records:
1. Use the selection boxes and associated fields to enter specific search information to limit the
search, or do not enter any search limits to view all Message Table information.
2. Press the START SEARCH softkey . The Message Search View screen displays with the
search results. Records are ordered by ID, with the lowest ID number shown first.
3. Press the UP and DOWN navigation keys to scroll up and down the screen in order to view all
data and all records listed.
4. Modify or add Message Table information by editing, inserting, or deleting information as
described previously under Tare Table.
5. Press the PRINT softkey to print the list.
6. Press the EXIT softkey to return to the Search Screen.
3.6.1.4. Target Table
Target comparison can be used in two types of applications: material transfer and over/under. Material
transfer applications require that a control device deactivate when a target value is achieved.
Over/Under applications classify a load placed on the scale platform as above or below the target
value.
Values in the active record registers such as the target weight and tolerance, which can be manually
modified or loaded from the Target Table, are compared with either the displayed or gross scale
weight.
Use the Target Table setup screen to select the mode, tolerance type, and output type for use in target
comparison.
More detailed information about the target table can be found in Chapter 2, Operation.
3.6.1.4.1. Mode
Use the Mode selection box to select the type of application for target comparison. Selections include:
 None
 Material Transfer
 Over/Under
3.6.1.4.2. Tolerance Type
Use the Tolerance Type selection box to select the type of tolerance to be used for target
comparison. Positive tolerance and negative tolerance values can be entered as a deviation weight

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-31
value in the same units as the target (Target Deviation) or as a percent of the target value (% of
Target).
When over/under mode is selected, a weight value choice is available as a tolerance type. In this
mode, there is no target value used – only under limit and over limit values are used as zone edges
for OK.
3.6.1.4.3. Output Type
The Output Type field is available only when the Material Transfer mode is selected. Use the Output
Type selection box to select the type of two-speed output to be used in the target application.
Selections include:
Concurrent Both outputs operate at the same time
Independent One output operates at a time (Fast Feed then Feed)
Press the CLEAR softkey to reset the entire target table. Log-in must be at supervisor level or
above to clear the table.
To view Target Table records:
1. Press the VIEW TABLE softkey The Target Search screen displays.
2. Use the selection boxes and associated fields to enter specific search information to limit the
search, or do not enter any search limits to view all Target Table records.
Search fields that display in the Search Field selection box vary depending on the selections
made on the Target Table setup screen.
3. Press the START SEARCH softkey . The Target Search View screen displays with the search
results. Only records with non-null values display. Records are ordered by ID, with the lowest
ID number shown first.
4. Press the UP, DOWN, LEFT, and RIGHT navigation keys to scroll up and down and across the
screen in order to view all data and all records listed.
Target Table records include different information, depending on the selections made on the Target
Table setup screen. For example, Over Limit and Under Limit data displays only when the
Over/Under mode is selected with Weight Value tolerance type.
Target Table records can include the following fields:
ID Identification number for the active record
Description Descriptive identification for the active record
Target The desired measured value for a weighment
Target Units The units of measure for the target weight (the target record units and the
comparison units can be a combination of primary, second or third units)
Spill The amount of material delivered after the control device is signaled to stop
Positive
Tolerance
The highest acceptable tolerance above a target value

3-32 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
Negative
Tolerance
The lowest acceptable tolerance below a target value
Fine Feed Value For two-speed applications, the fine feed value determines when the Fast
Feed output turns off.
Modify or add Target Table information by editing, inserting, or deleting information as described
previously under Tare Table.
Press the PRINT softkey to print the list. Login must be at supervisor level or above to print the
table.
Press the EXIT softkey to return to the Search Screen.
3.6.2. Operation
Operation setup screens include:
 Target
 Comparators
 Totalization
 ID
3.6.2.1. Target
Use this setup screen to select which live measurement data stream (source) to use as input for
target comparison and to enable or disable latching.
3.6.2.1.1. Source
Select which source to use as input for target comparison, either Displayed Weight or Gross Weight.
3.6.2.1.2. Motion Check
This parameter is shown only if target mode is set to Over/Under (at Application > Memory > Target
Table). The default value is disabled. If Motion Check is enabled, the discrete outputs for Over,
Under and OK are only activated when a no-motion condition is detected.
3.6.2.1.3. Latching
If the Target Table mode selection is Material Transfer, output latching can be selected. When output
latching is enabled, the target comparison outputs remain latched (OFF) after the output switching
threshold is exceeded until the latch is reset by a start input (softkey or discrete input).
If latching is disabled, the outputs will operate as coincidence-type outputs with no interlocks.
3.6.2.2. Comparators
The Comparators screen permits the configuration of simple targets controlled either by coincidence
or by comparison with a range, and used as an assignment for Discrete I/O outputs. Comparators
are controlled either by coincidence or by comparison with a target or range. The source for
comparison can be the Gross Weight, Displayed Weight, Rate or assigned by a custom TaskExpert
application.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-33
3.6.2.2.1. Home Screen Comparators Softkey
The Comparator softkey can be assigned to the weight display home screen. It opens the
Comparators view screen, which displays the ID, Description, Limit, Unit, High Limit and Active
values for each Comparator. An EDIT softkey opens a screen that display’s the selected
Comparator’s Description, and permits its Limit and High Limit values to be modified. Changes can
be saved ( ) or discarded ( ) when returning to the Comparators view screen.
3.6.2.2.2. Comparators Configuration Screen
Source, Active, Description and Limit values can only be set from the Application > Operation >
Comparators screen. This screen displays all five Comparators, and contains columns for ID,
Description, Source, Limit, Unit, High Limit, and Active condition. From here, it is possible to EDIT
a Comparator, delete one , or clear all Comparators.
Once parameters have been set in the Comparator Edit screen, they can be saved by pressing OK
, or discarded by pressing ESCAPE . In either case, the view returns to the Comparators
screen.
3.6.2.2.3. Source
The default value for Source is None, meaning that the comparator is disabled, and no other
parameter fields are displayed. Other choices for the Source are Displayed Weight, Gross Weight,
Rate and Application. If a weight field is selected as the Source, the weight unit will be primary
units. If Rate is selected as the Source, then the unit will be the same as selected for Rate in the
Scale > Rate branch. If Application is selected, no other parameter fields are displayed. This
selection allows a TaskExpert custom application to control the source assignment for the
Comparators.
3.6.2.2.4. Active
The Active setting determines the range of other options available in the Comparator Edit screen.
These options are summarized in Table 3-2.
When Active is set to less than (<, the default value), less than or equal to (<=), equal to
(=),greater than or equal to (>=), greater than (>), or not equal to (< >), the output is active
depending on the relationship between the current Source value and the Limit. When Active is set to
within (>_ _<) or outside (_< >_) a range, the source value is compared to the target value range
set in the Limit and High Limit fields.
Table 3-2: Comparator Configuration
Source None*, Displayed Weight, Gross Weight, Rate
Active <*, <=, =, >=, >, <
>
>_ _< (within), _< >_ (outside)
Description 20 character alphanumeric string
Limit
Target value Lower target value for the comparison
range
High Limit
n/a Upper target value for the comparison
range

3-34 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
3.6.2.2.5. Description
The description is an alphanumeric string that is used to identify the type and purpose of the
comparator. This string appears, together with the automatically assigned ID number (1-5), in the
list that displays when the Comparator softkey is pressed.
3.6.2.2.6. Limit
The Limit either sets the target weight value to which the actual source value is compared, or the
lower target value for the range to which the currently measured source value is compared.
3.6.2.2.7. High Limit
The High Limit is available only for Range mode, and sets the upper target value for the range to
which the currently measured source value is compared. Its value must be higher than the Limit – if
a lower value is entered, the terminal will show an Invalid Value message when the OK softkey
is pressed. The value is expressed in the programmed Rate unit.
3.6.2.3. Totalization
Knowing how many weighing transactions were performed and how much material was processed
during a particular period of time is useful information for many weighing applications.
The IND560x terminal provides both grand total (GT) and subtotal (ST) registers and counters.
Counters have a limit of 1,500,000 and registers will accumulate up to 11 digits of weight
including any decimal places to the right of the decimal point. For example, a scale programmed
for 500 x 0.1 kg will accumulate weight values up to 9999999999.9 (11 total digits). If either of
these limits is exceeded, an error message will display and the totals must be reset before
additional weights or counts will be added.
Use the Totalization setup screen to select parameters for totalization operations, including which
source to use as input for totalization, settings for grand totals and subtotals, and to enable or
disable the conversion of second or third unit weights for totalization.
3.6.2.3.1. Mode
Select which source to use as input for totalization comparison – either Displayed Weight or Gross
Weight. A selection of None disables totalization.
3.6.2.3.2. Clear GT on Print
GT can be configured to clear automatically after printing the Totals report. If Clear GT on Print is
enabled, the subtotal also clears automatically after printing the Totals report.
3.6.2.3.3. Subtotal
ST can be separately disabled while GT continues to accumulate weights. Choose to either enable
or disable the subtotal register.
3.6.2.3.4. Clear ST on Print
Clearing the subtotal on print and not clearing the grand total on print allows the subtotal register to
totalize sub-sets of weighments and to be reset while the grand total continues to track the grand
total of weight. Choose to Clear ST on Print or not by selecting enabled or disabled in the selection
box.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-35
3.6.2.3.5. Convert Weight
The total registers always store weights in primary units. If Convert Weight is disabled, scale
weights other than primary units are not accumulated. If Convert Weight is enabled, then the weight
is converted to primary units, then accumulated.
3.6.2.4. ID
The ID function is a simple but powerful means of facilitating specific data input from the operator
or causing a specific action to take place. The prompt list may include up to 20 steps, each step
containing a command that determines the action the IND560x terminal will take when the step is
executed.
The sequence can be programmed to run once (for example, when initiated by the pressing the ID
softkey ) or to loop continuously until terminated. It can also be initiated and restarted
automatically, triggered by weight input from the scale. In either case, the operator works though a
prompted series of actions or data entries. For example, the operator could place a package on the
scale, be prompted to enter his or her name, enter a pre-set tare value, scan a bar code on the
package, and generate an automatic print before finally removing the package from the scale. The
printed data could include the net weight value, together with the scanned package information and
the operator’s name. The content and format of the printed information is determined by the template
assigned to the demand output.
3.6.2.4.1. ID Mode – None
If None is selected as the ID Mode, then the ID function is disabled.
3.6.2.4.2. ID Mode – Automatic
Automatic Mode triggers the ID sequence when a weight above the Threshold weight value is
placed on the scale and a no-motion state is detected. The sequence trigger rearms again after the
weight falls below the Reset weight value. Press the EXIT softkey to exit the sequence.
3.6.2.4.3. ID Mode – Softkey
In this mode, the ID softkey triggers the sequence, which either runs once (if Looping is
Disabled) or continuously (if Looping is Enabled). In either case, pressing the EXIt softkey exits
the sequence.
3.6.2.4.4. Configuring ID Sequence Steps
Once the Mode Parameters have been set, ID sequence steps can be viewed, created, edited or
deleted by pressing the VIEW TABLE softkey .
1. The ID Search screen displays, permitting the search results to be filtered by #, Type, Length or
Prompt.
2. Either select a field to search and enter a comparative operator (<, <=, =, >=, >, <>) and
alphanumeric information in the Data fields; or leave the wildcard asterisk (*) in the Data field
to view all existing steps.
3. Press the START SEARCH softkey to display the ID View screen. From here, ID steps can
be edited , created , erased and printed .

3-36 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
4. The # value is automatically assigned. It determines the order in which the steps will execute.
Steps may be inserted into an existing sequence by selecting an entry and pressing the NEW
softkey to define a new step. The newly created step will appear in the sequence before the
selected entry, and all subsequent # values will increase by one. To move a step in the
sequence, it must be deleted and re-entered in the appropriate position.
5. The ID Edit screen is accessed by pressing the NEW softkey or the EDIT softkey . The
parameters that can be set for each step depend on the Type of step selected are shown below:
Type Parameters Options
Alphanumeric
Numeric
Prompt Enter up to 20 characters as the text or prompt for the step to
display on-screen during the ID sequence.
Length
Enter numeric value (0 - 40) to constrain the length of the
response input after the prompt.
If length of 0 is entered, the prompt is shown without an entry field.
The step is then displayed as an instruction in the sequence. In this
case, ENTER must be pressed to move to the next step. During
entry, once the length value is met, the terminal will not accept
further input; incorrect entries can be deleted using the CLEAR key.
Clear Data
Disabled, Enabled. Permits the entered response (e.g., Operator’s
name) to be retained from one cycle to the next. If disabled, the
next time the field displays it contains the data entered during the
previous cycle, and pressing ENTER accepts the value and moves
to the next step.
If enabled, the entry field will be blank when viewed the next time.
Clear Tare
Print
Tare – Auto
None
Tare – Preset
Prompt See Prompt, above
Length See Length, above
Clear Data See Clear Data, above
6. As many as 20 steps can be created. They can be printed at any point using the PRINT softkey
. Return to the setup menu tree by pressing the EXIT softkey three times.
3.6.3. Discrete I/O
Discrete I/O setup screens include:
 Inputs
 Outputs
3.6.3.1. Inputs
The Discrete Inputs screen displays discrete input assignments, including the input assignment
address, polarity, and function. Only records with non-null values display.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-37
To clear the entire table, press the CLEAR softkey , then press OK to confirm the action. Login
must be at supervisor level or above to clear the table.
Press the UP and DOWN navigation keys to scroll up and down the screen in order to view all of
the possible discrete input assignments.
To modify or add discrete inputs:
1. Press the UP and DOWN navigation keys to select (highlight) a discrete input.
2. Press the EDIT softkey to open the setup screen for editing an input assignment or press the
INSERT softkey to open the setup screen to create a new input assignment.
3. Enter the input assignment address. The input address is shown as [x.y.z] where x indicates
the input location, y indicates the slot address of the I/O option, and z indicates the input
position. The input address digits are:
Location – The first digit represents whether the I/O is local (0) or remote (1–3).
Slot Address – The second digit will always be a 1 for the IND560x internal I/O and a 0 for the
remote I/O (ARM 100).
Position – The third digit refers to the position (1–4) of the discrete input option (internal or remote)
that is being assigned to a function.
Valid address numbers are:
Local – 0.1.1, 0.1.2, 0.1.3, 0.1.4
Remote #1 – 1.0.1, 1.0.2, 1.0.3, 1.0.4,
Remote #2 – 2.0.1, 2.0.2, 2.0.3, 2.0.4,
Remote #3 – 3.0.1, 3.0.2, 3.0.3, 3.0.4
Examples:
Address 0.1.1 = Local discrete board , position 1.
Address 1.0.3 = Remote address #1, position 3.
4. The inputs can be programmed to accept either a + True or – True polarity level as “ON”. Use
the Polarity selection box to select + True or – True.
5. Use the Assignment selection box to select an input assignment. Selections are:
 None  Disable Setup  Target Pause/Abort  Unit Switching
 Blank Display*  Enter  Target Start/Resume  Unit - primary
 Calibration Test  Print  Trigger 1  Unit - second
 Clear Tare  SmartTrac  Trigger 2  Unit - third
 Disable Keypad*  Tare  Trigger 3  Zero
* Blank Display and Disable Keypad are designed to be continuously and semi-continuously activated
by a simple press-and-hold or key and lock mechanism
6. Press the OK softkey to accept the entry.
7. Press the ESCAPE softkey to return to the Discrete Inputs screen.

3-38 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
8. Press the DELETE softkey to delete an input assignment.
3.6.3.2. Outputs
The Discrete Outputs screen displays discrete output assignments, including the output assignment
address and function. Only records with non-null values display.
To clear the entire table, press the CLEAR softkey , then press OK to confirm the action. Login
must be at supervisor level or above to clear the table.
Press the UP and DOWN navigation keys to scroll up and down the screen in order to view all of
the possible discrete output assignments.
To modify, add, or delete discrete outputs, follow the same procedures described for inputs. Valid
address numbers are:
Local – 0.1.1, 0.1.2, 0.1.3, 0.1.4, 0.1.5, 0.1.6
Remote #1 – 1.0.1, 1.0.2, 1.0.3, 1.0.4, 1.0.5, 1.0.6
Remote #2 – 2.0.1, 2.0.2, 2.0.3, 2.0.4, 2.0.2, 2.0.6
Remote #3 – 3.0.1, 3.0.2, 3.0.3, 3.0.4, 3.0.5, 3.0.6
Examples:
Address 0.1.1 = Local discrete board , position 1.
Address 1.0.3 = Remote address #1, position 3.
The selections for output assignments are:
 None  Comparator 3  Feed  Over Zone
 Center of Zero  Comparator 4  Motion  Tolerance - OK
 Comparator 1  Comparator 5  Net  Under Zero
 Comparator 2  Fast Feed  Over Capacity  Under Zone
3.6.4. TaskExpert
The TaskExpert screens allow you to set up a TaskExpert program, and to access custom setup
screens defined by your TaskExpert application.
Note that TaskExpert setup screens are not accessible unless the TaskExpert option is
enabled in the terminal. TaskExpert is enabled in the terminal by installing the TaskExpert
Application Software key (iButton).
3.6.4.1. Start
The TaskExpert Start screen allows you to list your TaskExpert programs and associate each of
them to a Task number. Each task can be set to run automatically on startup or after leaving Setup,
or started manually using the Task List , or the Task 1, Task 2 or Task 3 softkeys.
Press the CLEAR softkey to reset the table.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-39
To modify the TaskExpert Start list:
1. Use the arrow keys to select the task to be edited or deleted. With the record selected:
Press the EDIT softkey to modify a record. Note that the Task number cannot be modified.
Press the DELETE softkey to delete a record.
2. To create a new task press the NEW softkey . A TASK EXPERT EDIT screen will display.
3. The Task number will be assigned automatically. Enter the name of the TaskExpert file in the
File Name field.
4. Set Auto Start to Enabled or Disabled. If Enabled is selected, the task will run each time the
indicator is powered up or when Setup is exited.
5. Set Manual Start to Enabled or Disabled. This will cause the program to run when selected and
started.
6. Press to save changes or to exit without saving.
3.6.4.2. Custom Setup
The Custom Setup menu is defined by the TaskExpert application. For details, refer to the TaskExpert
Reference Manual, provided with TaskExpert.
3.6.5. Reset
The Reset setup screen resets setup values to factory default settings for the Application branch
setup. Login must be at supervisor level or above to perform a reset.
3.6.5.1.1. Application Reset
To initiate a reset, press the OK softkey . If the reset was successful, a verification message that
reads “Reset Successful” displays If the reset was not successful, an error message that reads
“Reset Failure” displays. If the reset fails, try to initiate the reset again. If the reset continues to fail,
contact a local METTLER TOLEDO representative for assistance.
Press the ESCAPE softkey to exit without resetting.
Application reset does NOT include the reset of information in Alibi Memory or tables. This data
can only be reset by selecting Maintenance, Reset All.

3-40 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
3.7. Terminal
Setup screens for Terminal setup include:
 Device*  Transaction counter
 Display*  Users
 Region  Softkeys
Leaf nodes that are marked with an asterisk (*) above
will have different parameters depending on the
connected scale type (analog or IDNet).
Use these setup screens to configure:
 External battery compatibility and power saving options  Language
 Trax EMT™ ID fields  Transaction counter settings
 SmartTrac display settings  Users
 Time and date format and settings  Softkeys
3.7.1. Device
Use the Device setup screens to enable harsh/analog terminal compatibility with the external NiMH
battery pack and to enter data for the three terminal IDs and terminal serial number.
3.7.1.1. Battery Operation
This parameter is only shown in IND560x models with analog main boards. The default value is
Disabled. Setting this parameter to Enabled will cause the IND560x to use the Load Cell parameters
established at Scale > Type. Enabling Battery Operation also unlocks an Auto Power Off parameter
at Terminal > Device.> Display.
3.7.1.2. Terminal ID
Up to three terminal IDs can be entered on the Device setup screen. When the Terminal ID text box
is selected, the softkeys become alpha keys. Use the alpha keys to enter up to three terminal IDs.
Press the ENTER key to accept the ID entered. These ID fields display in the information recall when
the SYSTEM INFORMATION softkey is pressed after the RECALL softkey .
3.7.1.3. Serial Number
When the Serial Number text box is selected, the softkeys become alpha keys. Use the alpha keys
to enter the terminal’s serial number. The serial number is located on the back of the terminal. Press
the ENTER key to accept the serial number entered.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-41
3.7.2. Display
Use the Display setup screen to set the screensaver timeout and the SmartTrac size settings.
3.7.2.1. Screensaver
Enter the number of minutes (up to two digits) that must elapse with no scale motion and no
keypad activity before the screensaver is shown (replaces the view on the display screen). If motion
is detected or any key is pressed, the screen saver automatically exits and its time is reset. A
keystroke used to exit the screen saver mode is ignored for all other purposes.
A setting of 0 in the Screensaver field will disable the screensaver.
3.7.2.2. Backlight
The IND560x is equipped with a backlight for improved viewing in some conditions. The backlight
can be enabled or disabled. If the backlight is enabled, an additional selection for Backlight
Timeout appears.
3.7.2.3. Backlight Timeout
Allowable values for Backlight Timeout are 0 (always on) to 99 minutes. The default setting is 10
minutes.
3.7.2.4. Auto Off Timer
If the Battery Operation parameter at Terminal > Device has been Enabled, users can choose to
activate an auto-power off feature that will help to optimize single charge battery life. The Auto Off
Timer parameter monitors scale motion and keypad operation. If no keypad activity occurs within
the selected time value, the terminal automatically turns itself off. This auto-shutdown will occur
with no warning to the operator.
Choices are:
 Disabled (default)
 10 minutes
 30 minutes
 60 minutes
3.7.2.5. SmartTrac Size
SmartTrac refers to a graphical display visualization used to represent measured values. The
display is either a bar graph (if material transfer is selected as the target mode) or an over/under
display (if over/under is selected as the target mode).
The graphical display varies in terms of the amount of data displayed. While the SmartTrac graphic
remains constant in size, the “small” setting includes minimal information, while the “large” setting
displays the description as well as the target and limit values.
Examples of different sizes of SmartTrac displays are as follows:

3-42 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
Material Transfer Mode Over/Under Mode
Small
Medium
Large
Figure 3-16: SmartTrac Display Sizes, Material Transfer and Over/Under Modes
Use the SmartTrac Size selection box to select a setting for the SmartTrac graphical display.
Selections include None, Large, Medium, and Small.
If this parameter is set to None, SmartTrac will not display when the SmartTrac softkey is
pressed at the home screen. Instead, a message will display: “This Function is disabled.”
3.7.2.6. Rate Display
If a Rate is configured at Scale > Rate, it may be displayed on the home screen or not, depending
on whether this value is set to Enabled or Disabled. The option to display Rate does not appear
when SmartTrac is set to any value other than None.
3.7.3. Region
The region setup screens enable configuration of:
Time and date format
Time and date settings
Language
3.7.3.1. Format Time & Date
Selection boxes on this setup screen enable formatting of:
3.7.3.1.1. Time Format
12:MM (12-hour clock with hour and minutes displayed)

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-43
12:MM:SS (12-hour clock with hour, minutes, and seconds displayed)
24:MM (24-hour clock with hour and minutes displayed)
24:MM:SS (24-hour clock with hour, minutes, and seconds displayed)
3.7.3.1.2. Date Format
DD MM YY (Two-digit day, month, year)
DD MMM YYYY (Two-digit day, three-character month, four-digit year)
MM DD YY (Two-digit month, day, year)
MMM DD YYYY (Three-character month, two-digit day, four-digit year)
YY MM DD (Two-digit year, month, day)
YYYY MMM DD (Four-digit year, three-character month, two-digit day)
3.7.3.1.3. Date Field Separator
/ (slash)
- (dash)
. (period)
(space)
None
3.7.3.2. Set Time & Date
Enter the hour, minutes, day, month, and year on this setup screen’s text fields and selection
boxes. The terminal automatically adjusts the date for a leap year, and a battery backup maintains
the time and date settings in the event of a power outage.
Manual setting of the time is necessary for daylight savings time adjustments.
3.7.3.2.1. Hour
Use the numeric keypad to enter the hour in the Hour field text box. Use the AM/PM selection box to
select AM or PM. The AM/PM selection box only displays if the time format is set to 12:MM or
12:MM:SS on the Format Time & Date setup.
3.7.3.2.2. Minutes
Use the numeric keypad to enter the minutes in the Minutes field text box.
3.7.3.2.3. Day
Use the numeric keypad to enter the day in the Day field text box.
3.7.3.2.4. Month
Use the Month selection box to select the month.
3.7.3.2.5. Year
Use the numeric keypad to enter the year in the Year field text box.

3-44 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
3.7.3.3. Language
Use the Language setup screen to specify the language for terminal operations.
3.7.3.3.1. Display Messages
Use the Display Messages selection box to select the language for messages that display on the
terminal. Choices are:
 English
 French
 German
 Italian
 Spanish
 Custom
 Custom2
3.7.3.3.2. Keypad Selection
Use the Keypad Selection selection box to select the language for the keypad characters that display on
the terminal. This selection determines which international characters are available in the alphabetic
data entry softkeys. Choices are:
 Dutch
 English
 French/German
 Nordic/German
 Spanish/Italian/Portuguese
 Russian
3.7.4. Transaction Counter
The transaction counter is a seven-digit register that tracks the total transactions that are completed on
the terminal. When the value reaches 1,500,000, the next transaction causes a roll-over to 0000001.
Use the Transaction Counter setup screen to configure transaction counter operations.
3.7.4.1. Transaction Counter
Use the Transaction Counter selection box to enable or disable the transaction counter.
3.7.4.2. Counter Reset
Use the Counter Reset selection box to enable or disable manual resetting of the counter.
3.7.4.3. Next Transaction
The value for the counter for the next transaction displays in the Next Transaction field. The RESET
COUNTER softkey can be assigned to the home screen, giving direct access to the Transaction
Counter screen. If Counter Reset is enabled, the counter can be set to 0000000001 by pressing the
RESET softkey. A value greater than one can also be entered manually in the Next Transaction field.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-45
3.7.5. Users
The IND560x terminal is pre-configured with two usernames: “admin” and “anonymous”. Neither of
these usernames has a password. Without passwords, there will be no security challenge for entering
setup and making changes. If a password is entered for “admin”, a login will be required before
changes can be made in setup. These two default usernames cannot be deleted but passwords may
be entered for them. All functions of the terminal will be available to all users until a password is
entered for the default admin user.
To modify and/or enter a password for a username:
1. Move focus to the username to be edited.
2. Press the EDIT softkey to access the User Edit screen.
3. Press the DOWN navigation key to view the second screen, which displays the Password and
Confirm Password fields.
4. Enter the desired password in the Password and Confirm Password fields.
5. Press the OK softkey to accept the password as entered.
6. Press the ESCAPE softkey to exit without saving the password.
Passwords are case-sensitive.
All numbers and all characters available in the alpha keys are valid for use in passwords.
Be sure to store a record of the password in a safe place. Without the correct password access
to the setup menu will not be possible.
To enter a new username and password:
1. Press the INSERT softkey to access the User Edit screen.
2. Enter the desired username in the Username field.
3. Use the Access selection box to assign the appropriate access level to the user. The following
access levels are available:
 Administrator – An administrator has unlimited access to all set-up parameters in the
terminal. There can be multiple administrator users.
 Maintenance – Next lower level of access.
 Supervisor – Next lower level of access.
 Operator – Lowest level of access.
4. Press the DOWN navigation key to view the second screen, which displays the Password and
Confirm Password fields.
5. Enter the desired password in the Password and Confirm Password fields.
6. Press the OK softkey to accept the username and password as entered.
7. Press the ESCAPE softkey to exit without saving the username and password.
8. Press the DELETE softkey to delete a username from the table on the Users screen.
Usernames "admin" and "anonymous" cannot be deleted.

3-46 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
9. To restore the factory default setting (by deleting all users except the “Admin” and “Anonymous”
users and restoring the passwords to null), press the CLEAR softkey and then OK to
confirm the action.
3.7.6. Softkeys
Add or rearrange the softkeys on the terminal's main menu on the softkeys setup screen. Appendix
E, Softkey Mapping, provides a more detailed explanation of the softkeys.
When the softkeys setup screen opens, focus is on the softkey position numbers located above the
icons.
Two softkeys, INFORMATION RECALL and SETUP , are automatically placed in the softkey
setup. The default setting for the placement of these softkeys is in positions 9 and 10.
The INFORMATION RECALL and SETUP softkeys must always be assigned a position. They can be
moved or multiple copies of them can exist but there must always be at least one appearance of
them. They cannot be deleted. All other softkeys may be added or deleted as desired.
Adding a softkey to the home pages of the terminal does not automatically enable the softkey’s
function. Most softkeys must also be enabled in setup too. For instance, adding the UNITS
SWITCHING softkey does not automatically enable units switching – the units must also be
enabled in the Scale branch of setup. If a softkey has been added in setup but does not appear on
the home page, check that the setup parameters for that function have been enabled.
The following softkeys are available for assignment to the home page:
 None  Recall Info*  Target Start
 Alibi Memory  Repeat Print  Target Table
 Calibration Test  Reports  Task 1**
 Comparators  Reset Transaction Counter  Task 2**
 Custom Trigger 1  Setup*  Task 3**
 Custom Trigger 2  SmartTrac  Task List**
 Custom Trigger 3  Tare Table  Time & Date
 ID  Target  Unit Switching
 MinWeigh  Target Control  x10 Display
* Cannot be removed from home screen or edited.
** Only available when TaskExpert™ or an application is installed.
Press the UP, DOWN, LEFT and RIGHT navigation keys to navigate among the softkey position
numbers. Softkeys can be added, removed, and positioned using the softkeys:
Edit
Changes the softkey in the selected position to another softkey or to none,
which leaves the softkey position blank. Editing a blank position does not move
the position of following softkeys.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-47
Insert Inserts a softkey into a selected position. All other softkeys located at or after
that position increase position number by one.
Delete Deletes a softkey in a selected position. All other softkeys located at or after that
position decrease position number by one.
Clear Clears all softkey assignments except INFORMATION RECALL and SETUP
softkeys. These will be shown in positions 1 and 2 respectively.
3.7.7. Reset
The Reset setup screen resets setup values to factory default settings for the Terminal branch of
setup.
3.7.7.1.1. Terminal Reset
To initiate a reset, press the OK softkey . If the reset was successful, a verification message that
reads “Reset Successful” displays If the reset was not successful, an error message that reads
“Reset Failure” displays. If the reset fails, try to initiate the reset again. If the reset continues to fail,
contact a local METTLER TOLEDO representative for assistance.
Press the ESCAPE softkey to exit without resetting.
3.8. Communication
Setup screens for Communication setup include:
 Templates  Connections  Network
 Reports  Serial  PLC
Use these setup screens to configure:
 Input and Output Template Setup
 Report Structure
 Connection Assignments
 Serial Port Parameters
 Network Parameters, including Email Alert settings
 PLC Interfaces
3.8.1. Templates
The IND560x terminal utilizes five output templates (1,000 bytes in
size) to define the format of data output following a weighing operation.
An input template is also available to receive a string input (such as
from a bar code gun) and use it as data entry for ID 1, Tare, Tare ID or

3-48 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
Target ID. When the Fill-560 Application Software is installed, the input template can also be
assigned for Target Weigh-In or Target Weigh-out.
A template strings setup screen is also available to configure strings of characters that are
frequently used in templates.
3.8.1.1. Input
Use the Input Template to remove extra characters from an input string of data. Setup parameters
include:
Preamble Length – Defines how many characters are skipped at the beginning of an input string
before the desired data.
Data Length – Defines the maximum length of a string. All characters beginning after the preamble
through the data length selection are used as the input.
Postamble Length – Defines the number of characters (before but not including the termination
character) that will be stripped off the data string. All other data from the preamble length to the
termination character minus the postamble length are used as the input string. When using an input
that is always the same fixed length, this field remains blank.
Termination Character –+ Used to signal the end of the string input. The termination character can
be any ASCII control character. If None is selected, the three-second timeout feature terminates the
entry. In addition to None, selections available from the selection box include:
SOH STX ETX EOT ENQ ACK BEL BS
HT LF VT FF CR SO SI DLE
DC1 DC2 DC3 DC4 NAK SYN ETB CAN
EM SUB ESC FS GS RS US
Assignment – Determines how the input data should be used. Choices are:
ID1 Input populates the ID1 Shared Data variable.
Keypad Used in conjunction with ID Mode – data that would normally be entered using the
numeric keypad or on-screen alpha keys may be scanned in during an ID Mode
sequence.
Tare Enters data as a preset tare value
Tare ID Uses value as an ID lookup in the tare table
Target ID Uses value as an ID lookup in the target table
Target
Weigh-in
Uses value as an ID lookup in the target table and assigns it as the Active Weigh-in
target (Fill-560 feature only)
Target
Weigh-out
Uses value as an ID lookup in the target table and assigns it as the Active Weigh-out
target (Fill-560 feature only)
There is also a three-second timeout feature that tracks the amount of time between characters. If
this three-second time is exceeded, the string is considered terminated.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-49
3.8.1.2. Output
The Output Template setup screen enables configuration of the output data formats and a header or
footer for the Repeat Print function.
To add the “DUPLICATE” designation to a repeat print of an output template, select either Header or
Footer from the Repeat Print Field dropdown menu highlighted in Figure 3-17.
Figure 3-17: Repeat Print Field Setup
To view elements in a template or create a new template, select the desired template from the
Template selection box. Five templates are available (Template 1–Template 5).
Press the VIEW TABLE softkey to access the selected template’s view screen, which lists each
element configured for the template.
Press the COPY softkey to open the copy screen, which enables the elements from one
template to be copied to another template.
Press the CLEAR softkey to clear all elements from the selected template. A warning screen
displays as a precaution.
Press the EXIT softkey to return to the menu tree.
Each element is defined by the data type, actual data, justification and length. Elements may be
edited, added (inserted), or deleted individually. A leading zero fill function is available for certain
fields.
3.8.1.2.1. Editing
1. Press the UP and DOWN navigation keys to select a record to edit.
2. Press the EDIT softkey to open the setup screen for the record where edits can be made.
The element number cannot be edited.
3. Select the element type from the Type selection box. Four element types are available:
CR/LF – A combination carriage return and line feed.
SDName – The name of a Shared Data variable from the terminal.
Special Character – Control characters from a standard ASCII chart. Refer to Appendix G, ASCII
Standard and Control Characters.
String – A string of characters to print.

3-50 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
Press ENTER to confirm the element type selection. Depending on which type is selected,
additional configuration options will appear.
4. Press the OK softkey to return to the template setup screen when editing of the element is
complete.
5. Press the ESCAPE softkey to return to the template setup screen without saving the
element edits.
Template Editing Options
Table 3-3 lists the editable fields that appear for each Element Type and Format selection. Fields
that are not available are indicated with a dash (–).
Table 3-3: Template Element Editing Options
Element
Type Data Repeat Count Format Length Leading Zero Fill
CR/LF  Blank**   
SDNAME Blank*
(Refer to Table 3-4) 
Default  
Centered, Left Blank** 
Right Blank** Enabled/Disabled
Special
Character ASCII character list    
String Blank* Blank**
Default  
Center, Left, Right Blank** 
* Alphanumeric entry field
** Numeric entry field
Repeat Count
A numeric value that sets the number of times the Data will repeat. For example, if the string data
entered is an asterisk (*), a repeat count value of 20 would produce a row of 20 asterisks:
********************.
Format & Length
The Format & Length options are Default, Centered, Left, and Right. These describe the justification
of the data within the length of the field.
The Default value is fixed, and Length cannot be defined here.
For SD name and String data, the Centered, Left and Right formats are accompanied by a Length
field. After choosing the format, focus moves to the Length entry box. This value determines how
long the element will be – the data will be positioned within a string of the length entered. Refer to
the Notes on Format & Length, below.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-51
Leading Zero Fill
In the case of Right-formatted SDName data, Leading Zero fill may be enabled or disabled.
Normally, Leading Zero Fill will be disabled, which means spaces will be used to fill in leading
non-significant data. If enabled, Leading Zero Fill will fill in any leading spaces of the data with
zeroes. The following examples of 4 characters in output data defined as 8 characters long show
the use of Leading Zero Fill:
Leading Zero Fill disabled:
Leading Zero Fill enabled:
Notes on Format & Length
If data for an element includes more characters than the value entered for Length, the output data
will be cut off. For example, if the number of characters entered under format on the Output
Template Edit Screen is six, and data for a string that contains eight characters is entered, only six
of the characters from the data string will print. If data for a string that contains four characters is
entered, all of the characters from the data string will print.
The characters and alignment of the printed data are defined by the selected format:
Example 8-character string of ABCDEFGH in a 6-character element:
Center: Left: Right:
B C D E F G A B C D E F C D E F G H
Prints the middle six characters
from an eight-character string –
drops first and last characters.
Prints the first six characters from
an eight-character string – drops
the last two characters.
Prints the last six characters from
an eight-character string – drops
the first two characters.
Example 4-character string of ABCD in a 6-character element:
A B C D A B C D A B C D
Space Space Spaces Spaces
Prints all four characters from a
four-character string in a centered
position, leaving a space at the
beginning and a space at the end.
Prints all four characters from a
four-character string in a leftjustified position, leaving two
spaces at the end.
Prints all four characters from a
four-character string in a rightjustified position, leaving two
spaces at the beginning.

3-52 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
In Table 3-4, where dashes appear in the SDName (e.g. sd--08), this means that there are more
than one instance of the share data variable. The instance number appears in place of the dashes
(e.g. sd0308, for the third comparator’s target comparison operator). For a full list of Shared Data
fields, please refer to the IND560 Terminal Shared Data Reference (document # 64058660) found
on the documentation CD.
Table 3-4: Example Shared Data Names for Use in Templates
IND560x Data Field SDName Length
(A/N = alphanumeric)
Terminal
Scale ID cs0103 20 A/N
Base Serial Number ce0138 13 A/N
Terminal Serial Number xs0105 13 A/N
Terminal ID#1 xs0106 20 A/N
Terminal ID#2 xs0107 20 A/N
Terminal ID#3 xs0108 160 A/N
Current Date xd0103 11 A/N
Current Time xd0104 11 A/N
Transaction Counter Value xp0101 9 A/N
Grand Total Value tz0101 12 A/N
Grand Total Count tz0102 9 A/N
Subtotal Value tz0103 12 A/N
Subtotal Count tz0104 9 A/N
ID Mode Prompts 1-20 pr0131-pr0150 20 A/N
ID Mode Prompt Responses 1-20 pa0101-pa0120 51 A/N
Message 01, 02, etc. aw0101, aw0102.. 100 A/N
Templates 1 - 5* pt0101 pt0105 As programmed
Template Strings 1–20 pt0111  pt0130 51 A/N
Targets
Target Description sp0101 20 A/N
Target Value sp0105 12 A/N
Target Units sp0120 3 A/N
Fine Feed sp0110 12 A/N
Spill Value sp0109 12 A/N
+ Tolerance Value (weight) sp0111 12 A/N
 Tolerance Value (weight) sp0112 12 A/N
+ Tolerance Value (%) sp0114 12 A/N
 Tolerance Value (%) sp0115 12 A/N
Comparator names 1-5 sd--01 21 A/N

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-53
IND560x Data Field SDName Length
(A/N = alphanumeric)
Comparator target coincidence
values 1 - 5 sd--05 12 A/N
Comparator target comparison
operators 1-5 sd--08 5 A/N
Comparator second weight range
values 1-5 sd--09 12 A/N
Comparator second weight range
comparison operators 1-5 sd--10 5 A/N
Weights
Scale Mode (Gross/Net) ws0101 1 A/N (G or N)
Tare Source Description ws0109 2 A/N “T<space>”, or “PT”
Displayed Tare Wt. ws0110 12 A/N
Displayed Gross Wt. wt0101 12 A/N
Displayed Net Wt. wt0102 12 A/N
Displayed Wt. Units wt0103 3 A/N
3rd Weight Unit Gross Weight wt0104 13 A/N
3rd Weight Unit Net Weight wt0105 13 A/N
3rd Units wt0106 3 A/N
Displayed Rate wt0108 13 A/N
Rate Time Units cs0107 1 A/N
Rate Weight Units cs0108 1 A/N
Custom Unit Name cs0112 12 A/N
Custom Unit Conversion Factor cs0113 12 A/N
* Using a template field code within another template will insert the entire template into the new
template.
Printing Multiple Weight Units
Note that if a third weight unit is defined, it can be named in a template so that it appears in the
output data together with the displayed unit. However, if the third unit is currently displayed on the
terminal, it will appear twice in the data output. Therefore, if the output is to include both primary
and 3rd weight units, ensure that the primary unit data is displayed when the data output is initiated.
Secondary Units are not available for multiple unit printing.
Inserting
Press the UP and DOWN navigation keys to select a location in the list of elements to insert a new
element. Press the INSERT softkey to open the setup screen to create a new record. Configure the
parameters as described under Editing. All following elements are moved down one position.

3-54 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
Deleting
Press the UP and DOWN navigation keys to select an element to delete. Press the DELETE softkey
to delete the element.
Test Printing
Press the TEST PRINT softkey to perform a test print of the template being configured.
3.8.1.3. Strings
The Template Strings setup screen defines strings of characters that are frequently used in template
messages. Template strings can be viewed, edited, inserted, deleted, or printed.
To view template strings:
1. Use the Search Field selection box to select ID or String and enter information in the Data field
to limit the search, or do not enter any search limits to view all template strings.
2. Press the START SEARCH softkey . The String Search View screen displays the search
results. Only records with non-null values display. Records are ordered by ID, with the lowest
ID number shown first.
3. Press the UP and DOWN navigation keys to scroll up and down the screen in order to view all
strings listed.
Editing
Note that not all keyboard characters are available when entering information using the
IND560x keypad. However, characters such as ~ and | can be included in a template created
using the InSite software. They will be recognized by the IND560x when the template is
downloaded, will display correctly, and will be included in the template when it is printed.
The [, ] and {, } characters have been added to the list of characters available from the
terminal softkeys. These characters can be found in firmware versions 3.02 and higher.
1. Press the UP and DOWN navigation keys to select (highlight) the string record to be edited.
2. Press the EDIT softkey to open the record for editing. The edit screen will be shown with
focus on the “Text” label. This is the only editable field on the screen. The ID number cannot be
changed.
3. Press ENTER to access the text field. The alpha keys will display.
4. Use the alpha keys and the numeric keypad to edit the string. A string can be a maximum of
50 characters.
5. Press ENTER when complete.
6. Press the OK softkey to accept the edited record and return to the String Search View
screen.
7. Press the ESC softkey to return to the String Search View screen without accepting the
changes to the string.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-55
Inserting
1. Press the INSERT softkey to create a new record. An edit screen will be shown with “ID” and
“Text” labels.
2. Press the UP and DOWN navigation keys to select the field to enter.
3. Press ENTER to access the field. When entering the ID field, the value shown is the first unused
ID.
4. Use the numeric keypad to enter an ID number from 1 to 20.
5. Press ENTER when complete. If the ID entered already exists, an error display will be shown.
The error must be acknowledged and a new ID entered.
6. If the TEXT field is entered, the alpha keys will display. Use the alpha keys and the numeric
keypad to edit the text string.
7. Press ENTER when complete.
8. Press the OK softkey to accept the new record and return to the String Search View screen.
9. Press the ESC softkey to return to the String Search View screen without accepting the new
record.
Deleting
Press the DELETE softkey to delete a template string from the list.
Printing
1. Press the PRINT softkey to print the list.
2. Press the EXIT softkey to return to the String Search Screen.
3. Press the EXIT softkey again to return to the menu tree.
3.8.2. Reports
The reports setup screens enable configuration of the structure of tare and target table reports that
are generated by the IND560x terminal. The reports setup screens include:
 Format
 Tare Table
 Target Table
3.8.2.1. Format
Use the selection box for the Format field to select the width of the reports
 Narrow (40) – 40 character-wide reports
 Wide (80) – 80 character-wide reports
3.8.2.1.1. Header
The Header field specifies the number of blank lines (CR/LF) to be placed at the start of each report.

3-56 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
3.8.2.1.2. Title
The Title selection box enables a default title line to be printed at the top of the report. Time and date
will print in the format selected at Terminal > Region > Format Time & Date.
3.8.2.1.3. Record Separator
A repeated character may be selected as a separator between printed records in the report. This step
selects the character to be used. The character choices in the selection box are:
None (no separator between records) = (equal symbols)
* (asterisks) CR/LF (blank line)
- (dashes)
For example, if * (asterisks) is selected, the resulting line separator will appear as follows:
*************************************************************************
3.8.2.1.4. Footer
The Footer field specifies the number of blank lines (CR/LF) to be placed at the end of each report.
3.8.2.2. Tare Table
Use this setup screen to select which fields in the Tare Table will print when a Tare Table Report is
printed. The ID field is always printed; it cannot be disabled. Fields that can be enabled or disabled
include:
 Tare  n (number of transactions)
 Description  Total (accumulated total tare weights)
3.8.2.3. Target Table
Use this setup screen to select which fields in the Target Table will print when a Target Table Report
is printed. The ID field is always printed; it cannot be disabled. Fields that display and that can be
enabled or disabled vary depending on how the Target Table is configured (as described in the
Application branch under Memory setup).
Fields available include:
 Description  +/- Tolerances  Fine Feed
 Target  Spill  +/- Limits
If the Target mode is disabled, this branch cannot be expanded.
3.8.3. Connections
The IND560x comes standard with a single, intrinsically safe serial port (COM1). Additional ports
are available by:
 Installing the current loop or fiber optic interface and using COM4 and/or COM5 to
communicate with an ACM500 safe area communication module. When using the ACM500

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-57
module with the Ethernet/COM2/COM3 option installed, serial ports COM2 and COM3 and
Ethernet connections Ethernet1, Ethernet2, Ethernet3 and EPrint become available.
NOTE: Always power up the ACM500 before turning on the power to the IND560x. This will
ensure that the IND560x will communicate correctly with the ACM500 and recognize all
installed options.
 Installing the intrinsically safe current loop interface option into the IND560x and using COM4
and/or COM5 directly. In this application, the COM4 and COM5 ports can be used to
communicate with the ACM200 safe area communication module or with an IND226x used as
remote display.
 Installing the fiber optic interface option into the IND560x and using COM4 and/or COM5
directly. In this application, the COM4 and COM5 ports can be used to communicate with the
legacy dual channel fiber optic converter, A100, 8624, or directly to another IND560x or Puma
terminal used as a remote display.
3.8.3.1. COM4 and COM5
As explained above, the COM4 and COM5 communication ports can be used as standard serial
ports directly or can be dedicated to communicate with the ACM500 module. The first decision to
make regarding connections to the IND560x is determining how these two ports will be used. The
Connections screen (Figure 3-18) shows the connection options for COM4 and COM5.
NOTE: The COM4/COM5 Connection page is visible in setup even if NO interface board has
been installed into the IND560x. If NO interface board is installed, the assignments for COM4
and COM5 must be “Disabled”.
Figure 3-18: COM4 and COM5 Setup
3.8.3.1.1. To edit the COM4 or COM5 connection
1. Press the UP and DOWN navigation keys to select (highlight) the appropriate port, COM4 or
COM5.
2. Press the ENTER key to edit the desired port.
3. Use the UP and DOWN keys to choose a connection type from the selection box.
4. Press the ENTER key to accept and save the connection parameter.

3-58 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
Table 3-5: Available COM4/COM5 Port Connections
Port Connection Function
COM4
Disabled Port will not be used
Standard Port will be used directly with ACM200, Fiber Optic converter or IND226x
ACM500 Port will be used to communicate with ACM500 for COM2, COM3 or PLC connectivity
COM5
Disabled Port will not be used
Standard Port will be used directly with ACM200, Fiber Optic converter or a remote IND226x
ACM500 Port will be used to communicate with ACM500 for Ethernet TCP/IP connectivity
The correct selection depends on what the IND560x will be communicating with over COM4 and/or
COM5. Table 3-6 provides an overview of what the COM4 and COM5 settings should be for the
various peripheral communication devices that work with the IND560x.
Table 3-6: COM4/COM5 Setup for Various Communication Modules
Peripheral Communication
Device Correct COM Connection Application Notes
None Disabled No communication intended over
COM4/COM5
ACM200 Module
(Communication available
over current loop interface
only)
COM4 = Standard
and/or
COM5 = Standard
The ACM200 provides a single RS-232
port in the safe area. Two ACM200
modules can be supported
simultaneously, one each on COM4 and
COM5.
Legacy Fiber Optic Converter
0964-0043
0964-0052
0964-0058
0964-0059
0964-0060
COM4 = Standard
and/or
COM5 = Standard
*1 or 2 communication channels may be
required. Dependant on legacy setup.
The FO converter supports one or two
channels of RS-232 or 20mA
communication in the safe area. This
selection would be used when replacing
a Puma, ID3sTx, 8141 or 8525 terminal
that is communicating to a fiber optic
converter.
IND226x (Communication
available over current loop
interface only)
COM4 = Standard
and/or
COM5 = Standard
The IND226x with Interface-Remote
installed can be used as a remote
display inside the hazardous area. Only
one of the two available ports is required
for this functionality.
ACM500
COM4 = ACM500
and/or
COM5 = ACM500
*Only 1 channel required in some cases.
The ACM500 can provide PLC and
COM2/COM3/Ethernet communications
in the safe area. Refer to the Installation
Manual for details of which port is
required for which option.
After the appropriate COM4 and/or COM5 connection has been made, proceed to the Connections
View setup screen for further port assignments by pressing the VIEW TABLE softkey .
If the port connection type has changed for COM4 or COM5, the IND560x will force a power cycle
so that the IND560x will correctly communicate with any safe area communication devices. The
IND560x will display a popup message indicating that a power cycle is about to take place (Figure
3-19). At this point, the operation cannot be cancelled. Setup must be entered again to view the
port assignments.

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-59
Figure 3-19: COM4/COM5 Power Cycle Message
Proceed in setup to Communications>Connections and again press the VIEW TABLE softkey to
enter the Connections View area where ports can be assigned.
3.8.3.2. Reconnect Mode
The communication protocol between the IND560x and the ACM500 requires that the ACM500 be
powered up first. If the ACM500 experiences a power cycle, the message shown in Figure 3-20 will
display. In the past when this occurred, users were required to acknowledge this pop-up message
by pressing the ENTER key. Users were then further required to manually power cycle the IND560x
to re-establish communication between the terminal and the ACM500.
Figure 3-20: ACM500 COM Error Message
A new setting in v3.04 firmware and higher provides the user with two options to re-establish
communication automatically in the event of an ACM500 power loss. In setup, Communication >
Connections includes a new setting called Reconnect Mode. Available settings include Manual and
Automatic.
Manual When Manual is selected, the operator will be presented with two options when the
ACM500 communication error (Figure 3-20) is shown on-screen. Pressing ESC will
acknowledge and clear the error message, but no further action will be taken by the
IND560x terminal. If OK is pressed, the error message is acknowledged and the
IND560x will complete a power cycle in order to re-establish communication with
the ACM500.
Automatic If Automatic is selected, the ACM500 communication error pop-up message will not
display on-screen and the user will not be given a choice of when the IND560x will
go through a power cycle. Instead, once the IND560x detects that the ACM500 is

3-60 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
back on line, the IND560x will automatically power cycle itself.
3.8.3.3. Communication Port Assignment
The Connections View setup screen (Figure 3-21) defines what type of communication will occur over
each port. If no connections are programmed, nothing will be available on the COM ports or Ethernet
ports.
Figure 3-21: Connections View Screen
Ports available for assignment can include the standard serial port COM1, optional COM2, COM3,
COM4 and COM5 serial ports, the optional Ethernet ports, and the optional E-print port. The
availability of all optional ports depends on the selections for COM4/COM5 on the Connections page
as well as optional interfaces installed in the ACM500.
Various COM4/COM5 setup combinations exist. Table 3-7 shows the available ports available for
additional assignment based on the initial setup selections for COM4 and COM5.
Table 3-7: Ports Available for Various COM4/COM5 Combinations
COM4 = Standard / COM5 = Standard
Available Port COM1, COM4, COM5
Ports Not Available COM2, COM3, Ethernet 1,2,3, Eprint
COM4 = Standard / COM5 = ACM500
Available Port COM1, COM4, Ethernet 1,2,3, Eprint
Ports Not Available COM2, COM3, COM5
COM4 = ACM500 / COM5 = Standard
Available Port COM1, COM2, COM3, COM5
Ports Not Available Ethernet 1,2,3
COM4 = ACM500 / COM5 = ACM500
Available Port COM1, COM2, COM3, Ethernet 1,2,3, Eprint
Ports Not Available COM4, COM5
The following functions are available via softkeys on the Connections View screen:
 Editing  Inserting

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-61
 Deleting  Clearing
To edit or insert connection assignments:
1. Press the UP and DOWN navigation keys to select (highlight) a connection assignment in the
list.
2. Press the EDIT softkey to open the setup screen for editing a specific connection
assignment or press the INSERT softkey to create a new connection assignment.
3. Parameters configured in the connections screen include the port and the type of input or output
assignment. Depending upon these selections the remaining fields will vary but could include
the trigger, the template to be sent, and if checksum will be sent or not.
4. Press the UP and DOWN navigation keys to move the focus to the field to be edited or added.
5. Press the ENTER key to select a field to edit or add.
6. Use the selection boxes for each field to select the port, associated assignment, and desired
settings for the connection. Available selections are indicated in Table 3-8, organized by Port
and Assignment.
Table 3-8: Available Parameters, by Port and Assignment
Port Assignment Trigger Template Checksum # of Nodes
COM1
COM2
COM3
COM4
COM5
ASCII Input, CTPZ Input,
Reports, SICS, Shared Data
Server*, Remote Display
Continuous Output Disabled, Enabled
Continuous Template Template 1…5
Demand Output,
Totals Report Scale, Trigger 1…3 Template 1…5
Remote Discrete I/O 0, 1, 2, 3
Ethernet 1
Continuous Output Disabled, Enabled
Continuous Template Template 1…5
Demand Output,
Totals Report Scale, Trigger 1…3 Template 1…5
Reports
Ethernet 2
Ethernet 3 Demand Output Scale, Trigger 1…3 Template 1…5
EPrint
Continuous Output Disabled, Enabled
Continuous Template Template 1…5
CTPZ Input
Demand Output Scale, Trigger 1…3 Template 1…5
Print
Client
Continuous Output Disabled, Enabled
Continuous Template Template 1…5
Demand Output Scale, Trigger 1…3 Template 1…5
Connection options notes:
Not all choices are available for all connection assignments. Only valid choices are shown in the
selection boxes on the IND560x terminal.

3-62 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
The Shared Data Server assignment is only available on COM1. Serial connection to the Shared Data
Server is only possible on COM1. This assignment allows a user to connect to the Shared Data
Server over COM1 without turning SW2-1 to the “on” position.
IMPORTANT: Appropriate signal barrier is required when connecting an non-intrinsically safe device,
such as a PC, to the intrinsically safe COM1 on the IND560x. Entity parameters must be confirmed
to ensure the appropriate barrier has been selected.
The Trigger field displays only when the assignment selection is Demand Output. The Trigger 1, 2,
and 3 selections for Trigger enable connection of a discrete input or a PLC command to trigger the
output of the selected template. The discrete selection must be programmed separately (as described
in the Application branch under Discrete I/O).
The Template field displays only when the assignment selection is Demand Output, Continuous
Template, or Totals Report. Only one template can be assigned to each connection.
The default template for the Totals Report is Template 5.
The Checksum field is available for continuous outputs only.
The SICS selection provides some Level 0 and Level 1 interface commands.
If usage conflicts occur, a pop-up message will display and the IND560x terminal will not allow the
conflicted assignment to be saved.
EPrint offers a method to access the demand or continuous output data directly through the Ethernet
port. Shared Data Server login and commands are not required to register for the data. The data
output string does not include any Shared Data Server response messages, and reflects only the
configured demand template data or the continuous output string. The EPrint port is accessible only
through the Ethernet interface’s secondary port, which can be configured in setup at Communication
> Network > Port.
The # of Nodes refers to the number of ARM100 Remote I/O modules that will be connected in series
to the IND560x.
7. Press the OK softkey to accept the connection parameters and return to the Connections
setup screen when editing or adding the connection assignment is complete.
8. Press the ESCAPE softkey to discard the connection parameters and return to the
Connections setup screen without saving the connection assignment edits or additions.
Press the DELETE softkey to delete a connection assignment from the Connections list.
To clear all connection assignments in the Connections list, press the CLEAR softkey then
press OK to confirm the action.
Press the EXIT softkey to return to the menu tree.
3.8.4. Serial
Serial communication setup screens provide access to the communication parameters for the serial
ports COM1, COM2, COM3, COM4 and COM5.
NOTE: The COM2 and COM3 ports are only shown if the Ethernet/Serial option board has
been installed in the ACM500.
NOTE: The COM4 and COM5 ports can only be changed if their connection setup selection
is “Standard”.
 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-63
If “Remote Discrete I/O” is selected as the port assignment at Communication >
Connections, the port parameters are automatically preset and cannot be changed.
3.8.4.1. COM1, COM2, COM3, COM4 and COM5
Use the COM1, COM2, COM3, COM4 and COM5 setup screens to configure the parameters for
serial ports.
3.8.4.1.1. Baud
Use the Baud selection box to set the baud rate for the serial port. Options are:
300 1200 4800 19200 57600
600 2400 9600 38400 115200
3.8.4.1.2. Data Bits
Use the Data Bits selection box to set the data bits to either 7 or 8 for the serial port.
3.8.4.1.3. Parity
Use the Parity selection box to set the parity to None, Odd, or Even for the serial port.
3.8.4.1.4. Flow Control
Use the Flow Control selection box to set the flow control to either None or XON-XOFF (software
handshaking).
After an XOFF command is received on a serial port configured for XON/XOFF handshaking,
data remaining in the port’s buffer will still be sent. Port buffers contain the following amounts
of data:
 COM1: 16 bytes
 COM2: 64 bytes
 COM3: 64 bytes
 COM4: 64 bytes (when set as “Standard”, not “ACM500”)
 COM5: 64 bytes (when set as “Standard”, not “ACM500”)
3.8.4.1.5. Interface
Use the Interface selection box to select the serial port interface. Selections include:
 RS-232
 RS-422
 RS-485
COM4 and COM5 are limited to RS-232 only.

3-64 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
3.8.5. Network
Network setup screens include Ethernet, Port, FTP and Email Alert.
3.8.5.1. Ethernet
Ethernet is available for TCP/IP transfer of data, shared data server access, connection via the
EPrint port, email alerts, FTP and saving and restoring configurations with InSite™ SL. The Medium
Access Control (MAC) Address cannot be edited; it is shown for information only. Setup for Ethernet
allows static Internet Protocol (IP) addressing or DHCP (Dynamic Host Configuration Protocol). If
the (Dynamic Host Configuration Protocol) DHCP Client setting is enabled, the IP Address, Subnet
Mask, and Gateway Address fields are assigned automatically by the network and become
read-only in the setup screens.
Use the UP and DOWN navigation keys to select the field to be edited. Press the ENTER key to edit
the field or selection choice. The Ethernet branch includes the following fields:
3.8.5.1.1. DHCP Client
The DHCP client can be enabled or disabled. If disabled, the IP address must manually be
assigned in the following fields. If enabled, the terminal will be assigned an IP address by the
network server. The IP address can then be viewed in the following fields.
3.8.5.1.2. IP Address
Enter the IP address (or view if DHCP Client is enabled) for the IND560x terminal. After each group
of digits has been entered, press ENTER to proceed to the next group. The default value for the IP is
192.68.0.1.
3.8.5.1.3. Subnet Mask
Enter the subnet mask (or view if DHCP Client is enabled) for the IND560x terminal. After each
group of digits has been entered, press ENTER to proceed to the next group. The default value for
the subnet mask is 255.255.255.0.
3.8.5.1.4. Gateway Address
Enter the gateway address (or view if DHCP Client is enabled) for the IND560x terminal. After each
group of digits has been entered, press ENTER to proceed to the next group. The default value for
the gateway is blank.
After entry is complete, press the EXIT softkey to return to the menu tree
3.8.5.2. Port
The Port setup screen displays the primary Ethernet port number and also allows a secondary port
number to be configured for the same Ethernet interface. The primary port is reserved for the Shared
Data Server access. The secondary port allows for either Shared Data Server access or EPrint
connections, if configured. The primary and secondary ports can be used concurrently.
3.8.5.2.1. Primary Port #
The Primary Port number for the Ethernet TCP/IP interface is fixed at 1701, and is displayed as a
read-only field in the Port setup screen (Figure 3-22). The Primary Port number can be modified
 

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-65
through a Shared Data write. Refer to the IND560 Shared Data Reference Manual found of the
Documentation CD for details.
Figure 3-22: Secondary Port Setup
3.8.5.2.2. Secondary Port #
If a valid, user-defined Secondary Port number is entered, the Shared Data Server will also be
accessible on a second TCP/IP port. However, if an EPrint connection is configured in setup at
Communication > Connections, Shared Data Server access is no longer be available on this port,
which will be used entirely for the EPrint demand or continuous types of output. A change to the
Secondary Port number may require a manual power cycle of the terminal before the change
becomes active.
3.8.5.3. FTP
The setup screen for FTP displays usernames and access levels for file download. Access levels for
all parameters are shown in Appendix B, Default Settings. Usernames can be edited, inserted, or
deleted.
3.8.5.3.1. To edit or insert a username
1. Press the UP and DOWN navigation keys to select (highlight) a username in the table.
2. Press the EDIT softkey to open the setup screen for editing a username or press the INSERT
softkey to open the setup screen to create a new username.
3. Press the ENTER key to select the Username field. The alpha keys display.
The Username field is accessible only when entering a new FTP user.
4. Use the alpha keys to edit or enter the desired username.
5. Use the Access selection box to assign the appropriate access level to the user. The following
access levels are available:
 Operator
 Supervisor
 Maintenance
 Administrator
6. Press the DOWN navigation key to view the second screen, which displays the Password and
Confirm Password fields.
7. Enter the desired password in the Password and Confirm Password fields.

3-66 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
8. Press the OK softkey to accept the username and password as entered.
9. Press the ESCAPE softkey to exit without saving the username and password.
Press the DELETE softkey to delete a username from the table on the Users screen.
To clear all usernames from the list and leave only the default username “admin”, press the CLEAR
softkey then OK to confirm the action.
3.8.5.4. Print Client
The Print Client connection allows the IND560 to send data to a specific IP address on the same
network as the IND560. To set up Print Client:
1. At Communication >Network > Print Client – enter the Server IP Address and Server TCP Port
number. The Server IP Address is the IP address of the device where the IND560 is sending the
print information. The Server TCP Port is the port number of the device on the network.
2. At Connections, press the INSERT softkey to define a new connection.
3. Under Port, scroll to Print Client and press ENTER.
4. Under the Assignment and Checksum fields, make desired selection.
5. Press the OK softkey to save the settings.
3.8.5.5. Email Alert
The IND560x can be configured to send email alerts when it encounters a calibration change or
calibration failure that might indicate a hardware problem, or when calibration expires. Note that the
Ethernet connection used to send email alerts is separate from those configured at Communication
> Connections.
3.8.5.5.1. Parameters
This screen permits four parameters to be set:
SMTP Server IP
This value is set using four groups of numbers, representing the IP address of the mail server the
IND560x will use to send alerts. Default is 0.0.0.0.
Sender Email Address
This is an alphanumeric entry field permitting the IND560x’s email address to be set. Default is
blank.
Sender Name
This is an alphanumeric entry field, used to give the IND560x a unique name that will appear in the
“From” field of alert emails. Default is “IND560x.”
Subject Line
This alphanumeric entry field defines the Subject field of the email to be set. Default is “Scale Cal
Alert.”
 

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-67
3.8.5.5.2. Recipients
This screen is used to define recipients of email alerts, together with the condition under which they
will receive an email. The following actions are available on the Email Recipients screen:
Softkey Function
EXIT Returns to the setup menu tree.
EDIT Opens the Email Recipients Edit screen, permitting parameters for the
selected recipient to be edited.
NEW Opens the New Email Recipient screen, allowing a new recipient to be
defined.
DELETE Deletes the selected recipient.
TEST Sends an email alert message, titled “Email Alert Test Message,” to the
recipient currently selected.
The NEW and EDIT screens offer the same options:
Email Address
This alphanumeric entry field is used to enter the recipient’s email address. Default is blank. Up
to 6 recipients can be configured.
Note – use the decimal key on the IND560x front panel to enter the period in each
email address.
Alert On
Available options are None, All, Failures. The default is None. Table 3-9 lists all the events that
trigger an email alert. Messages sent when Alert On is set to Failures are indicated by an
asterisk.
Table 3-9: Calibration Alerts
Event Email Message Subject
Completion of CalFree calibration CalFree complete
After each step in step calibration Calibration complete
Calibration expired Calibration expired*
Failing a calibration test Cal. test failed*
Passing a calibration test Cal. test passed
Completion of a zero capture from the Calibration
page
Zero captured
Completion of a span capture from the Calibration
page
Calibration completed

3-68 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
3.8.6. PLC
The PLC parameters are only available when a PLC option is installed in the ACM500. PLC setup
screens include:
 Analog Output
 A-B RIO
 DeviceNet
 PROFIBUS
 EtherNet/IP
 Modbus TCP
 Data Format
3.8.6.1. Analog Output
The analog output of the IND560x terminal provides either 4 - 20 mA or 0 - 10 volt DC outputs
depending on how the connection is wired. The output can be configured to represent either the
displayed weight or the gross weight.
3.8.6.1.1. To configure analog output
1. Choose the appropriate source from the Source field selection box. Selections are:
 None
 Application (Source as defined by a Task Expert application)
 ABS - Displayed Wt.
 ABS – Rate*
 Displayed Weight
 Gross Weight
 Rate*
*Available only when Rate is enabled
2. Select either None or Scale from the Channel selection box. A selection of None disables the
output.
3. On the second screen of this setup branch, the zero and span values for the analog output can
be adjusted. These values can be for weight or rate, depending on the selected source. In the
Zero Value field, enter the value at which the “zero” output of the analog signal should occur.
4. If these preprogrammed values do not provide the exact outputs required, they may be trimmed
by using the ZERO and SPAN softkeys located at the bottom of the screen as follows:
5. Press the ZERO softkey to initiate the zero fine-tuning process or the SPAN softkey to initiate
the span fine-tuning process. In either case, a warning screen offers two options – OK to
continue with the operation, or EXIT to return to the Analog Output screen without making any
adjustment.
6. The analog output signal value will change during this procedure – ensure that the PLC or other
control equipment is disconnected. A warning message displays asking for verification to
continue.
 

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-69
7. Once the OK softkey has been pressed, the zero fine-tuning or span fine-tuning operation
begins. A message screen appears, showing the current value of the analog output signal. Use
the softkeys that display to adjust the signal if necessary as follows:
Bump Down
(faster adjustment down)
Nudge Down
(slower adjustment down)
Bump Up
(faster adjustment up)
Nudge Up
(slower adjustment up)
8. Press the EXIT softkey to return to the previous screen.
3.8.6.2. A-B RIO
The Allen-Bradley Remote I/O (A-B RIO) network is an Allen-Bradley proprietary network protocol
that permits certain PLCs to communicate to additional racks of input and output devices or to other
peripheral devices that implement the RIO interface.
Additional interface information and programming examples can be found in the IND560 Terminal
PLC Interface Manual, which is included on the documentation CD.
The A-B RIO setup screen is used to configure A-B RIO as a PLC interface option.
3.8.6.2.1. Node Address
Each IND560x Terminal connected to the network represents one physical node; however, the
addressing of the node is defined as a logical rack address. This address is determined by the
system designer, then configured in the IND560x Terminal by selecting the Node Address text box
and using the numeric keypad to enter the appropriate node address (0–62).
Prior to the Node Address text box is a selection field allowing the user to enter the node address in
either:
 Decimal
 Octal
Figure 3-23 shows the new A-B RIO setup screen for Node Address.
Figure 3-23: A-B RIO Node Address Setup Screen
3.8.6.2.2. Start Quarter (Group)
Each scale occupies a quarter rack (single group) in the RIO address space and the quarter
(group) may be defined as the first (0), second (2), third (4), or fourth (6) quarter (group) of a

3-70 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
rack. Designate the location of the PLC that is the highest quarter used in a logical rack by using
the selection box to select the appropriate start quarter address 1–4 (group 0–6).
3.8.6.2.3. Last Rack
The IND560x configuration selections enable designation of the last rack. Use the selection box to
select Disabled or Enabled for last rack designation.
3.8.6.2.4. Data Rate
Use the selection box to select the desired data rate. Selections available are:
 57.6 Kb  115.2 Kb  230.4 Kb
3.8.6.2.5. Block Transfer
Use the Block Transfer selection box to either Disable or Enable the block transfer mode of
operation.
3.8.6.3. DeviceNet
The DeviceNet Interface can be configured to run up to 500Kbits per second depending on cabling
and distances. Messages are limited to 8 un-fragmented bytes. The network can include up to 64
nodes including the master, commonly called the scanner. The DeviceNet setup screen is used to
configure DeviceNet as a PLC interface option.
Additional interface information and programming examples can be found in the IND560 Terminal
PLC Interface Manual, which is included on the documentation CD.
3.8.6.3.1. Node Address
Each IND560x Terminal connected to the network represents one physical node. This address is
determined by the system designer, then configured in the IND560x Terminal by selecting the Node
Address text box and using the numeric keypad to enter the appropriate node address (0–63). The
default address is 63.
3.8.6.3.2. Data Rate
Use the selection box to select the desired data rate. Selections available are:
 125 Kb  250 Kb  500 Kb
125 kB is the default value.
3.8.6.4. PROFIBUS DP
The PROFIBUS DP Interface supports discrete data transfer that enables bi-directional
communication of discrete bit-encoded information or 16-bit binary word (signed integer)
numerical values.
Additional interface information and programming examples can be found in the IND560 Terminal
PLC Interface Manual, which is included on the documentation CD.
The PROFIBUS setup screen is used to configure PROFIBUS as a PLC interface option.
 

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-71
3.8.6.4.1. Node Address
Each IND560x Terminal connected to the network represents one physical node. This address is
determined by the system designer, then configured in the IND560x Terminal by selecting the Node
Address text box and using the numeric keypad to enter the appropriate node address (0–125).
3.8.6.4.2. Shared Data
The Shared Data parameter enables or disables an extended message length that includes shared
data access. This data is added to the end of the standard scale slot information and extends the
message length.
Use the Shared Data selection box to select Disabled or Enabled for shared data message
communication.
3.8.6.5. EtherNet/IP
EtherNet/IP, short for “EtherNet Industrial Protocol,” is an open industrial networking standard that
takes advantage of commercial, off-the-shelf EtherNet communication chips and physical media.
This networking standard supports both implicit messaging (real-time I/O messaging) and explicit
messaging (message exchange). The interface enables the IND560x terminal to communicate with
EtherNet/IP Programmable Logic Controllers (PLCs) through direct connection to the EtherNet/IP
network at either 10 or 100 MBPS speed.
EtherNet / IP interface has the following features:
 User-programmable IP addressing.
 Capability for bi-directional discrete mode communications (Class 1 Messaging) of weight or
display increments, status, and control data between the PLC and the IND560x.
Additional interface information and programming examples can be found in the IND560 Terminal
PLC Interface Manual, which is included on the documentation CD.
3.8.6.5.1. MAC Address
The Terminal’s MAC address in the Ethernet / IP network is assigned automatically.
3.8.6.5.2. IP, Subnet Mask and Gateway Address
The Terminal’s IP Address, Subnet Mask and Gateway Address are chosen by the system designer.
Default values are as follows:
IP Address: 192.168.0.1
Subnet Mask: 255.255.255.0
Gateway Address: 0.0.0.0
3.8.6.6. Modbus TCP
Additional interface information and programming examples can be found in the IND560 Terminal
PLC Interface Manual, which is included on the documentation CD. The Modbus TCP setup screen
is used to configure this PLC interface option.

3-72 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
3.8.6.6.1. MAC Address
The Terminal’s MAC address in the Modbus TCP network is assigned automatically.
3.8.6.6.2. DHCP Client
The DHCP Client is enabled by default. If disabled, the IP address must be assigned manually. If
enabled, the terminal will be assigned an IP address by the network server.
3.8.6.6.3. IP, Subnet Mask and Gateway Address
The Terminal’s IP Address, Subnet Mask and Gateway Address are configured as described in
Ethernet setup, above.
3.8.6.7. Data Format
Use the Data Format setup screen to configure the PLC data format.
3.8.6.7.1. Format
Select the desired PLC format from the Format field selection box. Selections include:
 Divisions - PLC format that provides scale increment data transfer.
 Floating Point - PLC format that provides 32-bit floating point data transfer in 16-bit integer
format.
 Integer - PLC format that provides 16-bit integer data transfer.
3.8.6.7.2. Byte Order
Select the desired byte order field selection box. Selections are:
 Word Swap - Default value. Takes the single-precision floating point format and swaps the two
words in the 32-bit double word. This format is compatible with RSLogix 5000 processors.
 Byte Swap - Makes the floating point format compatible with S7 Profibus.
 Historic - Makes the floating point format compatible with PLC 5 (selection only available with
A-B RIO interface)
 Double Word Swap - Makes the data format compatible with the Modicon Quantum PLC for
Modbus TCP networks (this selection is only available with EtherNet/IP and Modbus TCP
interfaces.
Table 3-10: Available Selections, by PLC Interface
A-B RIO PROFIBUS DeviceNet Ethernet / IP
Modbus TCP
Standard x x x
Word swap X x x x
Byte swap X x x x
Historic X
Double Word Swap x x x
 

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-73
Table 3-11: Byte Order
Word Swap Byte Swap Double Word Swap Standard /
Historic
Weight value, terminal 1355
PLC 15 Bit # 0 15 Bit # 0 15 Bit # 0 15 Bit # 0
Number,
Integer/Division
Weight value
Word 0x054B Hex 0x4B05 Hex 0x4B05 Hex 0x054B Hex
Floating point value Word First weight 0x6000 Hex 0xA944 Hex 0x0060 Hex 0x44A9 Hex
Second weight
value Word 0x44A9 Hex 0x0060 Hex 0xA944 Hex 0x6000 Hex
3.8.6.7.3. Message Slots
This parameter refers to the number of slots of data that will be transferred. Typically only one slot is
used in a single scale application, but sometimes an application requires multiple fields of data in
a cyclic data transfer. In order to accomplish this, two, three, or even four message slots can be
assigned for the output. As the number of message slots is increased, so is the length of the
message, because each slot requires enough memory to transmit a complete set of data. More
detailed information can be found in the IND560x Terminal PLC Interface Manual, which is included
on the documentation CD.
Choose the number of message slots from the selection box. Press ENTER when done.
3.8.7. Reset
The Reset setup screen restores Communication branch setup values to their factory default
settings.
3.8.7.1.1. Communication Reset
To initiate a reset, press the OK softkey . If the reset was successful, a verification message that
reads “Reset Successful” displays If the reset was not successful, an error message that reads
“Reset Failure” displays. If the reset fails, try to initiate the reset again. If the reset continues to fail,
contact a local METTLER TOLEDO representative for assistance.
Press the ESCAPE softkey to exit without resetting.

3-74 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
3.9. Maintenance
The maintenance setup branch includes:
Configure/View of the log files and calibration test
Running diagnostics or the calibration test
3.9.1. Configure/View
The Configure/View setup branch includes the following screens:
 Change Log  Calibration Test
 Maintenance Log  Reset
 Calibration Management
3.9.1.1. Change Log
The change log file tracks all changes to setup and shared data. The Change Log is approximately
150k bytes in size. Each record could vary in length, but an average of about 2,500 records can
be saved.
Use the selection box on this setup screen to select Disabed or Enabled for the change log.
3.9.1.1.1. To reset all records in the change log file
1. Press the RESET softkey . A warning message displays asking for verification that all
configuration change records are to be reset.
2. Press the OK softkey . The configuration change records history is reset.
3. If the reset was successful, a verification message that reads “Reset Successful” displays. If the
reset was not successful, an error message that reads “Reset Failure” displays. If the reset fails,
try to initiate the reset again. If the reset continues to fail, contact a local METTLER TOLEDO
representative for assistance.
4. Press the ESCAPE softkey to return to the Change Log screen.
To view change log records:
5. Press the VIEW TABLE softkey . The Change Log Search screen displays.
6. Use the Search Field 1 and Search Field 2 selection boxes and associated data fields to enter
specific search information to limit the search by date, shared data (SDName), Time, or
Username, or enter * (the “find all” character) to view all Change Log information.
7. Press the START SEARCH softkey . The Log Search View screen displays with the search
results sorted in chronological order (The most recent change record displays at the end of the
file with focus. Older records may be viewed by pressing the UP navigation key.).
Information displayed in the Change Log View includes:
 Time Stamp
 Username
 

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-75
 Shared Data Field Name
 New Value
8. Press the PRINT softkey to print the list to all Report connections.
9. Press the EXIT softkey to return to the Change Log Search Screen.
3.9.1.2. Maintenance Log
The maintenance log is a record of routine maintenance or activities such as capture span and
capture zero. Use the selection box on this setup screen to select Disabled or Enabled for the
maintenance log.
The Maintenance Log is approximately 150k bytes in size. Each record can vary in size, but an
average of about 2,500 records can be saved. More details regarding the Maintenance Log can be
found in Appendix C, Table and Log File Structure.
Reset all records or view records in the Maintenance Log by following the same steps described for
Change Log.
Information that displays on the Maintenance Log Search View screen includes:
 Time Stamp
 Username
 Event
 Status
3.9.1.2.1. New Maintenance Log Entry
The NEW softkey on the Maintenance Log screen opens an ADD MAINTENANCE RECORD
screen, intended to record the upgrade or repair of equipment. From this screen, an Event may be
added, removed or replaced, and its status recorded in an 8-character alphanumeric entry field.
Press OK to confirm the change, or ESC to exit the screen without changing the
Maintenance Log.
3.9.1.3. Calibration Management
Calibration management includes a test interval to program how much time elapses or how many
weighments should be completed between calibration checks. When this time or number of
weighments is exceeded, an expiration action is initiated. The type of expiration action is also
programmable. This feature is disabled by entering zeroes into both the number of days and
number of weighments entry boxes.
Use the Calibration Management setup screen to configure parameters used for managing
calibration activities.
3.9.1.3.1. Test Interval
Specify test intervals in days or weighments in the corresponding field text boxes.
The expiration action will be triggered when the first of either of these values is reached. For
example, if 30 days and 3,000 weighments were programmed, as soon as 30 days or 3,000
weighments were exceeded (whichever were first), the expiration action would occur.

3-76 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
3.9.1.3.2. On Expiration
Use the On Expiration selection box to configure the activity to occur upon expiration of the
calibration after the next test date or number of weighments occurs. Settings include:
a. No Action
b. Alarm Only (displays an expiration message)
c. Alarm & Disable (displays expiration message and disables the scale)
Note that, regardless of how the On Expiration option is configured, a pop-up message will appear
over the home screen when calibration expires.
3.9.1.3.3. Last Date Tested, Next Test Date, and # of Weighments Left
The Last Date Tested and Next Test Date (if days is specified for intervals), or the number of
weighments left until the next service (if weighments is specified for intervals), automatically
calculate and display.
To reset the Last Date Tested to the current date:
1. Press the RESET softkey .
2. A warning message displays asking for verification that calibration management values are to
be reset.
3. Press the OK softkey . A status screen displays that shows the reset process status. The Last
Date Tested is reset to the current date. This reset also generates a recalculation of the Next Test
Date and/or # of Weighments Left parameters if test intervals have been entered for these fields.
4. If the reset was successful, a verification message that reads “Values Reset OK” displays. If the
reset was not successful, an error message that reads “Reset Failed” displays.
5. Press the ESCAPE softkey to return to the Calibration Management screen.
3.9.1.4. Calibration Test
The calibration test provides a prompting, scale test sequence to help lead the person performing
the test through a test sequence. A test load with +/ tolerance is programmable for each step
along with two lines of prompting directions to guide the test person through each step. Up to 25
steps can be programmed for the sequence.
Use the Test Load Units selection box to select calibration test load units. Only weight units
available for primary units are available for selection.
The entire calibration test sequence can be cleared by pressing the CLEAR softkey , then pressing
OK to confirm the action. When the calibration test sequence is cleared (empty), it still shows the –
End– step. The –End– record will always display as the last step of the calibration test sequence.
This field cannot be deleted and cannot be edited.
Press the VIEW TABLE softkey to access the current calibration test sequence setup. The
Calibration Test View screen displays.
When the Calibration Test View screen opens, the step numbers and their associated test loads,
tolerances, and prompts display. Calibration test steps can be configured by:
 

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-77
Editing Inserting Deleting
To edit a calibration step:
1. When the Calibration Test setup screen opens, the first step in the list will have focus. Use the
UP and DOWN navigation keys to select a step to edit.
2. Press the EDIT softkey to open the setup screen for the step where edits can be made.
3. The step's test load and tolerance data displays first. Information for Prompt Field 1 and Prompt
Field 2 (up to 20 characters per field that display as separate lines during the calibration test)
display on the next screen. Edit the test step’s test load, tolerance, and prompts as desired.
4. Press the OK softkey to accept the calibration step parameters as entered.
5. Press the ESCAPE softkey to return to the Calibration Test View screen without saving the
changes to the calibration step parameters.
To insert a calibration step:
6. Use the UP and DOWN navigation keys to select the step in the calibration test procedure where
the new step is to be inserted.
7. Press the NEW softkey to open the setup screen to create a new calibration test step.
8. Enter the step's test load and tolerance data and information for Prompt Field 1 and Prompt
Field 2.
9. Press the OK softkey . The current screen data is stored at the indicated step number, and
any existing steps move down one record to make room for the new step.
10. Press the ESCAPE softkey to return to the Calibration Test View screen without saving the
calibration step.
Use the UP and DOWN navigation keys to select a step to delete. Press the DELETE softkey
to delete the step.
Press the PRINT softkey to print the calibration test steps.
Press the EXIT softkey to return to the Calibration Test screen.
3.9.1.5. Reset
The Reset setup screen resets setup values to factory default settings for the Maintenance
Configure/View setup.
3.9.1.5.1. Maintenance Reset
To initiate a reset, press the OK softkey . If the reset was successful, a verification message that
reads “Reset Successful” displays. If the reset was not successful, an error message that reads
“Reset Failure” displays. If the reset fails, try to initiate the reset again. If the reset continues to fail,
contact a local METTLER TOLEDO representative for assistance.
Press the ESCAPE softkey to exit without resetting.

3-78 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
3.9.2. Run
The Run screens enable the operator to:
 Run the calibration test
 View and run diagnostic tests
3.9.2.1. Calibration Test
The name of the person who previously ran the calibration test displays on the Run Calibration Test
screen. Press the ENTER key to change the name. Use the alpha keys to enter a new name. Press
ENTER.
To view and modify test weight information:
1. Press the TEST WEIGHT INFORMATION softkey to access the Test Weight View screen where
test weight information such as weight values and serial numbers can be edited, inserted, or
deleted.
2. Press the EDIT softkey to change a weight value or ID for a test weight, or press the NEW
softkey to insert a weight value and ID for a new test weight.
3. Press the OK softkey to save the changes or the new test weight record.
4. Press the ESCAPE softkey to exit without saving.
5. Press the DELETE softkey to delete a test weight from the list.
6. To clear all test weight records from the list, press the CLEAR softkey , then OK to
confirm the action.
7. Press the EXIT softkey to return to the Run Calibration Test screen.
To run the calibration test:
8. Press the START softkey to initiate the calibration test. The currently displayed weight unit is
compared to the programmed calibration test procedure weight unit.
If the units do not match, the IND560x terminal will automatically check other programmed
units and switch to the corresponding weight unit. If the unit selected for calibration does not
match the primary, second, or third units, an error message will display and the test cannot be
run. If this occurs, reprogram the unit selection or the calibration test unit to ensure that they
match.
9. Live scale weight (active weight) displays on the first line (under the system line).
10. The calibration test step (for example, 1/5, which indicates step 1 of 5) displays under the
active weight display.
11. The next line displays the test load value and +/– tolerance values for the step.
12. Prompt 1 displays on the line under the test load.
13. Prompt 2 displays on the next line.
14. Follow the prompts on the display and add the required test weights.
15. Press the OK softkey to proceed to the next calibration test step.
 

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-79
16. If a tolerance fails at any step, a Calibration Test Tolerance Failure message displays. Press
ENTER to acknowledge the error. There are now three possible steps:
a. If the wrong test weights were added to the scale, adjust the weights and press the OK
softkey to repeat the steps.
b. If the test weights were correct and the scale needs to be recalibrated, press the ESCAPE
softkey to abort the calibration test.
c. Press the SKIP softkey (a new softkey that displays when a Calibration Test Tolerance
Failure occurs), to accept the calibration test error and continue to the next step. This step
of the test will still report as a failure, but the test can be run to completion if desired.
17. After progressing through all of the steps in the calibration test procedure, a Test Complete
message displays along with a Status message of either Pass or Fail. Press the PRINT softkey
to print the calibration test report to connections with a Report assignment.
18. Press the ESCAPE softkey during any step to abort the calibration test and return to the
Run Calibration Test screen.
3.9.2.2. Diagnostics
Diagnostic test setup screens include:
 Display Test  Discrete I/O Test
 Keyboard Test - Local
 Scale
- Remote I/O 1
- Load Cell Output - Remote I/O 2
- Calibration Values - Remote I/O 3
- Statistics  Network Test
 Serial Test  Reset Flash 2
3.9.2.2.1. Display Test
The Display Test screen displays an alternate off/on dot block pattern when first accessed. After the
alternate off/on cycle, any additional test data about the display such as memory or program
information displays.
Press the EXIT softkey to return to the menu tree.
3.9.2.2.2. Keyboard Test
The Keyboard Test screen enables testing of the terminal keys, including:
 Softkeys
 Scale function keys
 Navigation keys
 Numeric keys
Press any key. The screen will display the key last pressed.

3-80 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
Press the EXIT softkey to return to the menu tree.
3.9.2.2.3. Scale
Scale diagnostics setup screens include:
 Load Cell Output
 Statistics
 Calibration Values
Load Cell Output
The Cell Output screen displays the current number of counts (active value) for the scale. This
display is only available for analog load cells.
Press the EXIT softkey to return to the menu tree.
Calibration Values
The Calibration Values screen displays the current calibration values configured for the scale. The
number of test loads that display calibration values is determined by the Linearity Adjustment
setting configured for the scale (see the Scale branch, Calibration setup).
These calibration values can be recorded and then manually entered into a new replacement board
should a failure ever occur, which eliminates having to recalibrate the scale with test weights. While
this method is quick, it is not as accurate as placing test weights on the scale.
Use the UP and DOWN navigation keys to select a calibration value to be modified. Use the numeric
keypad to enter new values.
Press the EXIT softkey to return to the menu tree.
Statistics
The Scale Statistics screen displays statistical information for the scale such as weighments
(increments each time a transaction is triggered), overloads (increments when the applied load for
a single load cell exceeds its overload capacity), peak weight (the maximum weight recorded by
the scale), and zero commands (increments each time a zero command is received from an
operator or remotely).
Use the UP, DOWN, LEFT, and RIGHT navigation keys to view all information and records.
Press the EXIT softkey to return to the menu tree.
3.9.2.2.4. Serial Test
The Serial Test screen enables testing of the sending and receiving hardware on the serial ports
COM1, OOM2, and COM3.
 

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-81
Com Port
Use the Com Port selection box to select the COM1, COM2 or COM3 serial port for testing .
NOTE: COM4 and COM5 cannot be tested in this manner.
Press the START softkey to initiate transmission of the test string. After pushing the START
softkey , it changes to a STOP softkey . Press this key to end transmission of the test string.
In the serial test mode, the terminal will transmit a string “Testing COMX nn” out the selected serial
port where the "X" is the selected com port number (1, 2, or 3) and “nn” is a sequential two-digit
number (00–99). If a jumper is placed between the transmit and receive terminals on that port, the
same data displays in the receiving field.
If another device is connected to the receiving port, any ASCII data received displays in the receiving
field.
The serial test continues until the STOP softkey is pressed, which returns the user to the main
Serial Test screen.
Press the EXIT softkey to return to the menu tree.
3.9.2.2.5. Discrete I/O Test
Discrete I/O Test setup screens include:
 Local
 Remote I/O 1
 Remote I/O 2
 Remote I/O 3
Select the internal (local) or external (Remote I/O) to be tested from the menu tree. Press ENTER.
VERY IMPORTANT!: When any of the Discrete I/O Test screens are first accessed, a warning
message displays with instructions to remove output control power before proceeding with the
test. The Discrete I/O Test screens enable manual setting of any of the outputs to on or off for
testing, so it is necessary to remove output control power before proceeding.
Press the ESCAPE softkey to abort and not perform the test.
To continue the test:
1. Press the OK softkey . A real-time display shows the status of each of the inputs and
enables each of the outputs to be turned on and off. An input or output that displays is
turned off. An input or an output that displays is turned on.
2. Use the LEFT and RIGHT navigation keys to select an output to turn on or off.
3. Press the DISCRETE OFF softkey to turn the output off, and press the DISCRETE ON softkey
to turn the output on.
4. Press the EXIT softkey to return to the menu tree. The outputs return to their previous on/off
configuration before the tests were initiated.

3-82 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Configuration
5. If three remote I/O modules are set up in the terminal, the IND560x disables the ability to run
diagnostics on a local I/O board that may be installed.
3.9.2.2.6. Network Test
The Network Test screen enables Ethernet network testing and assists in determining the availability
of network options. The status of each line associated with the network setup displays during and
after testing, showing a status of either Testing, Pass, Fail, or Timeout.
Press the EXIT softkey to return to the menu tree.
3.9.2.2.7. Reset Flash 2
The reset Flash 2 function clears the Flash 2 memory of all TaskExpert program files, custom tables
and bitmap files specific to TaskExpert custom programs that have been downloaded to the
terminal. Only the A1.csv (Tare Table csv file), A2.csv (Target Table csv file) and A3.csv (Container
Tare Table csv file present in the Fill-560 only) will remain in place and intact during a reset of the
Flash 2 drive.
Press the START softkey to initiate a reset of the Flash 2 memory. The terminal will NOT prompt
the user for confirmation and the reset will begin immediately. A “Reset successful” pop-up will
appear when the reset is complete. Press ENTER to acknowledge. At this time, the terminal will
complete a power cycle.
Press the EXIT softkey to abort and not perform the reset.
3.9.2.3. Reset All Factory Default Settings
The Reset All setup screen resets all setup settings to factory default settings.
The Reset All step resets all parameters in the terminal, except metrologically significant settings
such as Scale type, capacity, etc.
When the Reset All screen is first accessed, a message displays that asks for verification to reset all
setup parameters to factory default settings. To continue with the Reset All, press the OK softkey
. If the reset was successful, a verification message that reads “Reset Successful” displays If the
reset was not successful, an error message that reads “Reset Failure” displays. If the reset fails, try
to initiate the reset again. If the reset continues to fail, contact a local METTLER TOLEDO
representative for assistance.
Press the ESCAPE softkey to exit without resetting.
3.10. Restoring Factory Default Settings
Factory default settings can be restored individually for branches such as scale, application, and
terminal, or globally with the Reset All screen under the Maintenance branch. The Reset screen is
the last branch in each major branch of the menu tree (except for Maintenance). To restore factory
default settings for Terminal, for example:
1. Press the SETUP softkey . The setup menu tree displays (see Figure 3-24).
 

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 3-83
Figure 3-24: Setup Menu Tree – Reset
2. Use the DOWN navigation key to move the focus to Terminal.
3. Press the RIGHT navigation key to expand the Terminal branch.
4. Use the DOWN navigation key to move the focus to Reset.
5. Press the ENTER key to open the Terminal Reset screen.
6. Press the OK softkey to reset the Terminal setup values to factory default settings.
7. A status message appears that verifies a successful reset.
8. Press the EXIT softkey to return to the setup menu tree display.
9. Repeat steps 2–8 to reset factory default settings for any major branch in setup.
Select Reset All under Maintenance to restore all setup settings to factory defaults. Note that
neither this, nor Scale Reset, includes the reset of Type, Capacity, Increment, or Calibration
data. Reset this data by setting SW2-1 to its ON position, and then performing a master reset
by turning SW1-1 and SW1-2 ON and then cycling power to the terminal.

64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 4-1
This chapter covers
Cleaning and Maintenance
Service
Troubleshooting
External Diagnostics (Web Server)
4 Service and Maintenance
The IND560x terminal is designed to provide years of dependable operation.
However, METTLER TOLEDO recommends that – as with any industrial
measurement equipment – the IND560x terminal and the connected scale
system be serviced periodically. Timely, factory specified maintenance and
calibration by a METTLER TOLEDO service technician will ensure and document
accurate and dependable performance to specifications.
If unexpected problems do occur, METTLER TOLEDO recommends that service be done only by
qualified, trained personnel. To assist in this repair, record as much information as possible about
what has happened including any error messages and physical responses of the terminal and/or
scale.
4.1. Terminal Cleaning and Maintenance
To clean the IND560x terminal’s keypad and cover:
Gently wipe the IND560x terminal’s keypad and cover with a clean, damp, soft cloth.
Use water or mild, non-abrasive cleaning agents.
Do not use any type of acids, alkalis or strong industrial solvents such as toluene or
isopropanol (IPA) that could damage the terminal’s finish.
Do not spray cleaner directly on the terminal.
Do not clean the terminal using high-pressure or high-temperature water.
Build-up of dust layers must be avoided.
Remove light dust deposits using a damp cloth with a gentle wiping motion.
Do not use compressed air or vacuum to remove dust layers.
Follow good housekeeping practices to keep the terminal clean.
Regular maintenance inspections and calibration by a qualified service technician are
recommended.

4-2 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Service and Maintenance
WARNING
IF THE IND560x KEYBOARD, DISPLAY LENS OR ENCLOSURE IS DAMAGED, THE DEFECTIVE
COMPONENT MUST BE REPAIRED IMMEDIATELY. REMOVE POWER IMMEDIATELY AND DO
NOT REAPPLY POWER UNTIL THE DISPLAY LENS, KEYBOARD OR ENCLOSURE HAS BEEN
REPAIRED OR REPLACED BY QUALIFIED SERVICE PERSONNEL. FAILURE TO DO SO COULD
RESULT IN BODILY HARM AND/OR PROPERTY DAMAGE.
The IND560x is a rugged stainless steel enclosed instrument; however, the font panel is a polyester
covering over sensitive electronic switches and a lighted display. Care should be taken to avoid any
punctures to this surface or any vibrations or shocks to the instrument. Should the front panel
overlay become punctured, the IND560x should be taken out of service until the overlay can be
replaced.
4.2. Service
Only qualified personnel should perform installation, programming, and service. Please contact a
local METTLER TOLEDO representative for assistance.
In general, once the IND560x is installed, programmed, and calibrated for a given application, only
routine calibration service is required.
WARNING
ONLY PERMIT QUALIFIED PERSONNEL TO SERVICE THE TERMINAL. EXERCISE CARE WHEN
MAKING CHECKS, TESTS AND ADJUSTMENTS THAT MUST BE MADE WITH POWER ON.
FAILING TO OBSERVE THESE PRECAUTIONS CAN RESULT IN BODILY HARM AND/OR
PROPERTY DAMAGE.
4.2.1. General Safety Guidelines for Service
When performing on-site service or maintenance, only carry out this work on the weighing terminal
in the hazardous area if the following conditions are fulfilled:
The certification papers (conformity certificates, control drawing, manufacturer certifications,
etc.) are present and can be used to confirm the correctness and safety of the installation
The intrinsically safe entity parameters and area approval of the individual components are in
accord with one another,
The customer has issued a permit ("spark permit" or "fire permit")
The area has been rendered safe and the owner's safety coordinator has confirmed that there is
no danger
The necessary tools and any required protective clothing are provided (danger of the build-up
of static electricity)
 

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 4-3
4.3. Display Messages
4.3.1. Application Key Error
If, on power-up, the terminal finds that the hardware application key status has changed, a fault
message (Figure 4-1) will display. For instance, if the Fill-560 hardware key was added to a
standard IND560x, the fault would be displayed at power up.
Figure 4-1: Application Key Fault Message
Even though the ENTER graphic is shown, the ENTER key does not function. Two responses to this
error are possible:
If the hardware key status was not changed on purpose, turn power off and determine what
caused the status to change. Fix the problem and apply power again. The terminal will retest
the hardware key status.
If the hardware key status was changed on purpose by adding or removing a key, perform a
manual Master Reset – refer to Master Reset under Troubleshooting, below.
NOTE: Back up the terminal’s current setup parameters before the hardware key status is
changed. The original setup can then be downloaded to the terminal after the Master Reset has
been performed.
4.3.2. COM4 Error / COM5 Error
A “COM4 Error” and/or “COM5 Error” message will display in the system line of the IND560x
(Figure 4-2) if:
The setting at Communication > Connections > COM4 or COM5 is ACM500, and the
IND560x does not register communication from the ACM500 when it is powered up, or
Communication is lost between the IND560x and ACM500.
For troubleshooting assistance, refer to the Internal Diagnostic Testing section later in this chapter.

4-4 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Service and Maintenance
Figure 4-2: COM4 and COM5 Communication Errors
4.3.3. Invalid Parameters
The IND560x might display an “Invalid Parameters” message in the system line when exiting the
setup menu (Figure 4-3). This will happen when the weighing mode has been set to “Over/Under”
at Application/Memory/Target Table/Mode and there is no active Target value established. As long
as a Target is setup up by recalling a record from the Target Table or manually setting up a value
with the Target softkey, the “Invalid Parameters” message will no longer appear.
Figure 4-3: Invalid Parameter Message
4.3.4. Battery Alarm
When the battery voltage falls to its shut-off level, a message will appear on screen – “ALARM –
Low Battery! Power off is in process!”. This message will display for ten seconds, after which the
terminal will power off automatically.
4.4. Setting a Custom Language
The IND560 terminal allows custom languages to be selected for display messages.
4.5. Troubleshooting
Troubleshooting activities for the IND560x include:
Internal Diagnostic Testing
 

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 4-5
4.5.1. Internal Diagnostic Testing
The IND560x provides several internal diagnostic tests that are accessible in setup mode. Press the
SETUP softkey to view the setup menu tree. Use the DOWN navigation key to scroll down the
menu tree to Maintenance. Press the RIGHT navigation key to expand the menu tree selections for
Maintenance. Scroll down and expand Run. Scroll down and expand Diagnostics. Available
diagnostic setup screens include the following.
4.5.1.1. Display Test
Displays an alternate off/on dot block pattern when accessed. After the alternate off/on cycle, the
display will return to the menu tree.
4.5.1.2. Keyboard Test
Enables testing of the keyboard. Each key that is pressed will be shown on the display. When
testing is complete, press the EXIT softkey to return to the menu tree.
4.5.1.3. Scale
4.5.1.3.1. Load Cell Output
Displays the current load cell output (active weight) for the scale. This is useful to determine exactly
how many counts the analog section is outputting and is not available with IDNet bases.
4.5.1.3.2. Calibration Values
Displays the current calibration values for the scale. If these values are recorded after a scale
calibration, and the Main PCB is replaced in the future, the calibration values can be manually
entered here to “transfer” the previous calibration to the new Main PCB. This is not available with
IDNet bases.
4.5.1.3.3. Statistics
Displays statistical information for the scale such as the total number of weighments, the number of
scale overloads, the peak weight weighed on the scale, total number of zero commands and failed
zero commands. These are very helpful when diagnosing scale problems.
4.5.1.4. Serial Test
Enables testing of the transmit and receive functions on serial ports COM1, COM2 and COM3 (if
installed in ACM500). COM4 and COM5 on the fiber optic and current loop interfaces of the
IND560x can also be tested in this manner with special conditions.
1. Place a jumper wire between the transmit and the receive terminals of the COM port to be tested
as shown in Figure 4-4.
2. Select the COM port (COM1, COM2, COM3) to be tested using the selection box for COM Port.
Only the ports installed are available for testing.
3. Press the START softkey and a data string is output repeatedly approximately once every
three seconds. The data is: [Testing COMx: nn] where “x” is the COM port and “nn” is an
incrementing value beginning at 00 and continuing to 99. Each transmission increments this
number by one.

4-6 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Service and Maintenance
4. The same data string that is transmitted displays as the receiving field. If another device is
connected to the receiving port, any ASCII data received displays in the receiving field.
5. To stop the serial port test, press the STOP softkey .
Figure 4-4: Jumper Wire Between Transmit and Receive Terminals on COM1
4.5.1.5. Discrete I/O Test
Provides a view of the status of the discrete inputs and allows enabling or disabling of the discrete
outputs for diagnostic purposes. Expand the Discrete I/O branch by pressing the RIGHT navigation
key. Choose if the test will be performed on the local (internal) discrete I/O option or one of the
remote ARM100 modules. Press ENTER when that branch is highlighted.
Next, a warning will display to remind the tester that the outputs can be turned on manually during
this test, so any control power to the discrete outputs should be removed.
WARNING
THE DISCRETE OUTPUTS OF THE IND560x TERMINAL WILL BE MANUALLY ENABLED
DURING THIS TEST. REMOVE OUTPUT CONTROL POWER SO EXTERNAL EQUIPMENT
WILL NOT BE ENERGIZED BY MISTAKE. EXERCISE CARE WHEN MAKING CHECKS, TESTS
AND ADJUSTMENTS THAT MUST BE MADE WITH POWER ON. FAILING TO OBSERVE
THESE PRECAUTIONS CAN RESULT IN BODILY HARM AND/OR PROPERTY DAMAGE.
When viewing the test screen, the input status will be shown at the top of the display and the output
status will be shown at the bottom of the display Figure 4-5). Focus is shown on Output #1 first.
Figure 4-5: Discrete I/O Test Screen
 

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 4-7
Focus can be moved to any of the other outputs by using the RIGHT and LEFT navigation keys.
The two softkeys at the bottom of the page will either enable or disable the highlighted
output.
Note that when the Discrete I/O Test screen is exited, all outputs will be turned off again.
4.5.1.6. Network Test
Enables testing of the Ethernet programming/hardware/firmware in the terminal. The following tests
are run during this diagnostic process:
TCP/IP Stack
This IP Address
Gateway address
Each step will indicate “Testing” during the test then indicate a Pass or Timeout status. If the status
is shown as Timeout, it indicates that the IP address or gateway address has not been programmed
in the terminal (refer to Communication | Network in Chapter 3, Configuration for information on
entering the IP and gateway addresses).
4.6. External Diagnostics
4.6.1. Web Server
The IND560x includes a built-in web server utility that provides valuable tools for examining the
operation of the terminal, its installed options and software. Information available in the Web Server
includes:
System hardware and software configuration
Resource utilization
Scale statistics and data
Web server help can also be used for the following functions:
Reading shared data variables, viewed either statically or dynamically
Writing shared data values
Remote terminal reset
The Web Server in the IND560x is accessed with a web browser via the Ethernet connection
installed in the ACM500. Access to the Terminal is provided by using its IP address. An IP address
can be manually assigned, or assigned automatically by the network, depending on how DHCP is
configured in setup at Communication > Network > Ethernet. If DHCP client is enabled, once the
terminal is connected to the Ethernet its IP address will appear in the Ethernet configuration screen.
After an IP address for the IND560x has been entered or automatically assigned using DHCP,
record the address and enter it as the URL in the browser.
If the connection was successful, the Web Server home page (Figure 4-6) is displayed.

4-8 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Service and Maintenance
4.6.1.1. Home Page
When the Web Server is first accessed, the Home page shown in Figure 4-6 appears in the browser
window.
Figure 4-6: Web Server Home Screen
Click the links on this page to visit the respective views, each of which is detailed in the following
sections.
At the bottom of each page accessed, a HOME button is shown which returns the view to the Home
page seen in Figure 4-6 when pressed.
 

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 4-9
4.6.1.2. View Pages
4.6.1.2.1. Help File
A help file, accessible from each page via the help button explains the full range of the utility’s
features. Refer to this file for a full explanation of each of the items visible in each page.
Figure 4-7 shows a portion of the help file.
Figure 4-7: Help File
At the bottom of the Help page, a CLOSE WINDOW button allows the user to close this browser
window without exiting the Web Server utility.

4-10 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Service and Maintenance
4.6.1.2.2. System Information
The system information page (Figure 4-8) provides information about asset information, installed
component part numbers, software versions and serial numbers, and installed application software.
The Channel Software column is not filled unless the terminal is using an IDNet scale base.
Figure 4-8: System Information Page
 

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 4-11
4.6.1.2.3. Resource Utilization
The resource utilization page (Figure 4-9) gives a snapshot of the current status of the terminal,
including power on time and information about the various types of memory used by the IND560x.
Figure 4-9: Resource Utilization Page

4-12 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Service and Maintenance
4.6.1.2.4. Scale Statistics
The scale statistics page (Figure 4-10) shows a variety of data for the scale.
Figure 4-10: Scale Statistics Page
 

 

manuals search engine
64062161 | 09 | 11/2016 METTLER TOLEDO IND560x User's Guide 4-13
4.6.1.2.5. Scale Data
The scale data page (Figure 4-11) shows information, including current configuration and
calibration values, for the scale. The signal quality display permits troubleshooting of the
connection between the IND560x and load cells. When quality is poor, a red bar is displayed. A
yellow bar indicates adequate quality, and green bar means the quality is good, as in the example
below.
Figure 4-11: Scale Data Page
4.6.1.3. Diagnostics & Maintenance Pages
4.6.1.3.1. Shared Data
The shared data page (Figure 4-12) allows the user to request a display of current values for up to
ten shared data variables at a time. Values are accessed by entering the shared data name in one

4-14 METTLER TOLEDO IND560x User's Guide 64062161 | 09 | 11/2016
Service and Maintenance
of the fields at left. When the Save Changes button at bottom is clicked, the Name fields are saved
and the Values refreshed. The button is in focus when a new entry has been made in a Name field.
Figure 4-12: Shared Data Page
To refresh the displayed values without changing any variable names, click on the Manual Refresh
button.
Values may also be refreshed automatically, at one second intervals, by clicking on the Start
button. Once Start is selected, the Stop button is in focus and can be clicked. The current status of
dynamic display (Stopped or Running) is shown in a field between the two buttons. Note that, as
indicated on the page, Auto Refresh must be stopped before new shared data names can be
entered.
For a complete list of available shared data names, refer to the IND560x Shared Data Manual. In
the example shown in Figure 4-12, the shared data information displayed is as follows:
Table 4-4: Shared Data Information
Name Data Displayed Interpretation
wt0101 29.2 Current gross weight value
xp0117 5585 Cumulative power on time, in minutes
ce0104 1 Number of ranges defined for the scale
ce0123 400000 Encoder counts at the middle calibration value for the scale
ce0125 16 GEO code set in the Scale 2 calibration page
nt0102 172.18.49.120 Ethernet IP Address
wt0103 kg Primary units set in the scale calibration page
When a string is longer than the data display width of 40 characters, the left/right scroll buttons
may be used to move through the string to view all of it.